PCB Piezotronics M132A31 Installation And Operating Manual Download Page 8

 

ICP

 is a registered trademark of PCB Piezotronics, Inc. 

 

 

 

Drawing Number: 21094 
Revision: A 
ECO #: 35764 

 

OPERATION MANUAL FOR MICROSENSORS 

132

 

Series 

 

 3 

4.1 

OPERATION BEYOND RANGE 
SPECIFICATION
 

 
Because  132  Series  Microsensors  can  be  installed 
using  RTV-like  potting  materials  to  hold  them  in 
place,  maximum  pressure  specifications  are  given  to 
prevent  the  sensors  from  being  extruded  from  the 
mounting 

holes. 

 

Static 

or 

slow 

dynamic 

overpressures are more apt to cause a problem than a 
short dynamic shock pulse.  Use of backing nuts will 
minimize the possibility of a problem. 
 
The  microsensors  are  capable  of  responding  to 
dynamic  overpressures  up  to  ten  times  the  specified 
range.  The output will be “clipped” (look flat-topped 
on the readout) after about 10 volts. 
 
The initial rise of the output will be accurate up to the 
point  of  clipping.    This  can  allow  for  an  accurate 
time-of-arrival  measurement,  even  though  the  peak 
pressure cannot be measured. 
 

 
5.0 

CALIBRATION 

 
A shock tube is used to calibrate the microsensors. 
 

 
6.0 

MAINTENANCE AND REPAIR 

 
Except  for  repair  of  wires,  the  microsensors  are  not 
field-repairable.    In  case  of  serious  malfunction, 
contact the PCB at (888) 684-0011. 
 
It  is  well  to  observe  the  following  precautions  in 
using the microsensors: 
 
1.

 

Do  not  exceed  specified  pressure  levels  without 
proper mounting. 

 
2.

 

Do  not  subject  microsensors  to  temperatures 
exceeding 174 

°

F (79 °C). 

 
3.

 

Do not apply voltage to sensors without current-
limiting diodes or other current protection. (PCB 
ICP

  power  supplies  feature  proper  voltage  and 

current protection.) 

 
4.

 

Do not apply more than 20 mA of current to the 
microsensors. 

 

Summary of Contents for M132A31

Page 1: ...pigtail conn for high frequency Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ... repair User servicing or repair is not recommended and if attempted may void the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment calibration cycles are typically established by the users own quality regimen When in doubt about a calibration cycle a good rule...

Page 3: ...ory customer service representative Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and mounting hardware are not covered by warranty Mechanical damage...

Page 4: ... O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ...te O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at le...

Page 6: ...ors is not attenuated nor is noise appreciably increased by the cable 2 0 DESCRIPTION 132 Series Microsensors feature small diameter pressure sensing surfaces incorporating very small 1 mm piezoelectric sensing elements The small size of the sensing elements imparts extremely fast response time to the microsensors The charge generated by deflection of the piezoelectric element when subjected to sh...

Page 7: ...on of indicator panel the IC amplifier is providing proper bias normally 11 VDC cable connections are normal and the system is ready to operate If the needle moves to the red area of the fault monitor meter output is zero and a short is indicated The short could be located in the amplifier cable connectors or in the power unit If the pointer moves into the yellow area of the fault monitor meter an...

Page 8: ...look flat topped on the readout after about 10 volts The initial rise of the output will be accurate up to the point of clipping This can allow for an accurate time of arrival measurement even though the peak pressure cannot be measured 5 0 CALIBRATION A shock tube is used to calibrate the microsensors 6 0 MAINTENANCE AND REPAIR Except for repair of wires the microsensors are not field repairable ...

Page 9: ...ess Steel Stainless Steel Sealing Epoxy Epoxy Electrical Connector Integral Twisted Pair Integral Twisted Pair Weight without clamp nut 0 018 oz 0 52 gm 3 All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc M Metric Mount N Ne...

Page 10: ......

Reviews: