CANISTER LOAD CELL OPERATION MANUAL 

 

3

 

5.0 ELECTRICAL INSTALLATION 

 
 

 
 
 

5.1 Electrical Drawing / Western Regional Std. 

All  load  cells  are  wired  following  the  Western  Region 
Standard.    All  models  utilize  strain  gages  configured  into  a 
Wheatstone  Bridge  Circuit  to  produce  the  primary  sensing 
element.  The  four-arm  Wheatstone  bridge  configuration  is 
shown below in Figure 3. 

 
 
 
 
 
 
 
 

The  gages  are  bonded  to  the  load  cell's  structure. 
Typically,  a  regulated  DC  or  AC  excitation  is  applied 
between A and D of the bridge. When a force is applied to 
the  load cell,  the Wheatstone  bridge  becomes  unbalanced, 
causing  an  output  voltage  between  B  and  C,  which  is 
proportional to the applied load.  This configuration allows 
for  temperature  and  pressure  compensation,  as  well  as 
cancellation  of  signals  caused  by  forces  not  directly 
applied to the axis of the applied load. Output is typically 
expressed in units of mV/V of excitation. 
 

5.2 Output Polarity 

The following figure describes the output polarity. 

 

 

 

 
 
 
 
 
 

 

 
 
 

5.3 Cable & Grounding Considerations 

Proper  grounding  and  shielding  is  required  to  prevent 
electrical  noise  in  strain  gage  load  cell  measuring  systems.  
The cable must be shielded twisted pairs with a drain wire. 
 
Cable shields must be grounded only at one end, for example, 
on  the  instrument  or  control  system  ground.    The  load  cell 
case is grounded by mechanical attachment to the structure to 
which it is mounted. 
 
The  instrument  or  control  system  is  grounded  through  its 
power cord.  Ground loops and measuring system wiring may 
result in unstable or noisy signals. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
A  simple  test  with  a  voltmeter  connected  between  the  power 
cord  ground  and  the  structure  on  which  the  load  cell  is 
mounted  can  confirm  that  the  structure  has  been  properly 
grounded.  If the power cord ground and structure ground are 
not  at  the  same  potential,  it  may  be  necessary  to  provide  a 
secure  structure  ground,  perhaps  by  driving  a  copper  rod  and 
attaching a ground strap. 
 
 

6.0 CALIBRATION 

 

Every canister load cell manufactured by PCB Load & Torque 
has  been  fully  calibrated  per  ISO/IEC  17025  procedures,  and 
meets  all  published  specifications.    Each  load  cell  will  come 
with  a  calibration  certificate  designated  with  matching  model 
and  serial  numbers.  PCB  Load  &  Torque  also  offers 
calibration services on an on-going basis. 

 
 

6.1 Calibration Certificate Description 

Calibration  reports  supplied  with  canister  load  cells  contain 
valuable  information  to  assist  the  customer  in  use  of  the 
equipment.  A  separate  calibration  report  is  provided  for 
tension  and  compression  calibrations.  Calibration  procedures, 
equipment, and reports comply with ISO/IEC 17025. 

Table 4

 

- Electrical Connections 

Figure 5

 

- Grounding 

 

Test Rig Ground 

Power Supply 

Instrument Ground 

Cable Shielding 

Load Cell 

 

- Signal White 

- Excitation Black 

+ Excitation Red 

+ Signal Green 

Figure 3 - Western Regional Strain Gage Committee Wiring Code 

Figure 4

 

- Output Polarity 

Load Cell Receptacle: 

PT02E-10-6P 

Mating Connector: 

PT02A-10-6S 

PCB Load & Torque P/N: 

181-012A 

 

 

Tensile Load 

= Positive 

Tensile Load 

= Positive 

 

Compression Load 

= Negative 

Compression Load 

= Negative 

Summary of Contents for M1102-05A

Page 1: ...Load Cell Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 6: ...ing Considerations 3 6 0 CALIBRATION 3 6 1 Calibration Certificate Description 3 6 1 1 Measured Output 4 6 1 2 Hysteresis 4 6 1 3 Best Fit Output 4 6 1 4 Strain Gage Measurements 4 6 1 5 Shunt Calibration Standard Resistor 4 6 1 6 Static Error Band SEB 4 7 0 SHUNT CALIBRATION DESCRIPTION 4 7 1 Resistor Value 4 7 2 Shunt Calibration Process 4 7 3 Estimating Shunt Resistor for a Given Load 4 8 0 MAI...

Page 7: ...nts 4 0 MECHANICAL INSTALLATION Loads applied to canister load cells must be even and centered Compensation linkages e g rod end bearing alignment couplers are recommended to help minimize load misalignment When installing fixtures be sure to thread the fixture to the load cell and not the load cell to the fixture If alignment fixtures are not used ensure loading surfaces are flat and parallel wit...

Page 8: ...instrument or control system is grounded through its power cord Ground loops and measuring system wiring may result in unstable or noisy signals A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structure has been properly grounded If the power cord ground and structure ground are not at the same potentia...

Page 9: ...leg of the bridge to apply the shunt resistor to for both tension and compression load simulation Typically tension is simulated by inserting the shunt resistor between the P and S connector leads Compression loading is simulated by inserting the shunt resistor between the S and P connector leads 7 1 Resistor Value Canister load cells have a nominal 2 0 mV V full scale output For a 700 ohm strain ...

Page 10: ...e signal leads 9 Check bridge balance 10 Keep a record of your observations correct problems or contact PCB factory for assistance 10 0 CALIBRATION REPAIR SERVICES PCB Load Torque offers calibration and repair services The PCB Calibration Laboratory in Farmington Hills Michigan is A2LA Accredited per ISO IEC 17025 Standard calibration certificates list five force points ascending and one point des...

Page 11: ...d No English Equivalent M6 x 1 6H Housing Material Painted Aluminum Painted Aluminum Sensing Element Strain Gage Strain Gage Deflection at Full Scale Capacity 0 003 in 0 08 mm Electrical Connector PT02E 10 6P PT02E 10 6P Electrical Connection Position Side Side All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right...

Page 12: ...ULL 1102 SERIES PT CONN CANNISTER LOAD CELL PTE 9 14 11 DECIMALS XX 01 XXX 005 DECIMALS X 0 3 XX 0 13 51277 PCB Load Torque Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Load and Torque Inc REVISIONS REV DESCRIPTION ECO A ADDED NOTE 1 WIRING CODE 38165 PT02E 10 6P RECEPTACLE T...

Reviews: