ROD END LOAD CELL OPERATION MANUAL 

 

5

 

8.0 MAINTENANCE 

 
Routine  maintenance  of  the  rod  end  load  cell  should  include 
cleaning  the  electrical  connectors,  housings,  and  mounting 
surfaces  with  solutions  and  techniques  that  will  not  harm  the 
physical  material  of  construction.    Make  sure  liquids  are  not 
allowed  to  migrate  into  devices  that  are  not  hermetically 
sealed.  Such devices should only be wiped with a damp cloth, 
and  never  be  submerged  or  have  liquids  poured  on  them.  
Never use a pressure washer on the load cells. 
 
Yearly calibrations are recommended to ensure that the unit’s 
outputs match the factory specifications. 
 
 

9.0 TROUBLE SHOOTING 

 
Proper performance of a load cell requires careful attention to 
both electrical and mechanical aspects of the measurement 
system.  A basic understanding of the electrical and 
mechanical installation requirements is recommended. 

9.1 Mechanical Trouble Shooting 

A mechanical checklist includes: 
 

1.

 

Check for proper installation of fixturing. 
 

2.

 

Check integrity of the fixturing. 

9.2 Electrical Trouble Shooting 

An electrical checklist should start with: 
 

1.

 

Check  cables  for  proper  wiring  and  make  sure 
connections are secure and proper. 

 
2.

 

Inspect for loose or dirty electrical connections. 

 
3.

 

Check for improper shield grounds. 

 
4.

 

Check  for  proper  grounding  of  the  structure  that  the 
load cell is mounted on. 

 
5.

 

Check  the  signal  conditioning  electronics  for  proper 
setup. 

 
6.

 

Check 

the 

insulation 

resistance 

of 

shielded 

conductors for short circuits. 

 
7.

 

Check  isolation  resistance,  load  cell  flexure  to 
conductors. 

 
8.

 

Check  load  cell  bridge  resistances,  (A-D)  excitation 
and (B-C) the signal leads. 

 
9.

 

Check bridge balance. 

 
10.

 

Keep a record of your observations, correct problems, 
or contact PCB factory for assistance. 

 

10.0 CALIBRATION / REPAIR SERVICES 

 
PCB  Load  &  Torque  offers  calibration  and  repair  services. 
The  PCB  Calibration  Laboratory  in  Farmington  Hills, 
Michigan is A2LA Accredited per ISO/IEC 17025. 
 
Standard  calibration  certificates  list  five  force  points 
ascending  and  one  point  descending.  Additional  data  points 
are available at extra cost upon request. A 60 K ohm (

±

0.1%) 

shunt  calibration  resistor  is  supplied  with  each  load  cell 
calibration. 
 
Certificate  information  includes  tabulated  measurement 
variable  data  zero  balance,  bridge  input/output  resistance, 
computer  nonlinearity  and  hysteresis,  static  error  band  (SEB) 
calculations and entries abilities and traceability statements. 

If an initial evaluation shows that a transducer requires repair, 
PCB will provide the customer with an estimate prior to taking 
any corrective action

 

10.1 RMA / Purchase Order 

Please  request  a  return  material  authorization  (RMA)  before 
sending  a  load  cell  back  to  the  factory  for  any  reason.  For 
calibration  services,  if  possible,  a  copy  of  the  purchase  order 
covering  the  requested  services  should  be  included  with  the 
returned load cell. 

 

11.0 WARRANTY 

 
Standard  warranty  on  rod  end  load  cells  covers  parts  and 
workmanship.    For  full  details,  refer  to  the  Warranty 
Statement supplied with each load cell. 
  
If  the  load  cell  is  defective  for  reasons  other  than  overloads, 
return  it  to  the  factory  for  detailed  evaluation.  Factory 
evaluation  may  show  that  the  load  cell  is  repairable  or  non-
repairable and if repair or replacement will be under warranty. 
If not under warranty, the customer will be contacted with the 
cost  of  repairs  and  recalibration.  Once  authorization  to 
proceed is received, a delivery date will be provided. 

 
 
 

Manual Number: 50253 

Manual Revision: A 

Revision Date: 6/6/13 

ECO Number:  41606 

 

Summary of Contents for LOAD & TORQUE 1380-01A

Page 1: ...tallation and Operating Manual For assistance with the operation of this product contact PCB Load Torque Inc Toll free 866 684 7107 24 hour SensorLine 716 684 0001 Fax 248 888 8266 E mail LTInfo pcbloadtorque com Web www pcbLoadTorque com ...

Page 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 6: ...rounding Considerations 3 6 0 CALIBRATION 3 6 1 Calibration Certificate Description 3 6 1 1 Measured Output 4 6 1 2 Hysteresis 4 6 1 3 Best Fit Output 4 6 1 4 Strain Gage Measurements 4 6 1 5 Shunt Calibration Standard Resistor 4 6 1 6 Static Error Band SEB 4 7 0 SHUNT CALIBRATION DESCRIPTION 4 7 1 Resistor Value 4 7 2 Shunt Calibration Process 4 7 3 Estimating Shunt Resistor for a Given Load 4 8 ...

Page 7: ...rod end load cells must be even and centered Compensation linkages e g rod end bearing alignment couplers are recommended to help minimize load misalignment When installing fixtures be sure to thread the fixture to the load cell and not the load cell to the fixture If alignment fixtures are not used ensure loading surfaces are flat and parallel with loads applied in line with the sensor Figure 2 D...

Page 8: ... The instrument or control system is grounded through its power cord Ground loops and measuring system wiring may result in unstable or noisy signals A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structure has been properly grounded If the power cord ground and structure ground are not at the same pot...

Page 9: ... load on a load cell The calibration certificate will indicate which leg of the bridge to apply the shunt resistor to for both tension and compression load simulation Typically tension is simulated by inserting the shunt resistor between the P and S connector leads Compression loading is simulated by inserting the shunt resistor between the S and P connector leads 7 1 Resistor Value Rod end load c...

Page 10: ...signal leads 9 Check bridge balance 10 Keep a record of your observations correct problems or contact PCB factory for assistance 10 0 CALIBRATION REPAIR SERVICES PCB Load Torque offers calibration and repair services The PCB Calibration Laboratory in Farmington Hills Michigan is A2LA Accredited per ISO IEC 17025 Standard calibration certificates list five force points ascending and one point desce...

Page 11: ...unting Thread 3 4 16 Female No Metric Equivalent Cover Material Anodized Aluminum Anodized Aluminum Sensor Material Anodized Aluminum Anodized Aluminum Sensing Element Strain Gage Strain Gage Deflection at Full Scale Capacity 0 0005 in 0 013 mm Electrical Connector PT02E 10 6P PT02E 10 6P All specifications are at room temperature unless otherwise specified In the interest of constant product impr...

Page 12: ...10 DECIMALS XX 01 XXX 005 DECIMALS X 0 3 XX 0 13 45488 PCB Load Torque Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Load and Torque Inc REVISIONS REV DESCRIPTION ECO B ADDED CHART FOR M1380 SERIES 04 26 18 PTE 48167 1 64 41 5 4 25 108 0 1 00 25 4 1 27 32 3 Ø 1 50 38 1 Ø 1 86 ...

Reviews: