PCB Piezotronics IMI SENSORS 638A06 Installation And Operating Manual Download Page 18

Model Number

638A 06

ICP® ACCELEROMETER

Revision: B
ECN #: 50419

P erfo rm an ce

EN G L ISH

SI

Sensitivity(± 10 %)

25 mV/g

2.54 mV/(m/s²)

[1]

Measurement Range

± 200 g

± 1,962 m/s²

Frequency Range(± 5 %)

600 to 24,000 cpm

10 to 400 Hz

[2]

Frequency Range

(± 3 dB)

240 to 90,000 cpm

4 to 1,500 Hz

Filter Type

Low Pass

Low Pass

Electrical Filter Corner Frequency

1.5 kHz

1.5 kHz

Electrical Filter Roll-off

12 dB/octave

12 dB/octave

Resonant Frequency

≥ 1,200 kcpm

≥ 20 kHz

[3]

Broadband Resolution(1 to 10,000 Hz)

1 mg rms

0.01 m/s² rms

[3]

Non-Linearity

± 1 %

± 1 %

[4]

Transverse Sensitivity

≤ 5 %

≤ 5 %

En viro n m en tal

Overload Limit(Shock)

1,000 g pk

9,810 m/s² pk

Temperature Range

-320 to +250 °F

-196 to +121 °C

Temperature Response

See Graph

See Graph

[3]

Electrical

Settling Time(within 1% of bias)

≤ 3 sec

≤ 3 sec

Discharge Time Constant

0.05 to 0.5 sec

0.05 to 0.5 sec

Excitation Voltage

18 to 28 VDC

18 to 28 VDC

Constant Current Excitation

1.6 to 20 mA

1.6 to 20 mA

Output Impedance

≤ 100 Ohm

≤ 100 Ohm

Output Bias Voltage

7 to 11 VDC

7 to 11 VDC

Spectral Noise(1 Hz)

600 µg/√Hz

5,890 (µm/sec

2

)/√Hz

Spectral Noise

(10 Hz)

120 µg/√Hz

1,180 (µm/sec

2

)/√Hz

[3]

Spectral Noise

(100 Hz)

36 µg/√Hz

355 (µm/sec

2

)/√Hz

[3]

Spectral Noise

(1 kHz)

25 µg/√Hz

250 (µm/sec

2

)/√Hz

[3]

Spectral Noise

(10 kHz)

6 µg/√Hz

59 (µm/sec

2

)/√Hz

Electrical Isolation

≥ 10

8

 Ohm

≥ 10

8

 Ohm

P h ysical

Size (Length x Height x Width)

2.25 in x 1 in x 1 in

57.4 mm x 25.4 mm x 25.4

mm

Weight

6.4 oz

181 gm

[3]

Mounting Thread

1/4-28 UNF

1/4-28 UNF

[5]

Mounting Torque

2 to 5 ft-lb

2.7 to 6.8 Nm

Sensing Element

Quartz

Quartz

Sensing Geometry

Shear

Shear

Housing Material

Stainless Steel

Stainless Steel

Sealing

Hermetic

Hermetic

[6]

Electrical Connector

2-Pin MIL-C-5015

2-Pin MIL-C-5015

Electrical Connection Position

Side

Side

[7]

[8]

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

ICP

®

 is a registered trademark of PCB Piezotronics, Inc.

EX

 - Hazardous Area Approval- contact factory for

specific approvals

M

 - Metric Mount

Supplied Accessory: Model 081M260 Thru bolt M6x1.0 (1)

OP TION A L  VERSION S

Optional versions have identical specifications and accessories as listed for the standard model except

where noted below. More than one option may be used.

N OTES:

[1]Conversion Factor 1g = 9.81 m/s².
[2]1Hz = 60 cpm (cycles per minute).
[3]Typical.
[4]Zero-based, least-squares, straight line method.
[5]1/4-28 has no equivalent in S.I. units.
[6]Design hydro tested to 1450 psi [10,000 kPa].
[7]See PCB Declaration of Conformance PS023 for details.
[8]See Model's “Instructions For Use” (IFU) for North American Conformity and Certification
Statements.

Date: 02/19/2020

Date: 02/19/2020

Date: 02/19/2020

Date: 02/19/2020

69962

Entered: LK

Engineer: gs

Sales: MC

Approved: BAM

Spec Number:

Phone: 800-959-4464
Fax: 716-684-3823
E-Mail: [email protected]

3425 Walden Avenue, Depew, NY 14043

SU P P L IED A C C ESSORIES:

Model 081A73 Captive mounting bolt 1/4-28 x 1.34" (1)
Model ICS-6 NIST traceable single axis amplitude response calibration from 10 Hz to upper +/- 3dB
point. (1)

Summary of Contents for IMI SENSORS 638A06

Page 1: ...4 to 1 5kHz side exit 2 pin conn Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 800 959 4464 24 hour SensorLine 716 684...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...ON A ECN NUMBER 49766 Operating Guide with Enclosed Warranty Information 3424 Walden Avenue Depew New York 14043 2495 Phone 716 684 0003 Fax 716 684 3823 Toll Free Line 1 800 959 4IMI Piezoelectric IC...

Page 7: ...ply low frequency and high frequency measurements Shear design sensors are preferred because of their inherent insensitivity to adverse environmental influences such as case or base strain and thermal...

Page 8: ...wer gain which is extremely close to unity and independent of input voltage Also the diode can be changed to supply higher currents for driving long cable lengths Constant current diodes as shown in F...

Page 9: ...nics However when operating AC coupled signal conditioners with readout devices having an input impedance of less than one megaohm thelow frequencyrangemaybeaffected 4 1 STANDARD STUD MOUNT This mount...

Page 10: ...derside of the mounting base Firmly press down on the assembly to displace any extra adhesive remaining under the base See Figure7 METHOD 2 Direct Adhesive Mount For restrictions of space or for conve...

Page 11: ...interference or radio frequency interference EMI or RFI To avoid ground loops thereshouldonlybeonegroundin the system It is recommended for permanent installations that the sensor becase isolatedand...

Page 12: ...ate Also any current not used by the cable goes directly to power the internal electronicsandwillcreateheat Thismaycausethesensorto exceed its maximum temperature specification For this reason donotsu...

Page 13: ...er the meter or LED and allow the sensor to power up If a faulty condition is monitored first check all system connections then check the functionality of the cable and signal conditioner If the syste...

Page 14: ...and former MIL STD 45662A and uses equipment directly traceable to NIST This assures an accurate calibration of relevant specifications The following includes a broad overview of the Back to Back Cal...

Page 15: ...A of constant current to extend battery life while continuous monitoring systems offer more current in order to drive longer cables Figure 16 shows a typical system for a sensor with integral electron...

Page 16: ...eive assistance via e mail at imi pcb com or visit our website at www pcb com 11 0 CUSTOMER SERVICE WARRANTY IMI a division of PCB Piezotronics guarantees Total Customer Satisfaction If at any time fo...

Page 17: ......

Page 18: ...UNF 5 MountingTorque 2 to5 ft lb 2 7 to6 8 Nm SensingElement Quartz Quartz SensingGeometry Shear Shear HousingMaterial Stainless Steel Stainless Steel Sealing Hermetic Hermetic 6 Electrical Connector...

Page 19: ...CASE ISOLATED ACCELEROMETER DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 KRM 12 13 18 KRM 12 13 18 RMB 12 13 18 68326 PCB Piezotronics Inc claims proprietary rights in the information disclosed here...

Reviews: