background image

Model Number

622B11

PRECISION INDUSTRIAL ICP® ACCELEROMETER

Revision: A

ECN #: 49082

Performance

ENGLISH

SI

Sensitivity(± 5 %)

100 mV/g

10.2 mV/(m/s²)

[2]

Measurement Range

± 50 g

± 490 m/s²

Frequency Range(± 5 %)

35 to 360,000 cpm

0.58 to 6000 Hz

[3]

Frequency Range

(± 10 %)

25 to 600,000 cpm

0.42 to 10,000 Hz

Frequency Range

(± 3 dB)

12 to 900,000 cpm

0.2 to 15,000 Hz

Resonant Frequency

1800 kcpm

30 kHz

[1]

Broadband Resolution(1 to 10,000 Hz)

50 µg

490 µm/sec

2

[1]

Non-Linearity

± 1 %

± 1 %

[4]

Transverse Sensitivity

≤ 5 %

≤ 5 %

Environmental

Overload Limit(Shock)

5000 g pk

49,050 m/s² pk

Temperature Range

-65 to +250 °F

-54 to +121 °C

Temperature Response

See Graph

See Graph

[1]

Enclosure Rating

IP68

IP68

Electrical

Settling Time(within 1% of bias)

≤ 5.0 sec

≤ 5.0 sec

Discharge Time Constant

≥ 0.8 sec

≥ 0.8 sec

Excitation Voltage

18 to 28 VDC

18 to 28 VDC

Constant Current Excitation

2 to 20 mA

2 to 20 mA

Output Impedance

<100 Ohm

<100 Ohm

Output Bias Voltage

8 to 12 VDC

8 to 12 VDC

Spectral Noise(10 Hz)

4.0 µg/√Hz

39.2 (µm/sec

2

)/√Hz

[1]

Spectral Noise

(100 Hz)

0.8 µg/√Hz

7.85 (µm/sec

2

)/√Hz

[1]

Spectral Noise

(1 kHz)

0.4 µg/√Hz

3.92 (µm/sec

2

)/√Hz

[1]

Electrical Protection

RFI/ESD

RFI/ESD

Electrical Isolation

>10

8

Ohm

>10

8

Ohm

Physical

Size (Hex x Height)

7/8 in x 2.06 in

22 mm x 52.3 mm

Weight(without cable)

3.3 oz

94 gm

Mounting Thread

1/4-28 Female

No Metric Equivalent

[5]

Mounting Torque

2 to 5 ft-lb

2.7 to 6.8 Nm

Sensing Element

Ceramic

Ceramic

Sensing Geometry

Shear

Shear

Housing Material

Stainless Steel

Stainless Steel

Sealing

Welded Hermetic

Welded Hermetic

Cable Length

10 ft

3 m

Cable Type

Polyurethane

Polyurethane

[6]

[7]

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

ICP

®

is a registered trademark of PCB Group, Inc. 

OPTIONAL VERSIONS

Optional versions have identical specifications and accessories as listed for the standard model 

except where noted below. More than one option may be used.

EX

- Hazardous Area Approval- contact factory for specific approvals

[8]

M

- Metric Mount

Supplied Accessory : Model M081A61 Mounting Stud 1/4-28 to M6 X 1 (1)

TO

- Temperature Output

Temperature Output Range

+36 to +250 °F

+2 to +121 °C

Temperature Scale Factor

5.56 mV/°F + 32

+10 mV/°C

Electrical Connections(Pin A)

Acceleration Output

Acceleration Output

Electrical Connections(Pin B)

Ground

Ground

Electrical Connections(Pin C)

Temperature Output

Temperature Output

NOTES:

[1] Typical.
[2] Conversion Factor 1g = 9.81 m/s².

[3] The high frequency tolerance is accurate within ±10% of the specified frequency.
[4] Zero-based, least-squares, straight line method.
[5] 1/4-28 has no equivalent in S.I. units.

[6] Twisted shielded pair.
[7] See PCB Declaration of Conformance PS023 or PS061 for details.

[8] EEx ia IIC T4.

SUPPLIED ACCESSORIES: 

Model 081A40 Mounting Stud (1)

Model ICS-1 NIST-traceable single-axis amplitude response calibration from 600 cpm (10 Hz) 
to upper 5% frequency (1)

Entered: LK

Engineer: gs

Sales: EGY

Approved: NJF

Spec Number:

Date: 2/4/2019

Date: 2/4/2019

Date: 2/4/2019

Date: 2/4/2019

37834 

3425 Walden Avenue, Depew, NY 14043

Phone: 800-959-4464
Fax: 716-684-3823

E-Mail: [email protected]

Summary of Contents for Imi Sensors 622B11

Page 1: ...celerometer Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 959 4464 24 hour SensorLine 716 684 0001 Fax 716 684 3823 E m...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...ON A ECN NUMBER 49766 Operating Guide with Enclosed Warranty Information 3424 Walden Avenue Depew New York 14043 2495 Phone 716 684 0003 Fax 716 684 3823 Toll Free Line 1 800 959 4IMI Piezoelectric IC...

Page 7: ...ply low frequency and high frequency measurements Shear design sensors are preferred because of their inherent insensitivity to adverse environmental influences such as case or base strain and thermal...

Page 8: ...wer gain which is extremely close to unity and independent of input voltage Also the diode can be changed to supply higher currents for driving long cable lengths Constant current diodes as shown in F...

Page 9: ...nics However when operating AC coupled signal conditioners with readout devices having an input impedance of less than one megaohm thelow frequencyrangemaybeaffected 4 1 STANDARD STUD MOUNT This mount...

Page 10: ...derside of the mounting base Firmly press down on the assembly to displace any extra adhesive remaining under the base See Figure7 METHOD 2 Direct Adhesive Mount For restrictions of space or for conve...

Page 11: ...interference or radio frequency interference EMI or RFI To avoid ground loops thereshouldonlybeonegroundin the system It is recommended for permanent installations that the sensor becase isolatedand...

Page 12: ...ate Also any current not used by the cable goes directly to power the internal electronicsandwillcreateheat Thismaycausethesensorto exceed its maximum temperature specification For this reason donotsu...

Page 13: ...er the meter or LED and allow the sensor to power up If a faulty condition is monitored first check all system connections then check the functionality of the cable and signal conditioner If the syste...

Page 14: ...and former MIL STD 45662A and uses equipment directly traceable to NIST This assures an accurate calibration of relevant specifications The following includes a broad overview of the Back to Back Cal...

Page 15: ...A of constant current to extend battery life while continuous monitoring systems offer more current in order to drive longer cables Figure 16 shows a typical system for a sensor with integral electron...

Page 16: ...eive assistance via e mail at imi pcb com or visit our website at www pcb com 11 0 CUSTOMER SERVICE WARRANTY IMI a division of PCB Piezotronics guarantees Total Customer Satisfaction If at any time fo...

Page 17: ......

Page 18: ...aling Welded Hermetic Welded Hermetic Cable Length 10 ft 3 m Cable Type Polyurethane Polyurethane 6 7 All specifications are at room temperature unless otherwise specified In the interest of constant...

Page 19: ...X SERIES ACCELEROMETER DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 KRM 2 4 19 KRM 2 4 19 GGS 2 4 19 37843 PCB Piezotronics Inc claims proprietary rights in the information disclosed hereon Neither i...

Reviews: