Page 1: ...Sensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail i...
Page 2: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...
Page 3: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...
Page 4: ...any case care must be taken such that the sensor s diaphragm is not oriented down to avoid ground reflections In some cases where flash temperatures such as those generated by blasts and shock fronts...
Page 5: ...hundreds or thousands of feet use a higher current up to 20 mA maximum Use of low capacitance cable i e RG 62 U and impedance matching variable resistors are suggested to improve long cable performanc...
Page 6: ...Aluminum Aluminum Diaphragm Invar Invar Sealing Epoxy Epoxy Electrical Connector BNC Jack BNC Jack Weight 12 4 oz 352 gm 2 All specifications are at room temperature unless otherwise specified In the...