- 5 - 

It is for this reason that PCB recommends that a recalibration cycle be established for each 
microphone. This schedule is unique and is based upon a variety of factors, such as: extent of 
use, environmental conditions, accuracy requirements, trend information obtained from previous 
calibration records, contractual regulations, and risk associated with incorrect readings. PCB 
recommends 12 to 24 month intervals, depending upon the above factors. In addition, it is also 
best practice to perform verification both before and after each test. PCB’s model CAL 250 
Acoustic Calibrator provides a simple method for accomplishing this. 

6.0 Service 

See the supplemental sheet, contained with this manual, for information on our service, repair 
and return policies, procedures and instructions. When unexpected problems arise, call our 24-
Hour SensorLine

SM

  (716-684-0001) to discuss your immediate dynamic instrumentation needs 

with a PCB Application Engineer.  

7.0 Warranty 

All equipment and repair services provided by PCB Piezotronics, Inc. are covered by a limited 
warranty against defective material and workmanship for a period of one year. Visit 

www.PCB.com

 for a complete statement of our warranty. 

 
PCB guarantees 

Total Customer Satisfaction

. If, at any time, for any reason, you are not 

completely satisfied with any PCB product, PCB will repair, replace, or exchange it at no charge. 
You may also choose within the warranty period to have your purchase price refunded.

 

 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 

3425 Walden Avenue, Depew, NY 14043-2495 

24-hour SensorLine

SM

: 716-684-0001 

E-Mail: 

[email protected]

  

 

            Fax: 716-684-0987 

Website: 

www.pcb.com

  

 

 

Toll-free: 800-828-8840 

 
 

ISO 9001 CERTIFIED  

  A2LA ACCREDITED to ISO 17025  

  AS9100 CERTIFIED 

 

In the interest of constant product improvement, specifications are subject to change without notice. 

 

 Copyright PCB Group, Inc. 2013 

PCB

 and ICP

 are registered trademarks of PCB Group, Inc. 

SensorLine

SM

 is a service mark of PCB Group, Inc. 

 
 
 

Manual Number: 46455 
Manual Revision: E 
ECO# 42623 
Printed in the U.S.A. 
 

Summary of Contents for ICP 130E21

Page 1: ...Microphone Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E ma...

Page 2: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Page 3: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Page 4: ...Calibrators etc you will ensure years of trouble free usage If after reading this manual you have additional questions concerning the Microphone or its application feel free to call a factory Applicat...

Page 5: ...ll ICP powered microphones require correct constant current excitation for proper operation For this reason use only PCB constant current signal conditioners or other approved constant current sources...

Page 6: ...e object you are measuring The Free Field type of microphone is best suited for applications where the sound is coming from one defined direction without any disturbing objects or hard surfaces that w...

Page 7: ...mV Pa Reference Freq 250 Hz 26 4 dB re 1V Pa Output Bias 8 6 VDC Temperature 71 F 22 C Relative Humidity 50 Condition of Unit As Found n a As Left New Unit In Tolerance Notes 1 This method involves co...

Page 8: ...ic instrumentation needs with a PCB Application Engineer 7 0 Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material an...

Page 9: ...aterial Stainless Steel Stainless Steel Electrical Connector Output 10 32 Coaxial Jack 10 32 Coaxial Jack Size Diameter x Length overall 0 28 in x 2 36 in 7 0 mm x 60 0 mm Weight 0 34 oz 9 7 gm 1 All...

Page 10: ......

Reviews: