background image

 

- 3 - 

normal operation and a “red” reading indicates a short circuit (example: a shorted cable). The 
signal conditioner provides a zero-based, AC-coupled, output signal that is compatible with most 
standard readout devices. 
 
Many FFT analyzers, data acquisition modules, and data collectors have the proper constant-
current excitation built-in for direct use with ICP

 microphones. Before using this feature, make 

sure the supply voltage and constant current are within acceptable limits for use with your 
particular microphone. (Check the enclosed 

Specification Sheet.) 

4.0 Installation and Operation 

There are many ways to support the microphone to take measurements, from simply placing the 
microphone in a clip to using a 2D array stand, as determined by your application and budget. 
After you remove the microphone from its package and have read this operating guide, attach the 
appropriate cable between the Signal Conditioner and the microphone. You will then need to 
attach a second cable to connect the Signal Conditioner to your readout device. This is a standard 
Coaxial cable with BNC connectors on each end. PCB offers these cables in a variety of lengths 
for any application. 
 
 

 

 

Figure 3 

 
All of the 130 series Array Microphones are the “Free-Field” type. This type of microphone is 
direction-sensitive and should be pointed directly at the object you are measuring. The “Free-
Field” type of microphone is best suited for applications where the sound is coming from one 
defined direction, without any disturbing objects or hard surfaces that will reflect the sound or 
cause reverberation. After completing the system setup, power on the signal conditioner and 
allow 1 to 2 minutes for the system to stabilize. The meter (or LED) on the signal conditioner 
should be reading “green”. This indicates proper operation and you may begin taking 
measurements. If a faulty condition is indicated (red or yellow reading), first check all system 
connections, then check the functionality of the cable and signal conditioner. If the system still 
does not operate properly, consult a PCB Application Engineer. 

 
 
 
 
 
 
 
 
 

When used in conjunction with a 
large number of signals, for 
simultaneous measurement, the 
microphones can be clipped onto a 
grid network to form a complete 
array. See Figure 3. 

Summary of Contents for ICP 130A24

Page 1: ...Microphone Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E ma...

Page 2: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Page 3: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Page 4: ...brators etc you will ensure years of trouble free usage If after reading this manual you have additional questions concerning the Microphone or its application feel free to call a factory Application...

Page 5: ...They utilize low impedance cable on which signals can be driven long distances with minimal signal loss 3 0 Powering All ICP powered microphones require correct constant current excitation for proper...

Page 6: ...adout device This is a standard Coaxial cable with BNC connectors on each end PCB offers these cables in a variety of lengths for any application Figure 3 All of the 130 series Array Microphones are t...

Page 7: ...a proper seal gently wipe the top of the microphone housing to remove any moisture or debris Apply the replacement water resistant patch to the end of the housing keeping it centered Finally assemble...

Page 8: ...mV Pa Reference Freq 250 Hz 26 4 dB re 1V Pa Output Bias 8 6 VDC Temperature 71 F 22 C Relative Humidity 50 Condition of Unit As Found n a As Left New Unit In Tolerance Notes 1 This method involves co...

Page 9: ...ic instrumentation needs with a PCB Application Engineer 7 0 Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material an...

Page 10: ...Excitation 2 to 20 mA 2 to 20 mA Output Bias Voltage 9 to 14 VDC 9 to 14 VDC Output Impedance 52 Ohm 52 Ohm Physical Housing Material Stainless Steel Stainless Steel Electrical Connector Output BNC J...

Page 11: ...FILLETS AND RADII 0 07 0 13 OUTLINE DRAWING 59447 1 OF 1 2X MODEL 130A24 DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 BB 5 4 15 ECB 5 4 15 DRK 5 4 15 bbehringer 12 46 5 4 2015 59447 PCB Piezotronics...