upon the coupling type or time constant of the external 
charge  amplifying  device.    This  characteristic  may  be 
fixed or adjustable, depending on the charge amplifier.  
Please  refer  to  the  appropriate  charge  amplifier 
specification  sheet  for  information  on  low-frequency 
characteristics. 
 

 

 

Figure 1.

  Assorted Mounting Configurations and Their 

Effects on High Frequency 
 

4.1 STUD MOUNT 

This  mounting  technique  requires  smooth,  flat  contact 
surfaces  for  proper  operation  and  is  recommended  for 
permanent and/or secure installations.  Stud mounting is 
also recommended when testing at high frequencies. 
 

Note:

  

Do NOT attempt mounting on curved, rough, or 

uneven  surfaces,  as  the  potential  for  misalignment  and 
limited  contact  surface  may  significantly  reduce  the 
sensor’s upper operating frequency range. 
 

 

 
Figure 2.

 Mounting Surface Preparation 

 
STEP 1:  

First, prepare a smooth, flat mounting surface, 

then  drill  and  tap  a  mounting  hole  in  the  center  of  this 
area  as  shown  in  Figure  2  and  in  accordance  with  the 

Installation  Drawing 

for  the  specific  sensor  that  is 

being mounted

 
A  precision-machined  mounting  surface  with  a 
minimum  finish  of  63 

in  (0.00016  mm)  is 

recommended.  (If it is not possible to properly prepare 
the  test  structure  mounting  surface,  consider  adhesive 
mounting  as  a  possible  alternative.)  Inspect  the  area, 
checking  that  there  are  no  burrs  or  other  foreign 
particles interfering with the contact surface. 
 

STEP  2:   

Wipe  clean  the  mounting  surface  and  spread 

on  a  light  film  of  grease,  oil,  or  similar  coupling  fluid 
prior to installation. 
 

 

 

Figure 3.

 Mounting Surface Lubrication 

 
Adding 

coupling 

fluid 

improves 

vibration 

transmissibility  by  filling  small  voids  in  the  mounting 
surface and increasing the mounting stiffness.  For semi-
permanent mounting, substitute epoxy or another type of 
adhesive. 
 

STEP 3:

  Screw the mounting stud into  the base of the 

accelerometer  and  hand-tighten.  Then,  screw  the 
sensor/stud  assembly  into  the  prepared  tapped  hole and 
tighten  to  the  recommended  mounting  torque  as 
indicated on the 

Installation Drawing

 

Note: It is important to use a torque wrench during this 
step.    Under-torquing  the  sensor  may  not  adequately 
couple  the  device;  over-torquing  may  result  in  stud 
failure.

 

 

4.2 ADHESIVE MOUNT 

Adhesive  mounting  is  often  used  for  temporary 
installation  or  when  the  test  object  surface  cannot  be 
adequately  prepared  for  stud mounting.  Adhesives like 
hot glue and wax work well for temporary mounts; two- 
part  epoxies  and  quick-bonding  gels  provide  a  more 
permanent mount. 
 

Note:  Adhesively  mounted  sensors  often  exhibit  a 
reduction in high-frequency range.   Generally, smooth 
surfaces  and  stiff  adhesives  provide  the  best  frequency 
response.

  

 

Summary of Contents for EX357C71

Page 1: ...elerometer Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E ma...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...F 482 C and extreme application up to 100 F 650 C for UHT 12 These accelerometers are ideal for structural testing machine monitoring and vehicular shock and vibration measurement tasks where high tem...

Page 7: ...urface consider adhesive mounting as a possible alternative Inspect the area checking that there are no burrs or other foreign particles interfering with the contact surface STEP 2 Wipe clean the moun...

Page 8: ...or most applications It is generally used only for machinery monitoring and other portable trending applications Both the accuracy and repeatability at low 5 Hz and high frequency 1 kHz ranges are que...

Page 9: ...o a precise sensitivity such as 10 0 or 100 0 mV g for ease of data analysis This is accomplished with most PCB laboratory charge amplifiers and some miniature in line units as well For fixed sensitiv...

Page 10: ...dure themselves While this may result in both a savings of time and money it has also been attributed to incorrect readings and costly errors Therefore in an effort to prevent the common mistakes asso...

Page 11: ...Figure 12 Typical Calibration Set Up Adhesively mounted sensors use similar practices However in this case a small portion of quick bonding gel or similar temporary adhesive is used to attach the tes...

Page 12: ...s Incorrect Understand Back to Back Calibration limitations Do not expect the uncertainty of calibration to be any better than 2 In fact the uncertainty may be as high as 3 or 4 for frequencies 10 Hz...

Page 13: ......

Page 14: ...DECIMALS X 0 8 XX 0 25 BB 9 2 15 ECB 9 2 15 GGS 9 2 15 bbehringer 14 57 9 2 2015 41341 PCB Piezotronics Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproducti...

Page 15: ......

Page 16: ......

Page 17: ...ONS METRIC DIMENSIONS IN BRACKETS 4 RECOMMENDED MOUNTING TORQUE 2 5 FOOT POUNDS 3 7 NEWTON METERS MOUNTING HOLE PREPARATION 218 5 54 X 300 7 62 MIN 1 4 28 UNF 2B X 200 5 08 MIN F M6 X 0 75 M6 X 1 00 M...

Page 18: ...tion Ces derni res ne sont pas couvertes par la pr sente attestation 12 Le marquage de l appareil ou du syst me de protection doit comporter les informations d taill es au point 15 Fontenay aux Roses...

Page 19: ...The equipment shall also bear the usual marking required by the manufacturing standards applying to such equipment 116 DOCUMENTS DESCRIPTIFS 16 DESCRIPTIVE DOCUMENTS Dossier de certification No 41124...

Page 20: ...ement la conception et la construction de l appareil ou du syst me de protection sp cifi conform ment a l annexe III de la directive 94 9 CE Des exigences suppl mentaires de la directive sont applicab...

Page 21: ...14 EC TYPE EXAMINATION CERTIFICATE LCIE 08 ATEX 6101 X 15 DESCRIPTION OF EQUIPMENT OR PROTECTIVE SYSTEM High temperature accelerometers Type 357 series The sensor is made of an housing with a piezo cr...

Page 22: ...stalled per product documentation 2014 34 EU 2011 65 EU w 2015 863 EU ATEX Directive RoHS Directive Standards to which Conformity is Declared Harmonized Standards EN 60079 0 2012 A11 2013 EN 60079 11...

Reviews: