441A152 PEAK METER     

 

Manual 23528 Rev. A 
ECN 50523 

Appendix A:  RS-232 Control 

 

For RS-232 communication, the port settings should be as follows: 

 
 

General 

RS-232,  DCE  interface,  XON/XOFF  flow  control, 
Asynchronous protocol 

Baud Rate 

9600 Baud 

Start Bits 

Data Bits 

Parity 

None 

Stop Bits 

 

The message format is as follows: 

 

Start  Byte 
(0x02) 

Destination ID 
(Rack/Slot) 

Command 

End Byte 

((0x03) 

Checksum 

 

Start Byte 

STX (0x02) - marks the start of a new message. 

Destination ID 

16 bits - rack code (0-3 ASCII) followed by slot address (0-9 ASCII). Messages meant 
for the master will have "MM" in this field. 

Command 

Contains  the  command  followed  by  any  associated  data.  See  below  for  command 
structure, and command set for examples. 

End Byte 

ETX (0x03) - marks the end of the message. 

Data Checksum  16 bits - the ASCII hex of the 8-bit sum, ignoring overflows, of all bytes in the message 

including SOT and ETX bytes. 

 
 

All messages sent to the rack will receive a response. A typical command transaction is shown below. 
 

 

Host    

 

Rack

 

 

Command 

 
 
 

 

 

 

 

 

       ACK [data] 

 

    NACK [reason] 

 

The  ACK/NAK  field  conveys  information  concerning  the  delivery  of  messages.  The  results  of  command 
operations are contained in the data field of an ACK message. NAK reason bytes are defined below. 

 

Start Byte (0x02) 

ACK (0x06) 

Data Field 

End Byte (0x03) Checksum 

The format of an ACK message. 

 

Summary of Contents for 441A35

Page 1: ...ol for Series 440 modules Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 71...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...41A43 441A45 AND 441A49 2 EQUIPMENT RATINGS 2 MULTIPLE CHASSIS CONFIGURATION 2 CHANGING THE FUSE 2 DISABLING COOLING FAN 2 AVAILABLE MODULES 2 AC POWER SUPPLY MODEL 441A101 3 DC POWER SUPPLY MODEL 441...

Page 7: ...tion High voltage electrostatic discharge ESD can damage electrical devices To avoid triboelectric transfer 1 Connect the cables only with the AC power off 2 Temporarily short the end of the cable bef...

Page 8: ...n 3 0 Ah NiMH rechargeable battery The optional Model 488A09 is a 12VDC 3 4A AC Power Adaptor that allows the 441A102 to be used with any AC line power in the world 100 240VAC 50 60Hz The 441A102 can...

Page 9: ...CHASSIS AND POWER SUPPLY 4 Manual 19680 Rev C ECN 50523 Appendix A Figures...

Page 10: ...CHASSIS AND POWER SUPPLY 5 Manual 19680 Rev C ECN 50523...

Page 11: ...Conditioner Card Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 09...

Page 12: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 13: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 14: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 15: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 16: ...3 OPERATION 3 SET UP AND MEASUREMENT 3 SETTING THE OPERATIONAL MODE 3 SETTING THE SENSOR SENSITIVITY 3 SETTING THE OUTPUT SENSITIVITY 4 SETTING THE TIME CONSTANT 4 ZEROING THE OUTPUT IN LONG TIME CON...

Page 17: ...s has its signal ground connected via the chassis to mains ground and b the housing of the transducer is isolated from grounded measurement sources If the measurement set up is mounted in a metal inst...

Page 18: ...of the charge amplifier the low pass filter and the output so overload in various parts of the circuitry are not masked by filtering The overload indicator is located on the front panel In ICP mode th...

Page 19: ...is selected in charge mode 443B102 only the display will briefly indicate Time Constant is 10 000 sec or Time Constant is 100 000 sec depending on the gain selected And then Autozero in Progress The...

Page 20: ...Features The 443B incorporates several features which allow the use of sensors with TEDS electronics The 443B can read the raw data from any TEDS sensor and is able to locally decode the data from sen...

Page 21: ...surements Using the Setup Memory The 443B has the capability to save and recall up to five different setups so that the user can easily switch between frequently used configurations To save a setup 1...

Page 22: ...associated data See below for command structure and command set for examples End Byte ETX 0x03 marks the end of the message Data Checksum 16 bits the ASCII hex of the 8 bit sum ignoring overflows of...

Page 23: ...cating with internal module T Time out message not sent rack slot address may be incorrect NAK Reason Bytes Command format is shown below Module Type Three ASCII printable characters This field identi...

Page 24: ...t is a common or CMM command Command String XYCMMMMOD Description X Rack number of the target module Range ASCII printable characters 0 through 3 Y Slot number of the target module Range ASCII printab...

Page 25: ...3 Y Slot number of the target module Range ASCII printable characters 0 through 7 CMM Module type for the target Specifying CMM in this argument indicates the intention of any module type that may be...

Page 26: ...ugh 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 CMM Module type for the target Specifying CMM in this argument indicates the intention of any module type that may...

Page 27: ...Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 Range 3 ASCII printable characters O...

Page 28: ...the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 XDCR Commands the...

Page 29: ...ge ASCII printable characters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443...

Page 30: ...the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 ICPM Commands module to charge mode ZZ specifies the constant current as follows 00 turns the constan...

Page 31: ...cifying C02 addresses a 443B102 SETF Commands module to change the low pass filter setting Z specifies the low pass filter setting as follows 0 turns the low pass filter off 1 sets the low pass filter...

Page 32: ...rough 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 LOWF Commands module to change the low frequency response Z specifies the...

Page 33: ...the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 INTG Commands module to engage integration circuitry Z specifies the integrator as follows 1 engages...

Page 34: ...rget module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 INTU Commands module to s...

Page 35: ...able characters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifyin...

Page 36: ...able characters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifyin...

Page 37: ...e ASCII printable characters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C02 Module type for the target Only type C02 443B102 can accept this command si...

Page 38: ...rough 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C02 Module type for the target Only type C02 443B102 can accept this command since the 443B101 does not have lon...

Page 39: ...of the target module Range ASCII printable characters 0 through 7 C02 Module type for the target Only type C02 443B102 can accept this command since the 443B101 does not have long time constant NULL C...

Page 40: ...ASCII printable characters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C02 Module type for the target Only type C02 443B102 can accept this command sinc...

Page 41: ...module Range ASCII printable characters 0 through 7 C02 Module type for the target Only type C02 443B102 can accept this command since the 443B101 does not have long time constant OFFS Commands the mo...

Page 42: ...haracters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C02 Module type for the target Only type C02 443B102 can accept this command since the 443B101 doe...

Page 43: ...s Filter Integration Units Reference Overvoltage Fault ICP mode only Zero Lock long time constant charge mode only Description Mode CHRG Charge mode ICP 0mA ICP mode with 0mA constant current voltage...

Page 44: ...r set to 3 0 kHz 10 kHz Low pass filter set to 10 kHz 30 kHz Low pass filter set to 30 kHz 100 kHz Low pass filter set to 100 kHz Integration Units SI SI units will be used for input and output Eng En...

Page 45: ...follows ICP mode with a constant current of 2 mA output sensitivity is 10 00 mV unit transducer sensitivity is 1 023 mV unit low frequency response is 2 0 Hz low pass filter set to 10 kHz integration...

Page 46: ...Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 TEDR Commands the module to return da...

Page 47: ...printable characters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 spe...

Page 48: ...t module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 TEDD Commands the module to...

Page 49: ...aracters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 a...

Page 50: ...aracters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 a...

Page 51: ...3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102...

Page 52: ...3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102...

Page 53: ...rget module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 TEDW tells the module to...

Page 54: ...in the application register or application register scratchpad if unlocked of a TEDS device Response String 8 bytes of hex data in ASCII format Example Command String 06C02RDAR Response String AABBCC...

Page 55: ...xample Command String 06C02WRARAABBCCDDEEFFAABB Response String 0 The 443B102 in rack 0 and slot 6 is commanded to write 8 bytes of hex data 06C02WRARAABBCCDDEEFFAABB to the application register of a...

Page 56: ...et module Range ASCII printable characters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument ad...

Page 57: ...as been received Command String XYC0ZLKAR Description X Rack number of the target module Range ASCII printable characters 0 through 3 Y Slot number of the target module Range ASCII printable character...

Page 58: ...443B101 443B102 43 Manual 19854 Rev D ECN 50523 Appendix B Figures...

Page 59: ...443B101 443B102 44 Manual 19854 Rev D ECN 50523...

Page 60: ...443B101 443B102 45 Manual 19854 Rev D ECN 50523...

Page 61: ...443B101 443B102 46 Manual 19854 Rev D ECN 50523...

Page 62: ...443B101 443B102 47 Manual 19854 Rev D ECN 50523...

Page 63: ...443B101 443B102 48 Manual 19854 Rev D ECN 50523...

Page 64: ...dual mode amplifier for charge Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 F...

Page 65: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 66: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 67: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 68: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 69: ...not recommended and if attempted may void the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement a...

Page 70: ...provided by PCB Piezotronics Inc are covered by a limited warranty against defective material and workmanship for a period of one year Contact PCB for a complete statement of our warranty Expendable i...

Page 71: ...WER SUPPLY Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E ma...

Page 72: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 73: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 74: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 75: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 76: ...41A43 441A45 AND 441A49 2 EQUIPMENT RATINGS 2 MULTIPLE CHASSIS CONFIGURATION 2 CHANGING THE FUSE 2 DISABLING COOLING FAN 2 AVAILABLE MODULES 2 AC POWER SUPPLY MODEL 441A101 3 DC POWER SUPPLY MODEL 441...

Page 77: ...tion High voltage electrostatic discharge ESD can damage electrical devices To avoid triboelectric transfer 1 Connect the cables only with the AC power off 2 Temporarily short the end of the cable bef...

Page 78: ...n 3 0 Ah NiMH rechargeable battery The optional Model 488A09 is a 12VDC 3 4A AC Power Adaptor that allows the 441A102 to be used with any AC line power in the world 100 240VAC 50 60Hz The 441A102 can...

Page 79: ...CHASSIS AND POWER SUPPLY 4 Manual 19680 Rev C ECN 50523 Appendix A Figures...

Page 80: ...CHASSIS AND POWER SUPPLY 5 Manual 19680 Rev C ECN 50523...

Page 81: ...dual mode amplifier for charge Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 F...

Page 82: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 83: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 84: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 85: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 86: ...ION 4 BIAS OPERATION 4 OPERATION 4 SET UP AND MEASUREMENT 4 MAIN MENU 4 SETTING THE OPERATIONAL MODE OUT MENU 4 SETTING THE DVM MODE DVM MENU 4 SETTING THE INPUT MODE IN MENU 5 SETTING THE ICP CURRENT...

Page 87: ...l mode amplifier module 443B102 and an AC power supply module 441A101 into one integrated device Model 444A52 eliminates the dual mode amplifier module and is intended for direct connection to ICP pre...

Page 88: ...k DVM output is deactivated if in BIAS mode Analog Output The analog output BNC on the front panel allows the user to observe the input signal with a storage oscilloscope or other recording device Thi...

Page 89: ...observe the correct mounting and installation instructions Connect it to the input connector of the Model 444A151 If using an ICP sensor allow proper settling time rule of thumb is 5 times the sensor...

Page 90: ...the unit is DC coupled Note In ICP mode off zero current can be selected which will put the unit in voltage mode Setting the Gain GAIN menu From the main menu using the arrow buttons to navigate place...

Page 91: ...easurements down to OHz Enabling Timed Reset AUTO menu From the main menu using the arrow buttons to navigate place the cursor on AUTO and press ENTER The following menu should appear Using the arrow...

Page 92: ...led by pressing the RIGHT arrow the cursor will move to Pressing the RIGHT arrow again will move the cursor to Continually the cursor will make its way through every digit as the RIGHT arrow is presse...

Page 93: ...er will have MM in this field Command Contains the command followed by any associated data See below for command structure and command set for examples End Byte ETX 0x03 marks the end of the message D...

Page 94: ...ing error occurred ETX was received before it was expected I Check sum error occurred while communicating with internal module T Time out message not sent rack slot address may be incorrect NAK reason...

Page 95: ...441A152 PEAK METER 10 Manual 23528 Rev A ECN 50523 Appendix B Figures Figure 1 BLOCK DIAGRAM...

Page 96: ......

Page 97: ...ol for Series 440 modules Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 71...

Page 98: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 99: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 100: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 101: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 102: ......

Page 103: ...Conditioner Card Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 09...

Page 104: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 105: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 106: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 107: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 108: ...10 kHz 0 07 V Hz 143 dB 6 Broadband Electrical Noise 1 Hz to 10 kHz Charge Input 9 V 101 dB 7 8 Broadband Electrical Noise 2 Hz to 22 4 kHz Charge Input 5 fC 5 fC 7 8 Spectral Noise 1 kHz 0 8 V Hz 122...

Page 109: ...WER SUPPLY Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E ma...

Page 110: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 111: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 112: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 113: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 114: ......

Page 115: ...ol for Series 440 modules Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 71...

Page 116: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 117: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 118: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 119: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 120: ......

Page 121: ...Conditioner Card Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 09...

Page 122: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 123: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 124: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 125: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 126: ......

Page 127: ...dual mode amplifier for charge Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 F...

Page 128: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 129: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 130: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 131: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 132: ......

Page 133: ...dual mode amplifier for charge Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 F...

Page 134: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 135: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 136: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 137: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 138: ......

Page 139: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 140: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 141: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 142: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 143: ...LE W BNC PLUG DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 NJF 02 18 20 NJF 02 18 20 BAM 02 18 20 SCHEMATIC DIAGRAM SHELL CENTER PIN 72786 PCB Piezotronics Inc claims proprietary rights in the inform...

Reviews: