calculation of output scaling
The differential 422 series contain high-gain, low-noise amplifiers connected in a charge amplifier
configuration. The charge output of the transducer is transferred to the feedback capacitors of the
amplifiers to develop a voltage which may be calculated using the following equation.
V
out
= 422 Scaling (mV/pC) x Transducer Sensitivity (pC/Engineering Unit)
where:
Transducer sensitivity = positive sensitivity minus negative sensitivity
V
out
= Output of 422 in mV/Engineering Unit
the Engineering Unit may be g’s, Newtons, psi, etc., depending on the sensor
Example: A (3 pC/g) differential server is used with a 422M175 (6 mV/pC). Calculate the output in
mV/g.
Answer: 6 mV/pC x 3 pC/g = 18 mV/g
PAGE 4