background image

 

 

 

 
Figure 11.

 Typical Test Accelerometer Response 

 
7.3 PCB CALIBRATION PROCEDURE 

Numerous  precautions  are  taken  at  PCB  to  insure 
accurate and repeatable results.  This section provides a 
brief overview of the primary areas of concern. 
 
Since  the  Back-to-Back  Calibration  technique  relies  on 
each sensor experiencing an identical acceleration level, 
proper  mounting  of  the  test  sensor  to  the  reference 
standard is imperative.  Sensors with mounting holes are 
attached  directly  to  the  reference  standard  with  a  stud 
tightened  to  the  recommended  mounting  torque.  A 
shouldered  mounting  stud  is  typically  used  to  prevent 
the stud from “bottoming out” in the hole.  
 
Both  mounting  surfaces  are  precision-machined  and 
lapped  to  provide  a  smooth,  flat  interface  according  to 
the  manufacturer’s  specification.    A  thin  layer  of 
silicone grease is placed between the mating surfaces to 
fill  any  imperfections  and  increase  the  mounting 
stiffness.  The cables are stress-relieved by first routing 
them  to  the  shaker  head,  securing  them  with  tape  or 
cable  ties,  then  routing  them  to  a  nearby  stationary 
location.  This reduces cable motion, which is especially 
important  when  testing  charge output sensors and helps 
to  prevent  extraneous  noise  or  stresses  from  being 
imparted  into  the  system.    A  typical  set-up  is  shown  in 
Figure 12.  

 

 

 

Figure 12.

 Typical Calibration Set-Up 

 
Adhesively  mounted  sensors  use  similar  practices.  
However, in this case, a small portion of quick-bonding 
gel  or  similar  temporary  adhesive  is  used  to  attach  the 
test  sensor  to  a  reference  standard  designed  with  a 
smooth, flat mounting surface. 
 
In  addition  to  mounting,  the  selection  of  the  proper 
equipment  is  critical.    Some  of  the  more  important 
considerations  include:  1)  the  reference  standard  must 
be  specified  and  previously  calibrated  over  the 
frequency  and/or  amplitude  range  of  interest;  2)  the 
shaker should be selected to provide minimal transverse 
(lateral)  motion  and  minimal  distortion;  and  3)  the 
quality of the meters, signal generator, and other devices 
should  be  selected  so  as  to  operate  within  the  limits  of 
permissible error. 
 

7.4 COMMON MISTAKES 

Most  calibration  errors  are  caused  by  simply 
overlooking  some  of  the  fundamental  principals  of 
dynamics.  This section attempts to address some of the 
more common concerns. 
 
For stud-mount sensors, always mount the accelerometer 
directly  to  the  reference  standard.    Ensure  that  the 
mounting  surfaces  are  smooth,  flat,  and  free  of  any 
burrs.    Always  use  a  coupling  fluid,  such  as  silicone 
grease,  in  the  mounting  interface  to  maintain  a  high 
mounting  stiffness.    Mount  the  sensor  according  to  the 
manufacturer’s  recommended  mounting  torque.    DO 
NOT  use  any  intermediate  mounting  adaptors,  as  the 
mounted  resonant  frequency  may  be  reduced  and 
thereby compromise the high-frequency performance.  If 
necessary, use adaptor studs. 
 

Summary of Contents for 357A08/NC

Page 1: ...ge output accel 0 3 pC g 10k Hz Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...F 482 C and extreme application up to 100 F 650 C for UHT 12 These accelerometers are ideal for structural testing machine monitoring and vehicular shock and vibration measurement tasks where high tem...

Page 7: ...urface consider adhesive mounting as a possible alternative Inspect the area checking that there are no burrs or other foreign particles interfering with the contact surface STEP 2 Wipe clean the moun...

Page 8: ...or most applications It is generally used only for machinery monitoring and other portable trending applications Both the accuracy and repeatability at low 5 Hz and high frequency 1 kHz ranges are que...

Page 9: ...o a precise sensitivity such as 10 0 or 100 0 mV g for ease of data analysis This is accomplished with most PCB laboratory charge amplifiers and some miniature in line units as well For fixed sensitiv...

Page 10: ...dure themselves While this may result in both a savings of time and money it has also been attributed to incorrect readings and costly errors Therefore in an effort to prevent the common mistakes asso...

Page 11: ...Figure 12 Typical Calibration Set Up Adhesively mounted sensors use similar practices However in this case a small portion of quick bonding gel or similar temporary adhesive is used to attach the tes...

Page 12: ...s Incorrect Understand Back to Back Calibration limitations Do not expect the uncertainty of calibration to be any better than 2 In fact the uncertainty may be as high as 3 or 4 for frequencies 10 Hz...

Page 13: ...Jack 3 56 Coaxial Jack Electrical Connection Position Side Side 4 All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the...

Page 14: ......

Page 15: ......

Reviews: