background image

1

1

2

2

3

3

4

4

A

A

B

B

CODE

IDENT. NO.

52681

DWG. NO.

SCALE:

SHEET

DRAWN

CHECKED

ENGINEER

TITLE

UNLESS OTHERWISE SPECIFIED TOLERANCES ARE:

DIMENSIONS IN MILLIMETERS

[ IN BRACKETS ]

XX

    

±

 

0.13

ANGLES

  

 

2 DEGREES

3425 WALDEN AVE. DEPEW, NY 14043

(716)  684-0001 E-MAIL:  [email protected]

DIMENSIONS  IN INCHES

DECIMALS     XX

  

 

±.01

ANGLES

  

 

2 DEGREES

FILLETS AND RADII 

 .003 - .005 

DECIMALS     X

   

 ±

 

0.3

FILLETS AND RADII 

0.07 - 0.13 

INSTALLATION DRAWING

66373

2 OF 2

3X

TRIAXIAL MEMS SHOCK

ACCELEROMETER

XXX   ±.005

KRM

1/16/17

ECB

1/16/17

GCD

1/16/17

   

   

4

FASTEN CABLE TO TEST STRUCTURE TYPICALLY WITHIN 2-3[51-76] OF SENSOR. THEN FASTEN AGAIN WITHIN 3-4[76-101] OF
PREVIOUS ATTACHMENT. BETWEEN THE TEST STRUCTURE AND A FIXED STRUCTURE, ALLOW A SERVICE LOOP LARGE 
ENOUGH TO PREVENT PULLING OF THE CABLE WHEN SHAKING. MORE ATTACHMENT POINTS WILL PROVIDE LESS NOISE 
IN THE RESULTING DATA. LOOSE CABLES OR PARTS ELSEWHERE ON THE TEST STRUCTURE CAN ALSO GENERATE "NOISE" 

ON THE SIGNAL RECEIVED FROM THE SENSOR

REVISIONS

REV

DESCRIPTION

DIN

 -SEE 

SHEET 

ONE-

 

66

37

3

PCB Piezotronics Inc. claims proprietary rights in

the information disclosed hereon.  Neither it nor any

reproduction thereof will be disclosed to others
without the written consent of PCB Piezotronics Inc.

2 - 3[51-76]            

3 -4[76 - 102]           

4

4

"SERVICE LOOP"

TO ALLOW MOTION
WITHOUT PULLING
CABLE

TEST STRUCTURE

TAPE
OR EQUIVALENT

(TYPICAL)

INTEGRAL CABLE TO
PIGTAIL END

Summary of Contents for 3503A1120KG

Page 1: ...SHOCK ACCELEROMETER Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0001 Fax 716...

Page 2: ...ty regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme sh...

Page 3: ...rchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers shou...

Page 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Page 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Page 6: ...icro electromechanical systems MEMS technology in which all structural and electronic components are manufactured in silicon using electronic microfabrication technologies MEMS devices are preferred i...

Page 7: ...ck the mounting torque after each measurement It s also recommended that a thread locking adhesive be applied to any threads if applicable Ceramic LCC leadless chip carrier packages available in this...

Page 8: ...ipment for repair it is recommended that the user confer with a factory application engineer or inter national representative to first troubleshoot the problem 11 0 Return Procedure To expedite the re...

Page 9: ...Cable Type 8 Cond shielded silcone jacket 8 Cond shielded silcone jacket Cable Termination Pigtail Ends Pigtail Ends Cable Length 10 ft 3 05 m Mounting Through Holes 2 Through Holes 2 Mounting Torque...

Page 10: ...ny reproduction thereof will be disclosed to others without the written consent of PCB Piezotronics Inc RECOMMENDED MOUNTING SURFACE SHOULD BE FLAT TO WITHIN 0003 008 TIR OVER 650 16 51 WITH A 32 08 F...

Page 11: ...EN FASTEN AGAIN WITHIN 3 4 76 101 OF PREVIOUS ATTACHMENT BETWEEN THE TEST STRUCTURE AND A FIXED STRUCTURE ALLOW A SERVICE LOOP LARGE ENOUGH TO PREVENT PULLING OF THE CABLE WHEN SHAKING MORE ATTACHMENT...