background image

 

 

 

 

 

ICP

 FORCE SENSOR OPERATION MANUAL 

 

 6 

Figure 12 - Cable Strain Relief 

 
 
 
 

FORCE RING INSTALLATION 

The  sensor  is  mounted  using  the  supplied  mounting  stud  and 
pilot  bushing.    The  supplied  beryllium  copper  (BeCu)  stud  is 
elastic  so  it  allows  force  transmission  to  the  sensor  while 
holding the sensor in place.  The pilot bushing centers the sensor 
about the mounting stud.  After installing the mounting stud in 
the lower of the two surfaces, the pilot bushing is threaded over 
the mounting stud.  The sensor is then placed over the stud and 
pilot bushing combination.  The pilot bushing should fit loosely 
inside of the sensor inner diameter, holding it in place.  Properly 
machined  holes  for  the  mounting  stud  will  ensure  proper 
vertical orientation of the sensor.  The upper surface should be 
installed  and  tightened  onto  the  mounting  stud.    Refer  to  the 
installation drawing for additional mounting details.   
 
When  installing  the  sensor  as  an  integrated  member,  it  is 
recommended that the supplied antifriction washers be used to 
eliminate the possibility of damage to the sensing surface of the 
sensor.  This type of damage may occur when imperfections in 
the mounting surface grind against the sensor surface while the 
mounting surfaces are being twisted during installation.  Refer 
to the installation drawing for additional mounting details. 
 
 

PRELOAD REQUIREMENTS FOR FORCE  RING  AND 
3-COMPONENT FORCE SENSORS  
 

PCB ICP

 Force Rings (Models 201B01 through 207C) and 3-

Component Force Sensors (Models 260A01 through A03)  are 
generally  installed  between  two  parts  of  a  test  structure,  as 
shown in 

Figure 4

.  During installation, the  sensor should  be 

pre-loaded  to  the  amount  specified  on  the  specification  sheet 
using the supplied elastic beryllium-copper stud.   Preloading in 
this  arrangement  ensures  that  the  sensor  will  perform  as 
calibrated and have good output linearity at the sensor

s lower 

operating range.  Use of a mounting stud of stiffer material or 
no stud will alter PCB calibrated sensitivity. 
 
The stud or bolt holds the structure together and applies preload 
to  the  force  ring  as  shown  in 

Figure  13

.    In  the  typical  link 

installation,  shown  on  the  left  in 

Figure  13

,  part  of  the  force 

between  the  two  structures  is  shunted  through  the  mounting 
stud.    The  amount  of  force  shunted  is  dependent  on  the  stud 
material,  but  may  be  up  to  5%  of  the  total  force  for  the 
beryllium-copper stud supplied with the sensor and up to 50% 
for steel studs.  This typical installation setup is used by PCB 
during  standard  calibrations  of  force  ring  models  201B01 
through  207C.    PCB’s  standard  calibration,  when  using  the 
BeCu stud, takes this shunted force into account with the final 
calibration value. 
 

A non-typical installation is shown on the right side of 

Figure 

13

.  In this installation, the stud or bolt used to apply the preload 

does not shunt part of the applied force.  The plate on top of the 
sensor has a clearance hole that the stud or bolt passes through.  
In this installation, the stud or bolt is not directly connected to 
the top plate by its threads, as it is in the typical installation, so 
it  does  not  shunt  any  force.  This  method  of  installation 
accommodates  mounting a group of sensors under a common 
plate. 
 

NOTE:  If  any  of  the  following  conditions  apply  to  the 
preloading  of  the  force  ring  in  the  actual  application,  the 
sensitivity  and  linearity  performance  of  the  sensor  will  not 
match the standard PCB calibration values. 
 

1.

 

Use of a stud or bolt other than the supplied beryllium-
copper stud. 

2.

 

Use of no stud or bolt. 

3.

 

Use  of  an  amount  of  preload  other  than  the 
recommended amount. 

4.

 

Use  of  installation  that  is  different  from  PCB  setup 
during calibration. 

5.

 

On  rare  occasions,  a  ring  sensor  may  be  installed 
WITHOUT  a  mounting  stud  (as  installed  in  precise 
fixture).    In  these  cases,  the  sensor  sensitivity  will  be 
HIGHER  than  that  shown  on  the  PCB  calibration 
certificate. 

 
In these cases, please contact a PCB applications engineer at 
800-828-8840 of the intended installation to discuss your 
special calibration requirements.

  

PCB can calibrate the 

sensor without a stud OR ask that a customer supplied 
mounting stud be provided for use in PCB’s calibration 
process. 
 
 

 

Figure 13 - Force Ring Sensor Installations 

 

PCB in-house calibration procedure requires the installation of 
a  force  ring  with  BeCu  stud,  in  the  typical  installation  setup 
above.    This  sensor  is  then  placed  in  series  with  a  NIST 

Typical Link Installation 

Non-Typical Installation 

Used for Force Limited 

Vibration Testing 

Summary of Contents for 208A14

Page 1: ...ensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 6: ...stallation outline drawing and specification sheet at the back of this manual for details and dimensions of the particular sensor model number s purchased The following pages give a brief description of the various sensor series available recommended mounting procedures operation and recommended calibration In addition to PCB s line of ICP sensors each ICP sensor series outlined has corresponding ...

Page 7: ...e ring mounting and preload requirements Figure 4 outlines some possible mounting configurations to which the ring series may be installed Figure 4 Series 201 to 207 ICP Ring Force Sensor Possible Installation Methods GENERAL PURPOSE RADIAL Model 208C01 C05 General Purpose Sensors are designed to measure compression and impact forces from a fraction of a lb N to 5 000 lbs to 22 24 kN Tension force...

Page 8: ... simultaneously measuring force in three 3 orthogonal directions X Y and Z Internally three sets of quartz crystals are cut oriented and preloaded so the maximum output from the crystal cut for the specific plane is obtained when a force is applied axially to the specific x y or z plane Sensor output from loads applied in the alternate transverse planes considered channel cross talk is negligible ...

Page 9: ...returning the sensor to PCB for recalibration Figure 8 SERIES 3 COMPONENT TRIAX FORCE SENSORS PENETRATION Penetration style sensors are similar to the axial models but are specifically designed for compression and impact force measurements in materials testing applications Smooth cylindrical housings and curved impact caps avoid cutting through specimens This enables measurements to be taken to de...

Page 10: ...ot facing or surface grinding Surface flatness should be held to within 0 001 TIR over the entire mating surface When mounting sensors between two plates care should be taken to assure mounting surfaces are flat and parallel Non parallel surfaces could place unwanted stress on internal components leading to premature sensor failure The protective cap should remain on the electrical connector durin...

Page 11: ...the left in Figure 13 part of the force between the two structures is shunted through the mounting stud The amount of force shunted is dependent on the stud material but may be up to 5 of the total force for the beryllium copper stud supplied with the sensor and up to 50 for steel studs This typical installation setup is used by PCB during standard calibrations of force ring models 201B01 through ...

Page 12: ...ice the installer can monitor the sensor output voltage directly for correct preloading Monitor the output from the Z axis connector when preloading 3 component type sensors These sensors require a preload in the Z axis that is 10 times their shear range Some models require this to be accomplished in steps not to exceed the usable voltage To prevent clipping of the signal increments should not exc...

Page 13: ...meters depending on the frequency of interest It is necessary to supply the sensor with a 2 to 20 mA constant current at 20 to 30 VDC through a current regulating diode or equivalent circuit contained in all PCB signal conditioners See Guide G 0001B for powering and signal conditioning information pertaining to all ICP instrumentation Most of the line powered signal conditioners manufactured by PC...

Page 14: ...tant in the system For this reason the readout device as well as the signal conditioner must be DC coupled Figure 16 Characteristic Discharge Time Constant Curve The discharge time constant represents the decay rate of an input signal One DTC represents the amount of time taken for the signal to decay to 37 of the initial peak value As illustrated in Figure 16 this is an exponential decay Approxim...

Page 15: ...turned to the factory for servicing in the event of serious malfunction Observe the following precautions in using the sensor A Do not exceed the maximum load levels for the force sensor see specification sheet B Do not subject the sensor to temperatures exceeding that of the specification normally 250 F 121 C C Do not apply voltage to the sensor without current limiting diodes or other current pr...

Page 16: ...xy Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Electrical Connection Position Axial Axial All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc N Negative Output Polarity Output Polarity Compression Negative Negat...

Page 17: ......

Reviews: