background image

 

 

 

 

 

ICP

 FORCE SENSOR OPERATION MANUAL 

 

 1 

1.0   INTRODUCTION 

 
ICP

  force  sensors  incorporate  a  built-in  MOSFET 

microelectronic  amplifier.    This  serves  to  convert  the  high 
impedance charge output into a low impedance voltage signal 
for analysis or recording.  ICP

 sensors, are powered from  a 

separate  constant  current  source,  operate  over  long  ordinary 
coaxial  or  ribbon  cable  without  signal  degradation.    The  low 
impedance voltage signal is not affected by triboelectric cable 
noise or environmental contaminants.  
 

 

Figure 1 - ICP

 

Sensor System Schematic 

 
Power to operate ICP

 sensors is generally in the form of a low 

cost, 24-27 VDC, 2-20 mA constant current supply. 

Figure 1

 

schematically  illustrates  a  typical  ICP

  sensor  system.  PCB 

offers  a  number  of  AC  or  battery-powered,  single  or  multi-
channel  power/signal  conditioners,  with  or  without  gain 
capabilities for use with force sensors. In addition, many data 
acquisition systems now incorporate constant current power for 
directly powering ICP

 sensors.  Because static calibration or 

quasi-static short-term response lasting up to a few seconds is 
often  required,  PCB  manufactures  signal  conditioners  that 
provide DC coupling.

   Figure 2

 summarizes a complete 2-wire 

ICP

 system configuration. 

 

 

Figure 2 Typical ICP

 

Sensor System

 

 
In  addition  to  ease  of  operation,  ICP

  force  sensors  offer 

significant advantages over charge mode types. Because of the 
low  impedance  output  and  solid-state,  hermetic  construction, 
ICP

 force sensors are well suited for continuous, unattended 

force  monitoring  in  harsh  factory  environments.  Also,  ICP

 

sensor  cost-per-channel  is  substantially  lower,  since  they 

operate  through  standard,  low-cost  coaxial  cable,  and  do  not 
require expensive charge amplifiers. 
 

Refer to the installation/outline drawing and specification sheet 
at  the  back  of  this  manual  for  details  and  dimensions  of  the 
particular  sensor  model  number(s)  purchased

.  The  following 

pages  give  a  brief  description  of  the  various  sensor  series 
available,  recommended  mounting  procedures,  operation  and 
recommended calibration. 
 
 
In addition to PCB’s line of ICP®  sensors, each ICP® sensor 
series  outlined  has  corresponding  charge  output  versions.  
Charge  mode  versions  with high output impedance are  suited 
for  higher  temperature,  metal-to-metal  and  very  high  shock 
applications.  These  models  can  also  be  used  for  applications 
where it is desirable to manually set the output range.   
 
In addition to standard products, PCB has the ability to design 
and  manufacture  custom  sensors/systems  for  specific 
applications. 
  
If questions arise regarding the operation or characteristics of 
the force sensor products as outlined in this manual, feel free to 
contact an experienced PCB applications engineer  toll-free at 
1-800-828-8840. 
 

2.0   DESCRIPTION 
 
IMPACT 

Series  200  Impact  Sensors  are  designed  to  measure 
compression  and  impact  forces  from  a  fraction  of  a  lb(N)  to 
50,000 lbs (to 22.4 kN).  The flat sensing surface located on the 
top  of  the  sensor  is  designed  to  measure  a  dynamic  force 
quickly applied axially to the sensor.   
 
As highlighted in 

Figure 3

 compression forces directed against 

the sensing surface produce a positive-going output.  A tensile 
output  could  be  obtained  if  a  static,  steady-state  load  were 
applied  to  the  sensor.    The  maximum  tensile  output  to  be 
measured would be that of the applied static, steady state load 
as that load is quickly removed. 
 
 This  force-directed  input  and  corresponding  output  will 
provide  a  positive  going  output  signal  in  ICP

  models  and  a 

negative  going  output  in  charge  mode  sensors.    If  desired, 
adding the prefix 

N

” 

to a model number upon order, sometimes 

desired for charge mode models, will indicate polarity reversal. 

Summary of Contents for 208A11

Page 1: ... comp 10 lb tension 500 mV lb Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 6: ...stallation outline drawing and specification sheet at the back of this manual for details and dimensions of the particular sensor model number s purchased The following pages give a brief description of the various sensor series available recommended mounting procedures operation and recommended calibration In addition to PCB s line of ICP sensors each ICP sensor series outlined has corresponding ...

Page 7: ...e ring mounting and preload requirements Figure 4 outlines some possible mounting configurations to which the ring series may be installed Figure 4 Series 201 to 207 ICP Ring Force Sensor Possible Installation Methods GENERAL PURPOSE RADIAL Model 208C01 C05 General Purpose Sensors are designed to measure compression and impact forces from a fraction of a lb N to 5 000 lbs to 22 24 kN Tension force...

Page 8: ... simultaneously measuring force in three 3 orthogonal directions X Y and Z Internally three sets of quartz crystals are cut oriented and preloaded so the maximum output from the crystal cut for the specific plane is obtained when a force is applied axially to the specific x y or z plane Sensor output from loads applied in the alternate transverse planes considered channel cross talk is negligible ...

Page 9: ...returning the sensor to PCB for recalibration Figure 8 SERIES 3 COMPONENT TRIAX FORCE SENSORS PENETRATION Penetration style sensors are similar to the axial models but are specifically designed for compression and impact force measurements in materials testing applications Smooth cylindrical housings and curved impact caps avoid cutting through specimens This enables measurements to be taken to de...

Page 10: ...ot facing or surface grinding Surface flatness should be held to within 0 001 TIR over the entire mating surface When mounting sensors between two plates care should be taken to assure mounting surfaces are flat and parallel Non parallel surfaces could place unwanted stress on internal components leading to premature sensor failure The protective cap should remain on the electrical connector durin...

Page 11: ...the left in Figure 13 part of the force between the two structures is shunted through the mounting stud The amount of force shunted is dependent on the stud material but may be up to 5 of the total force for the beryllium copper stud supplied with the sensor and up to 50 for steel studs This typical installation setup is used by PCB during standard calibrations of force ring models 201B01 through ...

Page 12: ...ice the installer can monitor the sensor output voltage directly for correct preloading Monitor the output from the Z axis connector when preloading 3 component type sensors These sensors require a preload in the Z axis that is 10 times their shear range Some models require this to be accomplished in steps not to exceed the usable voltage To prevent clipping of the signal increments should not exc...

Page 13: ...meters depending on the frequency of interest It is necessary to supply the sensor with a 2 to 20 mA constant current at 20 to 30 VDC through a current regulating diode or equivalent circuit contained in all PCB signal conditioners See Guide G 0001B for powering and signal conditioning information pertaining to all ICP instrumentation Most of the line powered signal conditioners manufactured by PC...

Page 14: ...tant in the system For this reason the readout device as well as the signal conditioner must be DC coupled Figure 16 Characteristic Discharge Time Constant Curve The discharge time constant represents the decay rate of an input signal One DTC represents the amount of time taken for the signal to decay to 37 of the initial peak value As illustrated in Figure 16 this is an exponential decay Approxim...

Page 15: ...turned to the factory for servicing in the event of serious malfunction Observe the following precautions in using the sensor A Do not exceed the maximum load levels for the force sensor see specification sheet B Do not subject the sensor to temperatures exceeding that of the specification normally 250 F 121 C C Do not apply voltage to the sensor without current limiting diodes or other current pr...

Page 16: ...ck 10 32 Coaxial Jack Electrical Connection Position Axial Axial All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc N Negative Output Polarity Output Polarity Compression Negative Negative W Water Resistant Cable Electrical C...

Page 17: ......

Reviews: