PCB Piezotronics 137B22B Installation And Operating Manual Download Page 8

Model Number

137B22B

ICP® PRESSURE SENSOR

Revision: B
ECN #: 51260

P erfo rm an ce

EN G L ISH

SI

Measurement Range(for ±5V output)

500 psi

3,447 kPa

Measurement Range

(for ±10V output)

1 kpsi

6,895 kpsi

[1]

Sensitivity(± 15 %)

10 mV/psi

1.45 mV/kPa

Maximum Pressure

5 kpsi

34,474 kPa

Resolution

1 mpsi

.007 kPa

[2]

Resonant Frequency

≥ 400 kHz

≥ 400 kHz

Rise Time(Incident)

≤ 6.5 µ sec

≤ 6.5 µ sec

Non-Linearity

≤ 1.0 % FS

≤ 1.0 % FS

[3]

En viro n m en tal

Temperature Range(Operating)

-100 to +275 °F

-73 to +135 °C

Temperature Coefficient of Sensitivity

≤ 0.05 %/°F

≤ 0.090 %/°C

Electrical

Discharge Time Constant(at room temp)

≥ 0.2 sec

≥ 0.2 sec

Excitation Voltage

20 to 30 VDC

20 to 30 VDC

Constant Current Excitation

2 to 20 mA

2 to 20 mA

Output Impedance

≤ 100 Ohm

≤ 100 Ohm

Output Bias Voltage

8 to 14 VDC

8 to 14 VDC

P h ysical

Sensing Geometry

Compression

Compression

Sensing Element

Quartz

Quartz

Housing Material

Aluminum

Aluminum

Diaphragm

Invar

Invar

Sealing

Epoxy

Epoxy

Electrical Connector

BNC Jack

BNC Jack

Weight

12.4 oz

352 gm

[2]

[4]

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

ICP

®

 is a registered trademark of PCB Piezotronics, Inc.

OP TION A L  VERSION S

Optional versions have identical specifications and accessories as listed for the standard model except

where noted below. More than one option may be used.

N OTES:

[1]For +10 volt output, minimum 26 VDC supply voltage required. Negative 10 volt output may be
limited by output bias.
[2]Typical.
[3]Zero-based, least-squares, straight line method.
[4]See PCB Declaration of Conformance PS023 for details.

Date: 11/13/2020

Date: 11/13/2020

Date: 11/13/2020

Date: 11/13/2020

55318

Entered: LK

Engineer: RPF

Sales: RWM

Approved: RPF

Spec Number:

3425 Walden Avenue, Depew, NY 14043

Phone: 716-684-0001
Fax: 716-684-0987
E-Mail: [email protected]

SU P P L IED A C C ESSORIES:

Model PCS-1 Calibration of pressure sensors up to 15k psi (except Series 113B50)

Summary of Contents for 137B22B

Page 1: ...Sensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail i...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...any case care must be taken such that the sensor s diaphragm is not oriented down to avoid ground reflections In some cases where flash temperatures such as those generated by blasts and shock fronts...

Page 7: ...hundreds or thousands of feet use a higher current up to 20 mA maximum Use of low capacitance cable i e RG 62 U and impedance matching variable resistors are suggested to improve long cable performanc...

Page 8: ...Diaphragm Invar Invar Sealing Epoxy Epoxy Electrical Connector BNC Jack BNC Jack Weight 12 4 oz 352 gm 2 4 All specifications are at room temperature unless otherwise specified In the interest of con...

Page 9: ......