PCB Piezotronics 132 Series Installation And Operating Manual Download Page 7

 

ICP

 is a registered trademark of PCB Piezotronics, Inc. 

 

 

 

Drawing Number: 21094 
Revision: A 
ECO #: 35764 

 

OPERATION MANUAL FOR MICROSENSORS 

132

 

Series 

 

 2 

 

 

Model 132A35 

 
 
 

 

 

 

Model 132A36 

 
 

 

 

 

Model 132A37 

 
 
 

3.0 

INSTALLATION 

 
Refer  to  the  installation  drawing  for  details.    If  the 
microsensors are to be subjected to pressures beyond 
their  specified  range,  be  sure  to  use  backing  nuts  to 
prevent  the  sensors  from  backing  out  of  their 
mounting  ports.  (Thread  adaptors  are  available  from 
PCB to facilitate mounting.) 

Connect the microsensors to an ICP

 power supply. 

Use standard two-conductor cable between the sensor 
and the ICP

 power unit. 

 
Next, connect the power unit to a readout device, e.g., 
oscilloscope,  recorder  or  high-speed  A  to  D 
converter. 
 

4.0 

OPERATION 

 
Switch the ICP

 

power unit on and observe reading of 

bias-monitoring  voltmeter  on  front  panel  of  power 
unit. 
 
If indicator is in green section of indicator panel, the 
IC  amplifier  is  providing  proper  bias  (normally  +11 
VDC),  cable  connections  are  normal  and  the  system 
is ready to operate. 
 
If  the  needle  moves  to  the  red  area  of  the  fault 
monitor meter, output is zero and a short is indicated.  
The  short  could  be  located  in  the  amplifier  cable 
connectors or in the power unit. 
 
If the pointer moves into the yellow area of the fault 
monitor  meter,  an  open  circuit  is  indicated  with  full 
power  supply  voltage.    An  open  circuit  could  be  the 
result  of  a  faulty  amplifier,  an  open  cable  or  open 
connectors. 
 
Allow the sensor to thermally stabilize for about one 
minute.    A  signal  drift  may  occur  when  the  output 
cable is connected to the readout instrument. 
 
The  drift  occurs  during  charging  of  the  coupling 
capacitor in the power unit.  The signal will stabilize 
in several minutes. 
 
 
 

Model 060A28 

Clamp Nut 

¼” Hex 

¼-28 Thread 

(supplied) 

2-ft Twisted Pair 
Cable Terminating into 
Model 070B09 Solder 
Connector Adaptor 
(supplied) 

2-ft Twisted Pair 
Cable Terminating 
into Pigtail 

#10-32 

Coaxial 

Connector  

O-Rings 
(2 supplied)  

O-Rings 
(2 supplied)  

Summary of Contents for 132 Series

Page 1: ...sor Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...rs is not attenuated nor is noise appreciably increased by the cable 2 0 DESCRIPTION 132 Series Microsensors feature small diameter pressure sensing surfaces incorporating very small 1 mm piezoelectri...

Page 7: ...n of indicator panel the IC amplifier is providing proper bias normally 11 VDC cable connections are normal and the system is ready to operate If the needle moves to the red area of the fault monitor...

Page 8: ...ook flat topped on the readout after about 10 volts The initial rise of the output will be accurate up to the point of clipping This can allow for an accurate time of arrival measurement even though t...

Page 9: ...ainless Steel Sealing Epoxy Epoxy Weight 0 45 oz 12 77 gm 5 Cable Termination Pigtail Pigtail Cable Type 030 Coaxial 030 Coaxial All specifications are at room temperature unless otherwise specified I...

Page 10: ...rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Piezotronics Inc REVISIONS REV DESCRIPTION DIN C REVIS...

Reviews: