background image

 

- 4 - 

 

Figure 4 

 
Microphone calibration provides, with a definable degree of accuracy, the necessary link 
between the physical quantity being measured and the electrical signal generated by the 
microphone. In addition, other useful information concerning operational limits, physical 
parameters, electrical characteristics or environmental influences may be determined. Under 
normal conditions, microphones are very stable. However, the microphone may be temporarily 
or permanently affected by harsh influences, such as moisture, dirt, being accidentally dropped 
or other unusual conditions that may cause the microphone accuracy to deviate from the normal 
specifications. This may manifest itself in a number of ways, ranging from a loss in frequency 
range, to total failure of the built-in microelectronic circuit due to high mechanical shock. 
 

~ Calibration Certificate ~

Model Number:

130E21

Serial Number:

37649

Description:

ICP® Microphone

Manufacturer:

PCB

Method 

1

 : Back-to-Back Reference

Calibration Data

Sensitivity:

48.1

mV/Pa

Reference Freq.:

250

Hz.

-26.4

dB re 1V/Pa

Output Bias:

8.6

VDC

Temperature:

71

°F

22

°C

Relative Humidity:

50

%

Condition of Unit

As Found: n/a
As Left:

New Unit, In Tolerance

Notes

1. This method involves comparing the test unit with a reference microphone in a GRAS 51AB calibrator.

2. Calibration is PTB Traceable and calibration records are on file.

3. Calibration is performed in compliance with ISO 9001, ISO 10012-1 and ANSI/NCSL Z540.3.

4. Due to state-of-the-art limitations, 4:1 calibration ratios are not possible on pressure measurement standards, microphones

and acoustic calibrators.  Calibration ratios for these types of devices are limited to 1:1.

5. See M anufacturer’s Specification Sheet for a detailed listing of performance specifications.

6. This certificate shall not be reproduced, except in full, without written approval from PCB Piezotronics, Inc

7. Calibrated per ACS-21.

Technician:

Date:

V

Headquarters: 3425 Walden Avenue, Depew, NY 14043

Calibration performed at: 10869 Highway 903, Halifax, NC 27839

Page 1 of  1

T EL: 888-684-0013

FAX: 716-685-3886

www.pcb.com

CAL58-3465911589.795

October 29, 2013

Scott Skibniewski

-10

-5

0

5

10

10

100

1000

10000

100000

R

e

sp

o

n

se

 D

e

v

ia

ti

o

n

 (

d

B

)

Frequency (Log Hz.)

Response Plot

Summary of Contents for 130E20

Page 1: ...Microphone Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E ma...

Page 2: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Page 3: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Page 4: ...Calibrators etc you will ensure years of trouble free usage If after reading this manual you have additional questions concerning the Microphone or its application feel free to call a factory Applicat...

Page 5: ...ll ICP powered microphones require correct constant current excitation for proper operation For this reason use only PCB constant current signal conditioners or other approved constant current sources...

Page 6: ...e object you are measuring The Free Field type of microphone is best suited for applications where the sound is coming from one defined direction without any disturbing objects or hard surfaces that w...

Page 7: ...mV Pa Reference Freq 250 Hz 26 4 dB re 1V Pa Output Bias 8 6 VDC Temperature 71 F 22 C Relative Humidity 50 Condition of Unit As Found n a As Left New Unit In Tolerance Notes 1 This method involves co...

Page 8: ...ic instrumentation needs with a PCB Application Engineer 7 0 Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material an...

Page 9: ...eel Stainless Steel Electrical Connector Output BNC Jack BNC Jack Size Diameter x Length overall 0 5 in x 2 62 in 12 7 mm x 66 6 mm Size Diameter x Length head 0 28 in x 1 10 in 7 mm x 28 mm Weight 0...

Page 10: ......

Reviews: