background image

 

- 4 - 

The  130A24  microphone  design  includes  a  removable  grid  cap 
and replaceable protective cover.  This unique feature allows the 
130A24  to  withstand  harsh  environments  while  extending  the 
lifespan  of  the  microphones’  critical  sensitive  components. 
Should  the  water  and  dust  resistant  cover  get  damaged  due  to 
prolonged  exposure  to  harsh  conditions,  a  removable  grid  cap 
allows the user to replace this protective cover quickly and easily 
in a few simple steps. First, unthread the grid cap from the housing. Next, peel off the existing 
cover from the end of the housing. Sharp edged tools can be used to aid in cover removal, but it 
is important to keep the tool at an angle that is parallel to the black covers surface area so that no 
contact  with  the  microphone’s  delicate  diaphragm  underneath  occurs  (Figure  4).  To  ensure  a 
proper  seal,  gently  wipe  the  top  of  the  microphone  housing  to  remove  any  moisture  or  debris.  
Apply  the  replacement  water-resistant  patch  to  the  end  of  the  housing,  keeping  it  centered. 
Finally, assemble the grid cap back onto the housing.  Optional 079A46 five pack replacement 
water resistant pads can be ordered from PCB. 
 

5.0 Calibration 

All PCB microphones come with a certificate of calibration and compliance with ISO 9001 and 
ANSI/NCSL Z540.3. In order to keep the array microphone cost effective, the 130 series 
calibrations are not A2LA accredited. For A2LA accredited microphones, please select our 377 
and 378 series of acoustic products. PCB microphones are traceable to one or more of the 
following National Labs (NIST, PTB or DFM), and calibration records are on file. PCB utilizes 
the “Back-to-Back Reference” form of calibration for the 130 series of array microphones. The 
microphone is checked against a test and measurement working standard microphone and must 
meet specifications before it can be shipped. An example of a calibration certificate is shown on 
the following page (Figure 5)
 

Figure 4

 

Summary of Contents for 130A24

Page 1: ...Microphone Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E ma...

Page 2: ...ty regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme sh...

Page 3: ...rchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers shou...

Page 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Page 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Page 6: ...brators etc you will ensure years of trouble free usage If after reading this manual you have additional questions concerning the Microphone or its application feel free to call a factory Application...

Page 7: ...They utilize low impedance cable on which signals can be driven long distances with minimal signal loss 3 0 Powering All ICP powered microphones require correct constant current excitation for proper...

Page 8: ...adout device This is a standard Coaxial cable with BNC connectors on each end PCB offers these cables in a variety of lengths for any application Figure 3 All of the 130 series Array Microphones are t...

Page 9: ...icrophone housing to remove any moisture or debris Apply the replacement water resistant patch to the end of the housing keeping it centered Finally assemble the grid cap back onto the housing Optiona...

Page 10: ...mV Pa Reference Freq 250 Hz 26 4 dB re 1V Pa Output Bias 8 6 VDC Temperature 71 F 22 C Relative Humidity 50 Condition of Unit As Found n a As Left New Unit In Tolerance Notes 1 This method involves co...

Page 11: ...ic instrumentation needs with a PCB Application Engineer 7 0 Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material an...

Page 12: ...dance 52 Ohm 52 Ohm Physical Housing Material Stainless Steel Stainless Steel Electrical Connector Output BNC Jack BNC Jack Size Diameter x Length overall 52 in x 2 75 in 13 2 mm x 69 9 mm Weight 1 8...

Page 13: ...FILLETS AND RADII 0 07 0 13 OUTLINE DRAWING 59447 1 OF 1 2X MODEL 130A24 DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 BB 5 4 15 ECB 5 4 15 DRK 5 4 15 bbehringer 12 46 5 4 2015 59447 PCB Piezotronics...

Reviews: