background image

Model Number

1220-11ADB

LOW PROFILE LOAD CELL, DUAL BRIDGE

Revision: NR

ECN #: 42842

Performance

ENGLISH

SI

Measurement Range(100 x 10^6 cycles)

25 klb

111 kN

[7][8]

Sensitivity(± 10 %)

4.0 mV/V

4.0 mV/V

[9][1]

Non-Linearity

≤ 0.05 % FS

≤ 0.05 % FS

Hysteresis

≤ 0.05 % FS

≤ 0.05 % FS

Non-Repeatability

≤ 0.02 %RO

≤ 0.02 %RO

Eccentric Load Sensitivity

0.1 %/in

0.1 %/25.4mm

Resonant Frequency

15 kHz

15 kHz

Side Load Sensitivity

0.1 %/in

0.1 %/25.4mm

Creep(in 20 minutes)

0.025 %

0.025 %

Static Error Band

≤ 0.05 % FS

≤ 0.05 % FS

[7]

Environmental

Overload Limit

37.5 klb

167 kN

Load Limit(Side Force, F

X

 or F

Y

)

12.5 klb

56 kN

[3]

Load Limit

(Bending Moment, M

X

 or M

Y

)

12.5k in-lb

1413 Nm

[3]

Load Limit

(Axial Torque, M

Z

)

12.5k in-lb

1413 Nm

[3]

Temperature Range(Operating)

-65 to +200 °F

-54 to +93 °C

Temperature Range

(Compensated)

+15 to +115 °F

-9 to +46 °C

Temperature Effect on Output(Maximum)

± 0.0008 %Reading/°F

± 0.0018 %Reading/°C

[4]

Temperature Effect on Zero Balance(Maximum)

± 0.0008 %FS/°F

± 0.0018 %FS/°C

[4]

Moisture Protection

See note

See note

[5]

Rust Resistance

See note

See note

[6]

Electrical

Bridge Resistance

350 Ohm

350 Ohm

[1]

Excitation Voltage(Recommended)

10 VDC

10 VDC

[2]

Insulation Resistance

>5k GOhm

>5k GOhm

Zero Balance

≤ 1 %RO

≤ 1 %RO

Physical

Size (Diameter x Height)

6.06 in x 3.50 in

153.90 mm x 88.90 mm

[10]

Weight

21 lb

9.53 kg

Mounting Thread

1 1/4 - 12 UNF-3B

No Metric Equivalent

Sensing Element

Strain Gage

Strain Gage

Mounting Torque

1400 in-lb

158 Nm

Deflection at Full Scale Capacity

0.001 in

0.03 mm

Electrical Connector

PT02E-10-6P (2)

PT02E-10-6P (2)

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

ICP

®

is a registered trademark of PCB Group, Inc. 

OPTIONAL VERSIONS

Optional versions have identical specifications and accessories as listed for the standard model 

except where noted below. More than one option may be used.

NOTES:

[1]

Nominal.

[2]

Calibrated at 10 VDC, usable 5 to 20 VDC or VAC RMS.

[3]

Singularly applied, i.e. no other extraneous loads.

[4]

Over compensated operating temperature range.

[5]

Gages, resistors and wiring coated against moisture, double diaphragms top and bottom, 
rubber gaskets under connectors.

[6]

All none load surfaces painted, load surfaces and mounting thread coated with dry film rust 
inhibitor.

[7]

FS - Full Scale.

[8]

Calibrated in tension and compression.

[9]

RO - Rated Output.

[10] See Outline Drawing 58695 for Complete Dimensions.

OPTIONAL ACCESSORIES: 

Model 084A56 Pre-Tension Stud
Model 084A90 Connector Shroud (2)
Model 181-012A PT06A-10-6S(SR)
Model 8311-01-10A CABLE

Entered: AP

Engineer: PE

Sales: KWW

Approved: JSD

Spec Number:

Date: 7/11/2014

Date: 7/11/2014

Date: 7/11/2014

Date: 7/11/2014

58781 

PCB Load & Torque, Inc.
24350 Indoplex Circle
Farmington Hills, MI 48335
UNITED STATES
Phone: 866-684-7107
Fax: 716-684-0987
E-Mail: [email protected]
Web site: http://www.pcbloadtorque.com

Summary of Contents for 1220-11ADB

Page 1: ...LL DUAL BRIDGE Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987...

Page 2: ...repair User servicing or repair is not recommended and if attempted may void the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps...

Page 3: ...ry customer service representative Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material and workmanship for a period...

Page 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Page 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Page 6: ...in Gage Strain Gage Mounting Torque 1400 in lb 158 Nm Deflection at Full Scale Capacity 0 001 in 0 03 mm Electrical Connector PT02E 10 6P 2 PT02E 10 6P 2 All specifications are at room temperature unl...

Page 7: ...13 58695 PCB Load Torque Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Load...

Page 8: ...LECTRICAL INSTALLATION 5 5 1 Electrical Drawing Western Regional Std 5 5 2 Cable Grounding Considerations 5 6 0 CALIBRATION 5 6 1 Calibration Certificate Description 5 6 1 1 Measured Output 6 6 1 2 Hy...

Page 9: ...the parts under test have completed the planned fatigue test cycles 3 0 OVERVIEW Dual bridge fatigue rated load cells are specifically designed for repetitive testing Applications include material te...

Page 10: ...ote that the performance of the load cell may be affected if the factory installed mounting base is removed or modified 4 2 Mounting Load Cell to a Standard Base or Custom Fixture If the load cell is...

Page 11: ...ctor Protector 1 Remove the 4 40 x 1 4 button head cap screws from the connector using a 1 16 hex wrench as shown in Figure 7 2 Carefully pull the connector from the load cell Take care to not break o...

Page 12: ...wiring may result in unstable or noisy signals A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structur...

Page 13: ...s are adjusted such that the error due to eccentric loading is less than 0 1 per inch of offset Side loads moments and torque must be limited to avoid electrical or mechanical damage to the load cell...

Page 14: ...tion certificate and adjust the span or gain until the display reads the force value stated on the certificate 4 Repeat steps 1 3 to verify that a valid calibration setting has been obtained 5 If poss...

Page 15: ...ridge resistances A D excitation and B C the signal leads 9 Check bridge balance 10 Keep a record of your observations correct problems or contact PCB factory for assistance 11 2 1 Estimating Bridge B...

Page 16: ...factory for any reason For calibration services if possible a copy of the purchase order covering the requested services should be included with the returned load cell 13 0 WARRANTY Standard warranty...

Reviews: