Drawing Number: 21086 
Revision: NR

 

 

 

 

OPERATION MANUAL FOR 

CONFORMAL BALLISTICS PRESSURE SENSOR 

Series 117B

 

the  curvature.  In  these  latter  cases,  the 
sensor  forward  side  is  identified  with  the 
legend 

FWD

 etched on the guide collar to 

which  the  slotted  clamp  is  attached.  The 
clamp  must  be  removed  to  find  this  legend. 
If  it  does  not  appear,  the  sensor  may  be 
mounted with either side toward the muzzle.  

 

2. 

Loosen  slotted  clamp,  but  do  not  remove 
clamp.  

 
3. 

Select  the  middle  thickness  (.014)  spacer 
from  the  set  of  nine  (065A19)  supplied  and 
place it around sensor barrel. 

 
4.  

Begin  threading  the  sensor  clamp  nut  into 
the  threaded  mounting  port,  sliding  slotted 
clamp fore and aft as needed to allow guide 
pin to fully enter hole.  

  

Continue to turn clamp nut into hole by hand 
or using 5/16 open end wrench. 
 
Do not tighten when sensor bottoms. 
 

5. 

Now tighten the screw closing slotted clamp. 

 
6. 

Using  open  end  wrench,  tighten  sensor 
clamp nut. 

 
NOTE:  It  is  not  necessary  to  put  large  amounts  of 

torque on this nut since a pressure seal is not 
necessary.  Approximately  5  to  10  ft.  lbs.  is 
sufficient.  

 
7.  

Now  inspect  flushness  of  diaphragm  with 
inside  surface  of  chamber.  This  can  be 
accomplished visually in most cases. 

 

If the diaphragm extends too deeply into the 
chamber,  select  a  thicker  spacer  and  repeat 
mounting procedure.  

 

If  the  diaphragm  is  too  deeply  recessed, 
select a thinner spacer and remount. 

Once  the  proper  thickness  is  found  for 
perfect  flushness,  the  sensor  may  be 
removed  and  reinstalled  using  this  same 
spacer and the proper depth will be achieved 
each time.  

 

NOTE:  For  best  accuracy  of  results,  use 
same charge amplifier for calibration and for 
actual 

operation. 

Use 

long 

TC 

for 

calibration,  medium  or  short  TC  for  best 
drift free operation. 

 

4.0  

POLARITY 

 

Polarity  of  the  Model  117B  is  negative  i.e.,  the 
charge  output  is  negative  for  increasing  pressure 
input,  making  it  compatible  with  inverting  type 
charge amplifiers. 
 

5.0 

CALIBRATION 

 

Calibration  of  the  Models  117B  is  facilitated  by  a 
calibration  adaptor  which  exactly  matches  the 
chamber  dimensions  of  the  cartridge  under  test.  An 
actual cartridge case is hydraulically pressurized with 
reference  pressure  to  obtain  a  point-by-point  sensor 
calibration. 
These  calibration  adaptors  can  be  fabricated  by  the 
user  or  can  be  purchased  from  PCB  as  our  Model 
090B calibration adaptor. Simply specify caliber and 
longitudinal location of sensor. 
 
For  most  rimfire  applications,  the  sensor  is  located 
.25 inches forward of the boltface. 
 
For center  fire cartridges a location .175 inches back 
from the cartridge shoulder is preferred. Consult PCB 
for recommendations on sensor locations if questions 
arise. 
 
Since most cartridge cases are tapered and diameter is 
dependent  upon  longitudinal  location,  this  location 
may  not  be  changed  after  the  sensor  is  fabricated 
since  it  is  essential  for  proper  operation  that  the 
sensor curvature exactly match chamber curvature.  

Summary of Contents for 117B66

Page 1: ...RESSURE SENSOR Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987...

Page 2: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Page 3: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Page 4: ...gement allows for tolerance in the location of the guide pin hole in an axial direction on the test barrel or calibration adaptor The precise depth adjustment is obtained by the use of the correct thi...

Page 5: ...ct a thinner spacer and remount Once the proper thickness is found for perfect flushness the sensor may be removed and reinstalled using this same spacer and the proper depth will be achieved each tim...

Page 6: ...to allow easy insertion and extraction the case must be expanded slightly by the internal pressure before force can be transmitted to the sensor See the enclosed guide An Improved Technique for Utili...