AN IMPROVED TECHNIQUE FOR  

UTILIZATION OF CONFORMAL BALLISTICS  

SENSOR CALIBRATION DATA

 

 4 

The straight line is the best line that can be constructed 
through  the  linear  portion  of  the  curve  as  shown  in 
Figure  4  and  the  pressure  intercept  (or  offset)  is  the 
point at which this line crosses the pressure axis, in psi. 
 
Instead  of  establishing  the  sensitivity  parameter  (which 
has  been  shown  to  have  little  meaning  for  this  sensor) 
the user now determines the slope as defined in Eq. 6  
 
and  the  accompanying  pressure  offset  by  constructing 
the best straight line as described previously. 
 
Pressures  P1  and  P2  are  arbitrarily  established  at  some 
point  well  above  and  well  below  the  expected  nominal 
pressure of the test. 
 
The  corresponding  outputs  at  these  pressures  should be 
an average of at least 20 cartridges taken from the same 
lot of ammunition as is under test. 
 
The main difference between utilization of the data with 
this method is that now a constant (pressure offset) must 
be  added  to  each  reading.    However,  this  can  be  easily 
handled by most data reduction methods in use today. 
 
Example: 
 
In actual practice, this is how a conformal sensor/charge 
amplifier system could be utilized: 
 
Assume  that  averaging  the  data  from  20  consecutive 
calibration  runs  performed  on  a  conformal  sensor 
yielded  a  slope  of  .04  pC/psi  and  an  intercept  of  7,500 
psi,  determined  by  constructing  the  best  fitting  straight 
line as described previously. 
 
Assume  also  that  average  expected  peak  pressure  is 
25,000 psi. 
 
Approximate full scale charge output is: 
 

pC

1000

pC/psi

.04

psi

25,000

 

Set range of charge amplifier output to 1000 pC/volt (or 
other desired range). 
 
Read peak output voltage of .800 volts during test. 
 
Peak indicated charge is then: 
 
1000 pC/volt x .80 volts = 800 pC  
 
Indicated peak pressure is: 
 

offset

slope

charge

peak

(pk)

Pressure

 

psi

7,500

pC/psi

.04

pC

800

P

 

 
P = 20,000 psi + 7500 = 27,500 Psi 
 
As  previously  stated,  this  method  will  yield  accurate 
results anywhere  within  the linear portion of  the sensor 
calibration  curve  without  inherent  errors  due  to 
variability of sensitivity with pressure level. 
 
Accuracy  should  be  close  to  that  expected  for  standard 
piezoelectric 

sensors 

within 

the 

limitation 

of 

transmissibility variations in the cartridge cases. 
 
Simplified  standardization  techniques  for  Conformal 
Sensor 
 
The  technique  described  by  the  previous  example 
outlines  one  basic  method  of  output  signal  utilization.  
This  method  in  actual  practice  may  be  cumbersome for 
repeated testing because of the conversions that must be 
made  from  measured  charge  output  to  indicated  peak 
pressure. 
 
Other  system  scaling  methods  exist  which  further 
simplify signal utilization. 

Summary of Contents for 117B205

Page 1: ...PRESSURE SENSOR Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 6: ...ngement allows for tolerance in the location of the guide pin hole in an axial direction on the test barrel or calibration adaptor The precise depth adjustment is obtained by the use of the correct thickness spacer selected from a set of 9 spacers of various thicknesses supplied with each sensor Once the proper thickness spacer is found removal and re installation now becomes a routine matter 3 0 ...

Page 7: ...ect a thinner spacer and remount Once the proper thickness is found for perfect flushness the sensor may be removed and reinstalled using this same spacer and the proper depth will be achieved each time NOTE For best accuracy of results use same charge amplifier for calibration and for actual operation Use long TC for calibration medium or short TC for best drift free operation 4 0 POLARITY Polari...

Page 8: ...r to allow easy insertion and extraction the case must be expanded slightly by the internal pressure before force can be transmitted to the sensor See the enclosed guide An Improved Technique for Utilization of Conformal Ballistics Sensor Calibration Data for methods of dealing with this topic 6 0 MAINTENANCE It is essential for normal operation of the Model 117B that the insulation resistance be ...

Page 9: ...Drawing Number 21086 Revision NR OPERATION MANUAL FOR CONFORMAL BALLISTICS PRESSURE SENSOR Series 117B 4 ...

Page 10: ...Drawing Number 21086 Revision NR OPERATION MANUAL FOR CONFORMAL BALLISTICS PRESSURE SENSOR Series 117B 5 ...

Page 11: ...mmunition under test to take into account transmissibility characteristics of the cartridge case for each lot of ammunition Because a certain amount of pressure is required to fully obturate the cartridge case in the test chamber a rather severe non linearity is exhibited over the first several thousand psi of input pressure i e until the cartridge case is pressed firmly against the inside chamber...

Page 12: ...or two percent of full scale is specified for quartz sensors An acceptable instrument is then defined as one whose data points all fall within this error band The sensitivity of an instrument so calibrated is simply the full scale output charge or voltage divided by full scale input pressure where F S output is determined by the straight line as defined above input S F output S F y Sensitivit Eq 1...

Page 13: ...ance between the cartridge case and the chamber wall Obturation of the cartridge occurs in this region However when obturation is complete the sensor output is then linear with pressure from this point up to maximum rated pressure In the past a common procedure has been to calibrate the conformal at the expected nominal pressure deriving a sensitivity at this point e g point Po Qo in figure 4 The ...

Page 14: ...he data from 20 consecutive calibration runs performed on a conformal sensor yielded a slope of 04 pC psi and an intercept of 7 500 psi determined by constructing the best fitting straight line as described previously Assume also that average expected peak pressure is 25 000 psi Approximate full scale charge output is pC 1000 pC psi 04 psi 25 000 Set range of charge amplifier output to 1000 pC vol...

Page 15: ... can be further refined to include a variable voltage source as a bias in series with the output of the charge amplifier This voltage bias can be set to simulate the offset pressure and then the true peak pressure can be read directly See figure 6 on the next page Example Assume a 20 round calibration lot and attendant graph yield a slope of 250 pC psi and an offset of 5 000 psi Proceed as follows...

Page 16: ...scale pressure to find full scale picocoulombs 40 000 psi x 250 pC psi 10 000 pC Set the charge calibrator for 10 000 pC and apply this signal to the input of the amplifier Now adjust the gain of the oscilloscope or charge amplifier if gain adjust is available to make full scale equal to a selected number of oscilloscope divisions For example 6 Div 40 000 psi Now the peak value can be determined f...

Page 17: ...Drawing Number 21210 Revision NR AN IMPROVED TECHNIQUE FOR UTILIZATION OF CONFORMAL BALLISTICS SENSOR CALIBRATION DATA 7 Adds offset pressure not available from PCB ...

Page 18: ...xy Epoxy Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Weight 0 45 oz 13 gm All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc OPTIONAL VERSIONS Optional versions have identical specifications and accessories as ...

Page 19: ...001 23 HOLD THIS DIMENSION TO OBTAIN FLUSH MOUNT 1 2 3 4 5 2 1 4 REVISIONS REV DESCRIPTION DIN B UPDATED 065A19 NOTE 50698 117 2205 90 PCB Piezotronics Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Piezotronics Inc 5 236 002 5 99 05 806 20 47 250 6 35 46 11 8 89 22 7 1 63 41 4...

Reviews: