background image

Model Number

115A04

CHARGE OUTPUT PRESSURE SENSOR

Revision: K

ECN #: 48697

[4]

Performance

ENGLISH

SI

Sensitivity(± 15 %)

1.40 pC/psi

20.3 pC/bar

Measurement Range

3.5 kpsi

241.4 bar

Maximum Pressure

4.35 kpsi

300 bar

Resonant Frequency

>125 kHz

>125 kHz

Non-Linearity

 0.3 % FS

 0.3 % FS

Environmental

Acceleration Sensitivity

<0.007 psi/g

<0.0005 bar/g

Temperature Range(Operating)

-4 to +662 °F

-20 to +350 °C

Temperature Range

(Survivable Limit)

-58 to +752 °F

-50 to +400 °C

Thermal Sensitivity Shift(20 to +250 °C)

2 %

2 %

[1]

Thermal Sensitivity Shift

(20 to +350 ºC)

5 %

5 %

[1]

Thermal Sensitivity Shift

(200 °C ± 50 °C)

1 %

1 %

[1]

Thermodynamic Error(

Δ

P mi)

<2%

<2%

[2]

Thermodynamic Error

(

Δ

P max)

<1%

<1%

[2]

Thermodynamic Error

(

Δ

P(short term drift))

<7.25 psi

<0.5 bar

[2]

Maximum Shock

2000 g pk

19,600 m/s² pk

Electrical

Capacitance(with cable)

120 pF

120 pF

[1]

Insulation Resistance(at room temp)

>10

13

 Ohm

>10

13

 Ohm

Physical

Sensing Element

UHT-12™

UHT-12™

Sensing Geometry

Compression

Compression

Housing Material

17-4 Stainless Steel

17-4 Stainless Steel

Diaphragm

17-4 Stainless Steel

17-4 Stainless Steel

Sealing

Welded Hermetic

Welded Hermetic

Electrical Connector

M4 x .35 Coaxial

M4 x .35 Coaxial

[3]

Cable Length

3.3 ft

1 m

Weight(without cable)

.085 oz

2.4 gm

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

ICP

®

is a registered trademark of PCB Group, Inc. 

OPTIONAL VERSIONS

Optional versions have identical specifications and accessories as listed for the standard model 

except where noted below. More than one option may be used.

NOTES:

[1] Typical.
[2] 9 bar IMEP; 1500 rpm
[3] Attached 006M40 cable with 070A54 BNC connector adaptor; o-ring sealed.
[4] See PCB Declaration of Conformance PS158 for details.

SUPPLIED ACCESSORIES: 

Model 065A69 Thermal baffle (1)
Model 066A17 Dummy plug (1)
Model 100-9169-50 FKM O-Ring (1)
Model PCS-1AP NIST Traceable Calibration, room temperature and 200°C. (1)

OPTIONAL ACCESSORIES: 

Model 040A39 Finishing tool (1)
Model 040A40 Extaction Tool (1)
Model 040A41 Mounting Key (1)

Entered: LK

Engineer: RPF

Sales: RWM

Approved: BAM

Spec Number:

Date: 9/24/2018

Date: 9/24/2018

Date: 9/24/2018

Date: 9/24/2018

41918 

3425 Walden Avenue, Depew, NY 14043

Phone: 716-684-0001
Fax: 716-684-0987
E-Mail: [email protected]

Summary of Contents for 115A04

Page 1: ...sure Sensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...ity regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme shock load or other environmental influence or prior to any critical test PCB Piezotronics maintains an ISO 9001 certified metrology laboratory and offers calibration services which are accredited by A...

Page 3: ...urchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers should direct all inquiries to their local distributor or sales office A complete list of distributors and offices can be found at www pcb com Customers within the United States may contact their local s...

Page 4: ... O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ...te O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at le...

Page 6: ...ion may be more difficult due to obstructions from other components deep bore holes or components which may be too hot to contact cabling When preparing mounting this style of sensor keep unsupported length at connector end of sensor to a minimum This will ensure the best quality measurements possible If a water jacket is to be breached the bore diameter tolerances of where the sensor s o rings wi...

Page 7: ...Never immerse the sensor in the ultrasonic bath without its cable connector 2 Prepare the bath with an aqueous alkaline pH 7 9 or cleansing solvent such as Isopropyl Alcohol and set the temperature to no more than 140 F 60 C maximum 3 Immerse the sensor into the bath only up to the sensor s shoulder or mounting hex 4 Set the ultrasonic output power so the power in the bath is no more than 50 W lit...

Page 8: ... Steel Sealing Welded Hermetic Welded Hermetic Electrical Connector M4 x 35 Coaxial M4 x 35 Coaxial 3 Cable Length 3 3 ft 1 m Weight without cable 085 oz 2 4 gm All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc OPTIONAL VERS...

Page 9: ......

Reviews: