OPERATION MANUAL FOR 

ICP

®

 HIGH PRESSURE SENSORS 

Series 108 & 109 

3.7 

CIRCUIT CONNECTIONS

 

 

The  figures  below  show  two  typical  circuit 
connections normally used to connect the 108 & 109 
Series to power units.   

 

The  first  figure  is  the  connection  scheme  used  for 
most  applications.    In  this  set  up,  the  signal  is  AC-
coupled from the bias voltage meaning the voltage at 
the "scope" terminal will be at a zero volt bias level. 

 

 

Normal Connection AC-Coupled Output

 

 

In  this  case,  the  system  discharge  time  constant  will 
be determined by the input resistance and capacitance 
of  the  power  unit.    Most  PCB  power  units  have 
approximately a 10 second discharge time constant. 
 

Alternate Connection to take Advantage of Sensor 
Discharge TC

 

 

The  second  figure  illustrates  an  optional  connection 
scheme  which  can  be  used  during  calibration  to  take 
full advantage of the sensor’s discharge time constant. 

 

 

 

With  this  arrangement,  the  readout  is  direct-coupled 
to  the  sensor  and  the  output  signal  will  be  added  to 
the DC bias voltage (11 VDC nominal). 

 

The  discharge  time  constant  will  then  be  determined 
only by the sensor and should be in the order of 2000 
seconds;  this  is  sufficient  for  most  types  of  static 
calibration methods. (See Calibration Section 4.0) 

    

 

Typical Connection

 

 

The  figure  above  shows  the  108  &  109  Series 
connected to a PCB Model 484B06 power unit.  The 
484B06  is  a  power  unit  that  can  be  AC  or  DC-
coupled. 

 

In  the  DC-couple  mode,  a  level-shifting  circuit 
removes  the  bias  level  and  provides  a  zero  adjust 
feature  at  the  front  panel.    The  system’s  discharge 
time  constant  is  determined  by  the  sensor  in  this 
mode. 

 

An  AC-coupled  mode  is  provided  for  normal  AC-
coupled operation. 
 

4.0 

CALIBRATION 

 

PCB  108  &  109  Series  sensors  are  supplied  with  a 
calibration 

certification 

from 

the 

factory.  

Recalibration services are provided at the factory for 
a nominal fee. 

 

Static calibration methods may be employed using the 
set-up shown in the Alternate Connection figure or by 
use of the Model 484B06 Power Unit as shown in the 
figure above – use the 484B06 in DC mode. 

 

Following  thermal  stabilization  of  the  sensor,  use  a 
high  pressure  pump  with  dial  reference  sensor  or  a 
dead  weight  tester  to  apply  pressure  in  desired 
increments to full scale.  Release pressure after taking 
the  reading  and  before  proceeding to the next higher 
pressure level. 

 

With  a  2000  second  discharge  time  constant,  1%  of 
the  signal  will  be  lost  in  20  seconds,  so  it  is 
imperative  that  the  pressure  setting  and  recording  of 
output be accomplished quickly. 

 

A  calibration  graph  can  be  plotted  using  output 
voltage vs. input pressure to determine sensitivity and 
linearity.

 

108 & 109 

SERIES 

002 SERIES 

      ICP

® 

POWER SUPPLY 

READOUT 

108 & 109 

SERIES 

ICP

® 

POWER SUPPLY 

 

002 SERIES 

READOUT 

ICP

®

 

POWER SUPPLY 

 

SERIES 002 

READOUT 

108 & 109 

SERIES 

Summary of Contents for 109C12

Page 1: ... psi accel comp ceramic coated integral diaphragm Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 6: ...ff diaphragm give these sensors a high natural frequency and linearity The microelectronic amplifier converts the high impedance voltage from the quartz package into a low impedance high level output signal See the Technical Information section of our website for more complete coverage of ICP instruments 3 0 INSTALLATION With a 3 8 24 M10x1 0 for metric mount mounting thread and flush diaphragm de...

Page 7: ... nicks and other imperfections which might cause leaks at high pressures 3 3 RECESSED MOUNT The recessed installation is shown in the Recessed Mount figure This type of mounting protects the sensor diaphragm from the effects of high flash temperatures and particle impingement due to blast effects thereby prolonging sensor life The recommended range of passage diameters is 0 090 to 0 125 inch 2 29 ...

Page 8: ...se of recessed installation In severe pyrotechnic environments sensor life may be severely limited with flush installation 3 5 FLASH TEMPERATURE EFFECTS The ceramic coating on the diaphragm of these sensors should render the flash thermal effect insignificant in most cases especially when recessed mounted However if more protection from flash thermal effects is required with the recessed mount the...

Page 9: ...e shows the 108 109 Series connected to a PCB Model 484B06 power unit The 484B06 is a power unit that can be AC or DC coupled In the DC couple mode a level shifting circuit removes the bias level and provides a zero adjust feature at the front panel The system s discharge time constant is determined by the sensor in this mode An AC coupled mode is provided for normal AC coupled operation 4 0 CALIB...

Page 10: ...w instrument several minutes to warm up and thermally stabilize When output from the power unit is connected to readout equipment a drift in the voltage signal will be noticed this is normal This voltage signal drift is caused by the charging of the coupling capacitor in the power unit and will cease within several minutes 6 0 POLARITY The 108 109 Series are designed to produce a positive output v...

Page 11: ......

Page 12: ...ghts in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Piezotronics Inc REVISIONS REV DESCRIPTION DIN F ADDED CERAMIC COATING NOTE 53103 CAUTION DO NOT TOUCH SEAL SURFACE WITH TAP WHEN TAPPING HOLE DIMENSIONS A B TO SUIT USERS REQUIREMENTS KEEP DIMENSION A AS SHORT AS POSSIBLE SEE OPERATING GUIDE FOR OPTIMUM P...

Reviews: