Page 24
Conf1G
PRG
SET
FUN
CaL1br
AnaLog
CapaC
SEns1t
nEt
dEad L
dspd1J
S1GnaL
CoUnts
opNodE
UpLoad
dnLoad
SET
FUN
0
0
0
0
0
0
0
0
0
0
SET
FUN
SET
FUN
SET
FUN
SET
FUN
SET
FUN
SET
FUN
SET
FUN
SET
FUN
SET
FUN
SET
FUN
PRG
Enter Value
PRG
PRG
Enter Value
PRG
PRG
Enter Value
PRG
PRG
Enter Value
PRG
PRG
Select Value
PRG
PRG
Read Value
PRG
Read Value
PRG
Select Value
PRG
PRG
PRG
Transmit setup data
to serial port
0
0
Receive setup data
from serial port
UpLoad
RECEIVING DATA FUNCTION
Receiving function from a serial of a file containing the setup data
that will be automatically set in the instrument.
dnLoad
SENDING DATA FUNCTION
Sending function from a serial of a file with the content of the setup
memory of the instrument.
EXAMPLE OF SETTING/CALIBRATION
You must weigh a tank, with empty weight of 750 kg and with a
capacity of 1000 liters, containing a product with a specific gra-
vity of 1.33 of which you want to read the weight with a display
resolution of 0.2 Kg.
Before proceeding with the configuration, you should make sure
that the load cells are connected properly to the unit and the tank
is empty. Then you can set the parameters.
They are used:
Nr 3 load cells with capacity of 1000 kg
Sensitivity of respectively 2.0015, 2.0008 and 1.9998 mV/V
(average value = 2.0007 mV/V)
Set the following values in the configuration parameters:
CAPAC
=
3000
SEnS1t
=
2.0007
nEt
=
1500
dEad L
=
0
dSpd1J
=
0.2
Make sure that the value read in the
S1GNAL
parameter corresponds
with the calibration weight of the system according to the following
proportion:
3000:2.0007=750:X
Where X is the value of the signal expressed in mV/V corresponding
to the theoretical value of the weight of the empty tank. The value
should be about 0.5 mV/V.
Now you can proceed to the calibration described in the following
paragraph, or you can exit the configuration menu by saving the
data entered.
The instrument should indicate the value corresponding to the weight
of the empty tank (for example 756.8).
You can re-access the configuration menu and enter the weight value
read in the
dEad L
parameter and enter the value 756.8.
Quit the configuration menu by saving the data.
For greater accuracy, prepare some sample weights or the pre-
weighed material on a certified scale and calibrate as described in
the next paragraph.