Page 25 / 44
5
Cleaning
Cleaning the device and its accessories is very important for compliance with hygienic
requirements and to avoid any cross-infection. Please always consider local regulations and
read this section carefully.
Before cleaning the device, the device must be switched off and removed from all connected
components (e.g. power supply unit).
Wipe the surface of the device with a cloth slightly dampened with mild detergent or
normal hospital bactericides or antiseptic solution. The following quantities of chemical
substances are allowed: ethanol: 70-80%, propanol: 70-80%, aldehyde: 2-4%. Do not immerse the
device and make sure that no liquid gets into the device. Dry the device with a lint-free cloth after
cleaning.
Disposable accessories (e.g. ear tips and other accessories marked for single use only on the package
label or data sheet) must be replaced between patients (or ears of the same patient) to avoid cross-
infection.
The ear probe test cavity must be used with a disinfected and clean new probe tip. In case of
contamination with pathological material or suspected dirt inside the cavity, please discontinue the
use of the test cavity. For external cleaning, please use a sterile alcohol wipe, typically containing
70% isopropyl alcohol.
It is recommended that parts which are in direct contact with the patient (e.g. headphone cushions)
are subject to standard disinfecting procedures between patients. This includes physical cleaning and
use of recognized disinfectants. The use of hygiene protective covers is recommended for
headphones (if available for the used headphone model).
For further information about cleaning instructions for accessories (e.g. ear probe) please refer to the
respective manual or data sheet of the accessory.
When using a cleaning agent, please refer to the manufacturer's data sheet of the cleaning agent for
the minimum time period in which the wipe has to be in direct contact with the surface of the device
or accessory to ensure effectiveness of cleaning.
The device and its accessories are provided non-sterile and are not intended to be sterilized.