Measurement Procedure

With the Xplorer GLX (Standalone Mode)

Follow these steps to use the colorimeter connected to a GLX 
without a computer. The colorimeter should be horizontal when it 
makes a measurement. If you connect it directly to the GLX with-
out a cable, keep the GLX horizontal.

1.

Connect the colorimeter to the GLX. The GLX detects the 
colorimeter and opens the Digits screen.

2.

Press 

 to highlight the data source field.

3.

Press 

 again to 

open the data source 
menu (right).

4.

Use the arrow keys to 
highlight the desired 
measurement and 
press 

. (Select the 

measurement corre-
sponding to the test kit that you will use.)

5.

(Optional) Perform the calibration procedure (see “Calibra-
tion” on page 3).

6.

Follow the instructions provided with your ezSample Snap 
Vial kit to mix your water sample with the reagent in the 
ampoule.

7.

Handling the ampoule by the tip, wipe the glass clean using a 
non-abrasive cleaning tissue. Avoid touching the glass after 
it is clean.

8.

Align the mark on the ampoule as illustrated and insert the 
ampoule into the colorimeter.

9.

Place the cap over the ampoule.

The GLX automatically monitors and displays live 
measurements in the Digits screen.

10.

(Optional) To start recording data, press 

.

To remove the measurement 
from the lower segment of the 
Digits screen: Press 

Press the down arrow to high-
light the data source field of 
the lower segment. Press 

 

again to open the menu. Use 
the arrow keys to highlight 

Clear Input

 and press 

With a Computer

Follow these steps to use the colorimeter with any PASPORT 
interface connected to a computer.

1.

Connect your PASPORT interface to the computer and start 
DataStudio.

2.

Connect the colorimeter to the PASPORT interface. 
DataStudio detects the colorimeter and creates a digits dis-
play showing all available measurement.

3.

(Optional) Perform the calibration procedure (see “Calibra-
tion” on page 3).

4.

Follow the instructions provided with your ezSample Snap 
Vial kit to mix your water sample with the reagent in the 
ampoule.

5.

Handling the ampoule by the tip, wipe the glass clean using a 
non-abrasive cleaning tissue. Avoid touching the glass after 
it is clean.

6.

Align the mark on the ampoule as illustrated and insert the 
ampoule into the colorimeter.

7.

Place the cap over the ampoule.

8.

Click 

Start

 to start recording and displaying data. (Or open 

the Experiment menu and select 

Monitor

 to display mea-

surements without recording them.)

DataStudio displays all available measurements. Read the one 
corresponding to the test kit that you used. (The other live mea-
surements are not meaningful and should be ignored.) 

To remove unneeded measurements from the digits display, click each 
measurement and press delete.