background image

®

Oxidation Reduction Potential Probe

Technical Support

2

012-15717A

(**SE-7287 100 ml or SE-7288 1000 ml Beakers)

Preparation of the Electrode

1. Combination ORP electrodes are shipped with a 

storage bottle with storage solution. Keep the 

solution bottle and solution for future use.

2. Remove the storage bottle from the electrode and 

thoroughly rinse the electrode with distilled water. 

Wipe carefully with a clean lab wipe.

Software Help

See the SPARKvue Help or PASCO Capstone Help for 

information about collecting, displaying, and analyzing 

data. 

In SPARKvue, select the HELP button in any screen 

including the Home Screen.

In PASCO Capstone, select PASCO Capstone Help 

from the Help menu, or press F1.

Checking Electrode Operation

1. Connect your ORP electrode to the input connector 

on the Wireless pH /ISE/ORP Sensor (or other pH 

measuring device). Ensure that the electrode 

connection is secure. 

2. Place the electrode into a beaker containing 

quinhydrone saturated pH 7.00 buffer. Stir gently. 

The mV reading E1 should be 86 ± 20 mV.

3. Remove the electrode from the buffer.  Rinse with 

distilled water and blot with a lab wipe.

4. Place the electrode into a beaker containing 

quinhydrone saturated pH 4.01 buffer. Stir gently. 

Record the mV reading E2. The difference between 

E2 and E1, (E2-E1), should be 175 ± 20 mV.

Reading a Sample with the Electrode

1. Rinse the electrode with distilled water and blot with 

a lab wipe. Place the electrode in a beaker 

containing the sample and a stir bar. Stir as before. 

Record the mV when the reading is sta-ble.

2. Remove the electrode from the sample, rinse the 

electrode with distilled water over the "waste" 

beaker. Blot the electrode dry with a lab wipe. The 

electrode is now ready to read the ORP readings of 

other samples.

Storing the Electrode

Short Term

Between measurements, immerse the ORP electrode in 

the storage solution.

Long Term

When storing for long periods, store the ORP electrode 

in the storage bottle which came with the electrode. 

Electrode Cleaning

Contamination of the sensing element often results in 

slow response and inaccurate readings. Clean the 

element by one of the following procedures:

1. Inorganic Deposits: Immerse electrode tip in 0.1 N 

HC1 for 10 minutes. Wash the tip with distilled 

water.

2. Organic Oil and Grease Films: Wash electrode tip in 

a liquid detergent and water.

3. After above treatment, soak the electrode tip in 

alcohol for 5 minutes and wipe dry, then, soak in 

quinhydrone saturated pH 4.01 for 15 minutes; rinse 

with distilled water afterwards.

NOTE: DO NOT ATTEMPT TO SAND OR POLISH THE 

SENSING ELE-MENT WITH SAND PAPER OR OTHER 

POLISHING MATERIAL!

Technical Support

For assistance with any PASCO product, contact 

PASCO at:

Limited Warranty

For a description of the product warranty, see the PASCO catalog.

Copyright

This PASCO scientific instruction sheet is copyrighted with all rights 

reserved. Permission is granted to non-profit educational institutions for 

reproduction of any part of this manual, providing the reproductions are 

used only in their laboratories and classrooms, and are not sold for profit. 

Reproduction under any other circumstances, without the written consent 

of PASCO scientific, is prohibited. Rev: 10/17

Address:

PASCO scientific

10101 Foothills Blvd.

Roseville, CA 95747-7100

Phone:

916-462-8384 (worldwide)

800-772-8700 (U.S)

Email:

[email protected]

Reviews: