Residual heat warning (HOT) 

 

To  prevent the risk of burns from the hotplate, this instrument has 

residual heat warning function. When heating is stopped, and the heating 

plate temperature is still above 50°C, the setting indicator will light up to 

warn that there’s potential risk of burns. When the hotplate temperature 

drops to below 50°C, the instrument will be automatically powered off. 

To turn off the instrument immediately, pull out the plug. In case of main 

power failure or disconnection, the residual heat warning will not work. 

 

Stirring 

 

Stirring is switched on  /  off by pressing the speed setting knob. The 

speed is set by rotating the knob in the range of 50-1500rpm.   

 

Faults 

 

 

The device can’t work when powered on 

-

 

Check if the power line is connected properly 

-

 

Check if the fuse is broken or loose 

 

Fault in power on self test 

-

 

Power off the device and then restart 

 

Actual speed can’t reach the set value 

-

 

Media in high viscosity  may probably cause abnormal speed 

reduction of the motor 

 

 

 

The device can’t be powered off when switched off 

-

 

Check if the residual  heat warning function  is still on and the 

hotplate temperature is above 50°C (the setting indicator is lit) 

 

 

Fault 

Fault description 

Solutions 

D2 light, 

D3 flash 

Internal temperature sensor 

PT1000 failure 

Re-start the device 

D1, D2 

light, 

D3 flash 

Motor failure 

Reduce the sample 

 

                     

Table 3 

 

D1 for “Heating”  indicator  light, D2 for “External temperature sensor” 

indicator light; D3 for “Power on” indicator light 

 

If the instrument still can’t work properly, please contact manufacturer/ 

supplier for technical support. 

 

Maintenance 

 

Proper maintenance can make the device work well and extend its life. 

 

 

Do not allow moisture to get into the device when cleaning. 

 

Disconnect the main plug when cleaning. 

 

Wear protective gloves when cleaning. 

 

Only use the recommended cleansing agents. 

Summary of Contents for PS-3401

Page 1: ...PS 3401 Operating Instructions Safety instructions 2 Inspection 3 Control 4 Trial run 6 Heating 6 Residual heat warning Hot 7 Stirring 7 Faults 7 Maintenance 8 Standards and regulations 8 Specifications 8 Warranty 9 PS 3401 Heater Stirrer ...

Page 2: ...ease the speed and reduce the speed if the medium splashes out of the container because the speed is too high the machine is not running smoothly the container moves on the heating plate Caution Only process and heat up any media that has a flash point higher than the adjusted safe temperature limit that has been set The safe temperature limit must always be set to at least 50 C lower than the fir...

Page 3: ...e mains voltage Don t cover the machine even partially e g with metallic plates or film which will result in overheating Ensure the heating plate is kept clean Protect the machine and accessories from bumps and impacts The minimum distance between the machines the minimum distance between the machine and the wall is 100 mm Inspection Unpack the device carefully and check for any damages that may h...

Page 4: ...012 15856A PS 3401 Instruction Manual 4 1 916 462 8384 800 772 8700 ...

Page 5: ...e safety temperature the machine stops heating It s the max temperature of 320 C by rotating the knob to the limit in the clockwise direction it ll be the min temperature of 50 C by rotating the knob to the limit in the anti clockwise direction 4 Temperature setting scale Reading the temperature setting value 5 Speed setting scale Read the speed setting value 6 Heating indicator It s lit when heat...

Page 6: ...e contact manufacturer supplier for technical support Warning DO NOT move the container when the device is working Heating With the digital temperature control the machine has two separate safe circuits The hotplate is kept at a constant temperature by a digital control circuit The hotplate temperature can also be monitored by another adjustable safe circuit The two temperature sensors PT1000 inte...

Page 7: ...use is broken or loose Fault in power on self test Power off the device and then restart Actual speed can t reach the set value Media in high viscosity may probably cause abnormal speed reduction of the motor The device can t be powered off when switched off Check if the residual heat warning function is still on and the hotplate temperature is above 50 C the setting indicator is lit Fault Fault d...

Page 8: ...lectronic product is subject to disposal and recycling regulations that vary by country and region It is your responsibility to recycle your electronic equipment per your local environmental laws and regulations to ensure that it will be recycled in a manner that protects human health and the environment To find out where you can drop off your waste equipment for recycling please contact your loca...

Page 9: ...0 320 Residual heat warning 50 C Dimensions W x D x H mm 250x148x85 Weight kg 1 9 Permissible ambient temperature C 5 40 Permissible relative humidity 80 Protection class acc to DIN EN60529 IP42 Table 5 Warranty The instrument is warranted to be free from defects in materials and workmanship under normal use and service for a period of five years from the date of invoice The warranty is extended o...