PS-3217 

Wireless Oxygen Gas Sensor 

012-16215A

To install the sensing element 

1. Unscrew the stopper from the sensor by turning

the stopper counterclockwise.

2. Remove the sensing element from the storage

bottle, then connect it to the sensor (Figure 1).

CAUTION: Make sure that the temperature and

humidity sensor does not get wedged under the

sensing element.

3. Screw the stopper back on the sensor.

NOTE: The sensing element only needs to be installed 

when using the sensor for the first time. The sensing 

element will not need to be removed until replacement is 

necessary (at least two years). 

Figure 1. Installation of the sensing element. 

To charge the battery 

1. Connect the micro USB cable to the micro USB

port on the sensor then connect the other end of

the cable to a USB charger.

2. Allow the battery to charge for at least three

hours.  The battery status  LED will turn green

when fully charged.

NOTE: The sensor can remain connected to a power 

source  when fully charged  without damaging the 

battery. The sensor automatically stops charging the 

battery when fully charged. 

To connect the sensor to software 

The sensor automatically connects with the software 

when connected to a device via USB. Use the procedures 

below to connect the sensor to the software via Bluetooth. 

SPARKvue 

1. Press and hold the power button until the

Bluetooth status LED starts flashing red.

2. Open SPARKvue and select Sensor Data.

3. Under Connected Devices, select the sensor that

matches your device ID.

PASCO Capstone 

1. Press and hold the power button until the

Bluetooth status LED starts flashing red.

2. Open Capstone and click Hardware Setup.

3. Select the sensor that matches your device ID.

To calibrate the sensor 

The sensor should be calibrated before using the sensor 

for the first time. It is not necessary to calibrate the sensor 

each time the sensor is used. Calibration is recommended 

if the O

2

 Concertation reading for an atmospheric sample 

is greater than ±1% from 20.9%. 

SPARKvue 

1. Collect an atmospheric air sample (see the

Collecting Samples procedure).

2. Connect the sensor to SPARKvue and set up

data collection.

3. Select the Hardware Setup icon.

4. Select the Calibration icon.

5. Select Calibrate.

6. Select OK to confirm calibration.