Figure 2.
Remo
Removve the st
e the stor
orage bottle
age bottle
1. Hold the pH Probe vertically so that the solution
will not spill out of the bottle.
2. Unscrew the plastic cap and remove the bottle.
Keep the storage solution for later use.
3. Push the bottle cap and the rubber washer up the
pH probe shaft to keep them out of the way
(Figure
Place the sensor in a solution
Place the sensor in a solution
Place the end of the pH probe in a solution when
taking measurements. Make sure that the bulb at the
end of the probe is completely immersed in the
solution.
Note: Do not immerse the BNC connectors in
the solution. The connectors are not rust proof and
can corrode.
To keep the probe in place, it is recommended to use
an Electrode Support (PS-3505). The Wireless Drop
Counter (PS-3505) can also be used to hold the probe.
Connect the sensor and collect data
Connect the sensor and collect data
Connect the sensor to SPARKvue or PASCO Capstone
to collect data.
SP
SPARK
ARKvue
vue
1. Turn on the sensor.
2. Open SPARKvue then click Sensor Data.
3. Select the Wireless pH Sensor that matches its device ID.
4. Select
the measurements you want to display.
5. Select a Template to display your data.
6. Click Start
to begin data collection.
P
PASCO Capst
ASCO Capstone
one
1. Turn on the sensor.
2. Open Capstone then click Hardware Setup
.
3. Select the Wireless pH Sensor that matches its device ID. Click
Hardware Setup
to close the panel.
4. Click Sensor Data
Sensor Data to display a graph.
5. Click Record
to begin data collection.
Calibr
Calibration
ation
Calibration is not always necessary, especially if you are measuring a
change in pH rather than absolute pH values. However, the pH Sensor
can be calibrated if more accurate measurements are necessary. See
the SPARKvue or Capstone online help for instructions on how to
calibrate the sensor.
St
Stor
orage and maintenance
age and maintenance
Cleaning the pr
Cleaning the probe
obe
For most applications, you can clean the probe using hot water with a
domestic cleaning detergent. The following table contains specific
cleaning recommendations for different types of solutions the probe
is exposed to.
SS
OLUTION
TYPE
R
R
ECOMMENDED
CLEANING
SOLUTION
lime and hydroxides
5-10% hydrochloric acid
organic fouling agents (fats, oils) limonene or acetone
algae and bacteria
diluted chlorine bleach
St
Storing the sensor and pr
oring the sensor and probe
obe
After each use, place the pH probe in the storage bottle filled with the
storage solution. The bulb at the end of the probe must be fully
immersed in the storage solution. It is not recommended to store the
probe dry since the probe will lose responsiveness. If needed, probe
restoration instructions are provided online on the Wireless pH
Sensor product page.
Replacement pH Storage Solution (SC-3507) can be purchased if any
solution is lost due to spillage or evaporation. If you wish to make
your own storage solution, instructions are provided online on the
Wireless pH Sensor product page.
For long-term storage, remove the battery to avoid damaging the
sensor due to battery leakage.
Replacing the batter
Replacing the batteryy
When the battery status light blinks red, the battery will soon need to
be replaced.
A
B
C
Figure 3. Sensor battery compartment. (A) First mark. (B) Second mark.
(C) Indicator.
To replace the battery:
1. Insert a coin into the battery compartment door. Turn
counterclockwise until the indicator (C) is aligned with the second
mark (B).
2. Remove battery compartment door, O-ring gasket, and battery.
3. Insert a new CR2032 battery in the compartment with the positive
(+) side facing up.
4. Insert the O-ring gasket in the compartment.
5. Place the door on the sensor, aligning the indicator with the
second mark. Turn the door clockwise until the indicator is
aligned with the first mark (A).