Model No. PS-2126A

 Collecting O

2

 Samples

2

012-11736E

3.

Carefully thread the deflector onto the Oxygen Sensing 
Element.

4.

Slide the rubber stopper over the deflector.

5.

Plug one end of the connecting cable into the Oxygen 
Sensing Element.

6.

Plug the other end of the connecting cable into the Sensor 
Amplifier.

7.

Connect the Sensor Amplifier to the PASCO interface.

8.

Refer to the documentation that came with the PASCO 
interface and software for further instructions for data 
collection.

Collecting O

2

 Samples

To collect an atmospheric sample, 
hold the bottle upright and open 
to the air. Insert the end of the 
sensing element into the bottle. 
Press the rubber stopper into the 
neck to seal the bottle.

To collect expired air or other gas 
samples, place the sensing ele-
ment inside a plastic bag. Press 
any atmospheric air out of the bag. Use a tube to fill the bag 
with the gas sample. Hold, tie, or clamp the bag closed round 
the tube and sensing element cable to contain the gas sample 
and exclude atmospheric air.

When doing an experiment that requires a faster than usual 
response time, the response time of the sensor may be 
increased by removing the white deflector and stopper from 
the Oxygen Sensing Element. This shortens the path to the 
sensor's membrane allowing it to be populated more rapidly 
with the sample. 

The purpose of the white deflector is to deflect air from an air 
stream into the sensing element and provide protection for the 
gas permeable membrane. It also holds the rubber stopper. 

While the white deflector is removed from the sensing ele-
ment, be careful not to place sharp objects near the membrane 
as puncturing the membrane could lead to damage to the sens-
ing element and electrolyte leaking from the membrane. 

Removal of the deflector is not required or recommended 
during typical operating conditions.

Warning: Do not allow the sensing element to contact liquids. Caution: 
Do not clean the sampling bottle in a dishwasher or autoclave.

Calibration

The sensor is factory calibrated. However, for greater accu-
racy, you can follow these steps to calibrate the sensor:

1.

With the sensor connected to a powered PASPORT inter-
face, place the sensing element in the empty sampling 
bottle. Outside air has an O

concentration 20.9%.

2.

Press the 

CAL (20.9%)

 button on the sensor and hold it 

for 3 seconds. The green light will flash on and off for 4 
seconds, indicating that calibration is in progress.

3.

After 4 seconds, the reading will stabilize at 20.9%. When 
calibration is complete, the light will stop flashing. 

Note: If the green light flashes rapidly after calibration, the sensing ele-
ment may be expired and it may need replacement.

Experiment: Oxygen 
Production from Liver Catalase

Equipment required:

 Oxygen Gas Sensor with sampling bottle, beef 

or chicken liver, hydrogen peroxide, pipette, mortar and pestle, ice 
bath, boiling water, and safety goggles.

Safety

: Always wear safety goggles when performing this experiment. 

Do not allow excessive pressure to develop in the sampling bottle.

1.

Place the liver in an open container and allow it to warm 
to room temperature. 

2.

(Optional) Calibrate the Oxygen Gas Sensor (see above).

3.

Grind the liver to a mushy consistency.

4.

Place 10 mL of hydrogen peroxide into the clean, dry 
sampling bottle.

Note: Perform step 5 through 7 carefully and quickly.

5.

With a pipette, extract 1 mL of blood/homogenate from 
the ground liver and release it into the sampling bottle.

6.

Insert the end of the sensing element into the bottle. Press 
the rubber stopper into the neck of the bottle just enough 
to close the bottle, but not too tightly

7.

Start data collection on the computer or interface.

8.

After about 300 seconds, stop data collection.

9.

Clean and dry the sampling bottle. Repeat steps 4 through 
8 with cold liver homogenate that has been chilled in an 
ice bath. Repeat again with liver homogenate that has 
been boiled in water.

10.

Compare your results from the three runs.