Pasco Scientific PS-2235 Reference Manual Download Page 4

®

Polarimeter

PS-2235

4

013-14434A

Insert the Sample Cell

Open the lid of the Polarimeter and place the Sample Cell in the 
chamber. The Sample Cell is an exact fit for the length of the 
space in the chamber. The direction of the Sample Cell in the 
chamber does not make a difference. Close the lid.

Calibration

1.

Pour distilled water into the 100 mm (1 dm) Sample Cell and 
insert it in the Polarimeter’s chamber. Close the lid.

2.

Start data collection and slowly rotate the analyzer with the 
thumb-wheel so you have data to cover 0° to 360°. For better 
analysis, you may want to take multiple passes in this data 
range.

3.

Stop data collection.

4.

Record the first angle above 0° where the light intensity is at 
a maximum.

To improve the accuracy of this selection, you may wish to apply 
a Gaussian fit.

The recorded angle indicates the position where the polarizer and 
analyzer are aligned without an optically active substance in the 
Sample Cell.

Data Collection

1.

Pour your sample into the 100 mm (1 dm) Sample Cell and 
put the Sample Cell in the Polarimeter’s chamber. Close the 
lid.

2.

Start data collection and slowly rotate the analyzer with the 
thumb-wheel so you have data to cover 0° to 360°. For better 
analysis, you may want to take multiple passes in this data 
range.

3.

Stop data collection.

4.

Record the first angle above 0° where the light intensity is at 
a maximum.

To improve the accuracy of this selection, you may wish to apply 
a Gaussian fit.

The recorded angle indicates the position where the polarizer and 
analyzer are aligned with an optically active substance in the 
sample cell.

The difference between the recorded angles represent the 
observed angle of rotation for your sample

Experiment Suggestions

Explore the optical activity of inorganic, organic and biochemical 
compounds.

Chiral compound (inorganic and organic) purity by optical 
rotation.

Solution concentration by optical rotation.

Determine enantiomeric excess in a mixture of dextro- and 
levo-rotatory compounds.

Kinetics studies where the reactants or products are optically 
active.

For more resources go to 

www.pasco.com/polarimeter

Battery 

Replacement 

If the Polarimeter fails to recharge, the battery may need to be 
replaced. Contact PASCO Technical Support for more 
information about battery replacement. It is recommended that 
the unit be returned to PASCO for battery replacement, if needed.

Troubleshooting the Polarimeter

If the Polarimeter loses Bluetooth connection and will not 
reconnect, try cycling the ON/OFF button. Press and 

briefly

 

hold the button until the status LEDs blink in sequence, and 
then release the ON/OFF button to turn on the Polarimeter. 

If the Polarimeter stops communicating with the computer 
software or tablet application, try restarting the software or 
application. If the problem remains, try cycling the ON/OFF 
button. 

As a last resort, push a pin into the Reset port on the bottom 
of the Polarimeter. 

Technical Support

For assistance with any PASCO product, contact PASCO at:

For more information about the product and the latest revision of 
this Reference Guide, visit the PASCO web site at 
www.pasco.com and enter “Polarimeter” or the product number 
(PS-2235) in the Search window.

Limited Warranty

The Polarimeter has a five year warranty. For a more thorough description 
of the product warranty, see the PASCO catalog.

Address:

PASCO scientific
10101 Foothills Blvd.
Roseville, CA 95747-7100

Phone:

+1 916 462 8384 (worldwide)
877-373-0300 (U.S.)

Web:

www.pasco.com

Email:

[email protected]

Summary of Contents for PS-2235

Page 1: ...USB Cable 100 mm Sample Cell Thumb wheel Chamber Holder Quartz Window Initial Step Charge the Battery Included Equipment Part Number Polarimeter PS 2235 100 mm Sample Cell USB Cable Required Item Data Collection Software see www pasco com software Replacement Item Replacement Sample Cell PS 2234 PASCO Capstone SPARKvue Mac OS X Mac OS X Windows Windows iOS Android Capstone UI 5400 SPARKvue PS 2400...

Page 2: ...the analyzer measures the angle and the photo detector records light intensity The data collection software produces a graph of light intensity compared to angle and can be used to analyze the data to determine various properties of the optically active substance The Sample Cell is precisely designed so that the distance inside the cell between the two quartz windows is 100 millimeters mm Therefor...

Page 3: ...y the stability of the mechanical assembly accuracy should be close to the encoder resolution or 0 09 Light Source The LED light source has an output between 585 nm and 596 nm or an optimal output of 589 nm The LED has a built in lens that directs the light output to an exit cone that is 15 off the center axis which increases the amount of usable light that is directed along the optical axis A sec...

Page 4: ...t the observed angle of rotation for your sample Experiment Suggestions Explore the optical activity of inorganic organic and biochemical compounds Chiral compound inorganic and organic purity by optical rotation Solution concentration by optical rotation Determine enantiomeric excess in a mixture of dextro and levo rotatory compounds Kinetics studies where the reactants or products are optically ...

Page 5: ... device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation CE Statement This device has been tested and found to comply with the essential requirements and other relevant provisions of the applicable EU Directives Product End of Life Disposal Instructions This electronic product is subject to disposal...

Reviews: