PAS Systems International PAS V Calibration Manual Download Page 11

 

 

c.  After 5 minutes, repeat the air blank making sure that the flashlight is on. If none 

of the bar-graph indicator lights come on (after observing for 15-20 seconds), the 
air blank check is OK. 

 

d.  If  however,  one  or  more  of  the  indicator  lights  come  on  again,  this  clearance 

process should  be repeated.  If after two attempts  to clear the unit, the indicator 
lights  are  still  coming  on  when  the  air  blank  is  performed,  there  are  two 
possibilities: 

 

i.  The  unit  requires  an  internal  zero  adjustment.  To  have  this 

adjustment made, the unit has to be returned to the manufacturer. 

 

ii.  The air in the test room contains a significant amount of alcohol. To 

determine if this is the case, air blanks should be made away from 
the  test  room  to  see  whether  the  unit  gives  a  zero  reading  when 
removed from the test area. 

 
5.      Once the air blank has been performed with a satisfactory result, the PASV unit 

is ready for sensitivity testing.     

 

PAS V Sensitivity Checking 

 

 
 

NOTE

If  the  low  battery  lamp  starts  flashing  at  any  point  in  the  procedure, 

recharge the battery  before proceeding.

     

 
1.      Turn on the flashlight to activate the fuel cell heater.  Leave it on for 5 minutes so 

that the fuel cell comes up to design temperature before starting the test.     

 

2.      Check that the inlet and outlet ports of the PAS V are free of obstruction. 

 

3. 

Attach the Calibration Connector Tube to the outlet on the dry gas regulator. 

 
 
 
 
 
 
 
 
 
 
 
 
 

NOTE:  Throughout  the  procedure,  the  test  conductor  should  check  for  a  flashing 
battery  light  to  ensure  that  the  battery  is  not  running  down  during  the  testing 
procedure. 

Summary of Contents for PAS V

Page 1: ...Passive Alcohol Sensor CALIBRATION MANUAL PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 www pasintl com Copyright 2015 by PAS Systems International Inc All Rights Rese...

Page 2: ......

Page 3: ...Bath Solution optional 2 Operating the Alcohol Wet Bath Simulator 3 Readying the PAS V 5 PAS V Sensitivity Checking 6 PAS V Calibration 7 Dry Gas Method Equipment Checklist for Dry Gas Sensitivity Che...

Page 4: ...ece that connects the outlet of the wet bath simulator to the intake port of the PAS V Available from PAS Systems A NHTSA approved Wet Bath Simulator 500 milliliters of 005 ethanol alcohol solution To...

Page 5: ...volumetric flask and fill the flask up with distilled water This completes your mixture of stock solution You may keep your stock solution in a refrigerator for mixing calibration solution as necessar...

Page 6: ...jar from the simulator by unscrewing it from the top assembly Caution DO NOT remove Top Assembly and expose the heating element to open air with power turned ON This will result in damage to the heati...

Page 7: ...hts come on after observing for 15 20 seconds the air blank check is OK d If however one or more of the indicator lights come on again this clearance process should be repeated If after two attempts t...

Page 8: ...sampling began 6 Disconnect the Calibration Assembly and observe the bar graph display All but one of the bars should light up There should be two green four yellow and two of the three red bars lit 1...

Page 9: ...activate the Sensor Control Button on the PAS V 6 Look for the green sensor light and the yellow pump PMP light and listen for the sound of the air pump that should come on immediately after the Sens...

Page 10: ...ct the inlet outlet ports of the PAS V to make sure that they are not clogged 3 Conduct an air blank a Turn on the flashlight and wait a minimum of 1 2 minutes for initialization process to complete b...

Page 11: ...r blanks should be made away from the test room to see whether the unit gives a zero reading when removed from the test area 5 Once the air blank has been performed with a satisfactory result the PASV...

Page 12: ...d light up There should be two green four yellow and two of the three red bars lit 15 20 seconds from the time the pump stops 9 Record the number of red bars that light up this should be two Repeat th...

Page 13: ...ough the opening and into the calibration adjustment pot screw Turn the calibration adjustment all the way up counter clockwise The adjustment is free to rotate approximately 270 Take care not to use...

Page 14: ...Problem Reported Checked By Action Taken Date Returned to Manufacturer Date Returned to Field The example shown provides a means for recording any reports from Testers of problems with the PAS V units...

Page 15: ......

Page 16: ...PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 www pasintl com 800 660 SNIF 540 372 3431 Printed 10 15...

Reviews: