Particulate Systems MA-1040 Operator'S Manual Download Page 2

                                                                                                                                                                                                     

© Particulate Systems 2011. All rights reserved. Printed in the U.S.A.

Summary of Contents for MA-1040

Page 1: ...MA 1040 Magnetic Analyzer Operator s Manual V2 0 P03 42801 00 March 2011...

Page 2: ...Particulate Systems 2011 All rights reserved Printed in the U S A...

Page 3: ...products sold by MICROMERITICS are the products of reputable manufacturers sold under their respective brand names or trade names We therefore make no express or implied warranty as to such products...

Page 4: ......

Page 5: ...f Contents Introduction 1 Description 1 Components and Connectors 2 Top Panel 2 Side Panel 3 Specifications 4 Setting up the Analyzer 5 Reference Standard Test 6 Cleaning the Sample Holder 7 Weighing...

Page 6: ......

Page 7: ...g insula tion The MA 1040 can also be used to detect low magnetic iron levels in food precious gems pharmaceuticals and many other materials The MA 1040 has been referenced in American National Standa...

Page 8: ...s of magnetic content in PPM WEIGHT GRAMS control Used to enter the sample weight Range 000 999 grams ZERO control Used to zero the reading on the display when the sample chamber is empty Sample Chamb...

Page 9: ...MA 1040 Magnetic Analyzer Description May 2011 3 Side Panel Power connector For connecting the power cord On Off switch For turning the analyzer on and off Power connector On Off switch...

Page 10: ...rature Stable between 10 and 45 C 50 to 113 F operating 10 and 55 C 14 to 131 F non operating Physical Height 16 76 cm 6 6 in Width 25 40 cm 10 in Depth 13 34 cm 5 25 in Weight 6 8 kg 15 lbs Electrica...

Page 11: ...r on a steel table If unavoidable place a 3 4 in thick piece of plywood underneath the instrument Ensure a constant room temperature fluctuations may result in inaccurate readings 1 Place the analyzer...

Page 12: ...ote of each reading 5 Take an average of the three readings from steps 3 and 4 The average for the three readings should be within 1 of the value printed on the reference standard or 0 5 PPM whichever...

Page 13: ...o zero 2 Using a proper lab technique such as stirring mix the sample to evenly distribute all grains 3 Remove the sample holder from the balance and fill with sample until it is just below the over f...

Page 14: ...rate for about three to five seconds after which time the display may drift 4 Remove and empty the sample holder Place the empty sample holder back into the sample chamber and record the reading shown...

Page 15: ...rder Part Number Item and Description P03 25803 00 Sample holder P03 34810 0010 Reference standard 10 PPM P03 34810 0050 Reference standard 50 PPM P03 34810 0100 Reference standard 100 PPM P03 34810 0...

Page 16: ......

Reviews: