BioCapt
®
Adjustable Height Quick Guide
PRIOR TO FIRST-TIME USE:
Unpack the impactor head(s)
using gloves
. Fingerprints on freshly
manufactured stainless steel may become lasting cosmetic marks.
Wash
the entire impactor surface
and mechanical parts.
Rinse
with DI or WFI water to remove residual detergent.
Dry
completely
before first autoclave sterilization.
PRIOR TO AUTOCLAVING:
Clean
the BioCapt impactor with chemicals compatible with 316L
stainless steel.
Remove
any cleaner residue and ensure all surfaces are dry to avoid discoloration.
Compatible chemicals:
•
Isopropyl alcohol (IPA) 70/30
•
Sodium Hypochlorite 1% or 6% (NaClO)
•
Hydrogen Peroxide, up to 6% (H
2
O
2
)
•
H
2
O
2
and acetic acid
•
Dec-Clean, typically 0.7 (Vol %) in WFI water
•
P2C-1, typically 1-5 Vol% in DI or WFI water
•
Quaternary ammonia solution
•
Phenol (C₆H₅OH)
•
Neutral detergent residue remover
BETWEEN SAMPLING:
Ensure no agar residuals are present on the impactor head.
Wash
the entire
impactor surface and mechanical parts. Allow cleaner residues to
evaporate until dry
.
Part Name
部件名称
有毒或有害的物质和元素
铅
(Pb)
汞
(Hg)
镉
(Cd)
六价铬
(Cr(VI))
多溴联苯
(PBB)
多溴联苯醚
(PBDE)
机械部件
X
O
X
X
X
X
O: 表示用于部件的所有同族物质中所含的有毒或有害物质低于SJ/T11363-2006规
定的限 度要求。
X: 表示用于部件的至少一种同族物质中所含的有毒或有害物质高于SJ/T11363-2006
规定的限度要求。
BioCapt
®
is a registered trademark of Particle Measuring Systems, Inc. © 2021 Particle Measuring Systems. All rights reserved. Reproduction or translation of any part of this work without the permission of the copyright
owner is unlawful. Requests for permission or further information should be addressed to Particle Measuring Systems, Inc. at 1-800-238-1801. P/N 1000026334 Rev A
1.
Turn locking pin counterclockwise with flathead
screwdriver until secure.
2.
To unlock, rotate pin clockwise.
L O C K / U N L O C K
LOCKING PIN (LOCKED)