BioCapt

®

 Adjustable Height Quick Guide

PRIOR TO FIRST-TIME USE:

  

Unpack the impactor head(s) 

using gloves

. Fingerprints on freshly 

manufactured stainless steel may become lasting cosmetic marks. 

Wash

 the entire impactor surface 

and mechanical parts. 

Rinse 

with DI or WFI water to remove residual detergent. 

Dry

 completely 

before first autoclave sterilization.

PRIOR TO AUTOCLAVING:

  

Clean

 the BioCapt impactor with chemicals compatible with 316L 

stainless steel. 

Remove

 any cleaner residue and ensure all surfaces are dry to avoid discoloration.

Compatible chemicals:

• 

Isopropyl alcohol (IPA) 70/30 

• 

Sodium Hypochlorite 1% or 6% (NaClO)

• 

Hydrogen Peroxide, up to 6% (H

2

O

2

)

• 

H

2

O

2

 and acetic acid

• 

Dec-Clean, typically 0.7 (Vol %) in WFI water

• 

P2C-1, typically 1-5 Vol% in DI or WFI water

• 

Quaternary ammonia solution

• 

Phenol (C₆H₅OH)

• 

Neutral detergent residue remover

BETWEEN SAMPLING:

  Ensure no agar residuals are present on the impactor head. 

Wash

 the entire 

impactor surface and mechanical parts. Allow cleaner residues to  

evaporate until dry

.

Part Name

部件名称

有毒或有害的物质和元素

(Pb)

(Hg)

(Cd)

六价铬

(Cr(VI))

多溴联苯

(PBB)

多溴联苯醚

(PBDE)

机械部件

X

O

X

X

X

X

O: 表示用于部件的所有同族物质中所含的有毒或有害物质低于SJ/T11363-2006规

定的限 度要求。
X: 表示用于部件的至少一种同族物质中所含的有毒或有害物质高于SJ/T11363-2006

规定的限度要求。

BioCapt

®

 is a registered trademark of Particle Measuring Systems, Inc. © 2021 Particle Measuring Systems. All rights reserved. Reproduction or translation of any part of this work without the permission of the copyright 

owner is unlawful. Requests for permission or further information should be addressed to Particle Measuring Systems, Inc. at 1-800-238-1801. P/N 1000026334 Rev A

1. 

Turn locking pin counterclockwise with flathead 

screwdriver until secure.

2. 

To unlock, rotate pin clockwise.

L O C K / U N L O C K