How to Calibrate Your AirSentry II IMS
Figure 1
AirSentry II Calibration Station
Calibration Station
The AirSentry II Calibration Station allows users to calibrate the AirSentry II IMS to a specific calibration
stream concentration. An internal pneumatic diagram of the calibration station is included on page 9 for
reference, if needed.
To set up the calibration station:
1.
Place the calibration station in the area where the AirSentry II IMS analyzers will be calibrated.
2.
Unscrew the two flathead screws on the oven cover.
Figure 2
Oven cover screws
3.
Remove the cover to the oven.
4.
Using the supplied 15/16” socket and nut driver, unscrew the Teflon® nut allowing access to the
span oven.