Parma VAF-A Operating Manual Download Page 6

РА1.007.001РЭ-001 

PARMA Ltd.

 

 

The present Operating Manual is intended for study of  technical data, 

device, operation principle and rules of application of voltamperephasometer 
“Parma ВАФ-A”. 

NORMATIVE REFERENCES 

GOST 12.3.019-80. Electric tests and measurements. General safety 

requirements. 

GOST 14254-80 (IEC 529-76(2-83)) Electric articles. Casings. Protec-

tion levels. Designations. Test methods. 

GOST 22261-94 Facilities of measurements of electric and magnetic 

values. General Specifications. 

GOST  26104-89  (IEC  348-78)  Electronic  measurement  facilities. 

Safety specifications, test methods. 

GOST  29254-91  Electromagnetic  compatibility  of  technical  compo-

nents.  Equipment  for  process  measurement  and  control.  Noise  immunity 
specifications and test methods. 

GOST  R  51318.22-99    (CISPR  22-97)  Electromagnetic  compatibility 

of  technical  equipment.  Radio  disturbance  from  information  technology 
equipment. Limits and test methods. 

GOST R 51317.4.2-99 (IEC 61000-4-2-95) Electromagnetic compati-

bility  of  technical  equipment.  Immunity    to    electrostatic  dis-
charge.Requirements  and test methods. 

GOST R 51317.4.4-99 (IEC 61000-4-4-95) Electromagnetic compati-

bility of technical equipment. Immunity  to  electrical fast transient/ burst. Re-
quirements and test methods. 

GOST 29073-91 Electromagnetic compatibility of technical equipment 

for  process  measurement  and  control.  Electromagnetic  noise  immunity. 
General provisions.  

DEFINITIONS,  DESIGNATIONS  AND  ABBREVIA-

TIONS 

The present Service Instructions contain the following terms: 
Device - Voltamperephasometer “Parma ВАФ-A”. 
GOST - Russian State Standards. 
Measuring  Clamp  -  current  measuring  clamp  for  measurment  chan-

nel. Identified by serial numbers to be represented on the display at switch-
ing on of the device, on the side of the corresponding channel. Or by serial 
numbers designated in the device logbook.  

Reference  Clamp  -  current  measuring  clamp  for  reference  channel.  

Identifying the same way as Measuring Clamp.  

Us (Is)   - Maximum scale value of voltage (current). 

РА1.007.001РЭ-001 

23 

PARMA Ltd.

 

7.6.4.4 

VAF is considered to pass the test if the range of effective 

value  of  direct  voltage  is  0.0  –  460  V  and  the  relative  permissible  error 
doesn’t exceed the value determined in p. 7.1. 

7.6.5  Verification  of  effective range  and   measurement  error of 

effective value of alternating voltage measurement. 

7.6.5.1 

Connect the device to UPPU-1M system.  

7.6.5.2 

Set input voltage frequency on 50 Hz. 

7.6.5.3 

Switch  the  device  on  and  make  sure  that  zero  value  of 

voltage is shown on indicator. 

7.6.5.4 

Make measurements in the following range points: 10, 100, 

200, 300, 460 V. 

7.6.5.5 

VAF is considered to pass the test if the range of effective 

value  of  alternating  voltage  is  0.0-460  V  and  the  relative  permissible  error 
doesn’t exceed the value determined in p. 7.1. 

7.6.6  Verification  of  effective range  and   measurement  error of 

alternating current  measurement. 

7.6.6.1 

Connect the device to UPPU-1M system. 

7.6.6.2 

Set the frequency of output voltage on 50 Hz. 

7.6.6.3 

Switch the device on and make sure that zero value of cur-

rent is shown on indicator. 

7.6.6.4 

Make measurements in the following range points: 100 mA, 

1, 3, 5, 7.5, 10.0 A. 

7.6.6.5 

VAF is considered to pass the test if the range of effective 

value of alternating current strength and the relative permissible error doesn’t 
exceed the value determined in p. 7.1. 

7.6.7  Verification  of  effective range  and   measurement  error of 

phase shift measurement. 

7.6.7.1 

Connect  the  device  to  UPPU-1M  system,  connect  voltage 

supplied  to  terminals  of  reference  channel  and  join  clip-on  instrument  of 
measurement  channel  to  lead  wire.  Install  clamp,  marking  plate  in  down 
position. 

7.6.7.2 

Set the frequency of output current and voltage on 50 Hz. 

7.6.7.3 

Set the output voltage of UPPU-1M system on 100 V, out-

put current on 5 A, current-voltage phase shift equal to zero. Control phase 
shift by phase shift meter F2-34. 

7.6.7.4 

Measure the phase shift. 

7.6.7.5 

Repeat  the  measurement  for  90  and  180  degree  phase 

shift. 

7.6.7.6 

Turn off the voltage. 

7.6.7.7 

Join  clamp  of  reference  channel.  Install  both  clamps  on 

lead wire of the system in the same manner. 

7.6.7.8 

Measure the phase shift. 

Summary of Contents for VAF-A

Page 1: ...ё РА1 007 001РЭ 001 PARMA Ltd Saint Petersburg VOLTAMPEREPHASOMETER Parma ВАФ А OPERATING MANUAL РА1 007 001РЭ ФГУ ...

Page 2: ...РА1 007 001РЭ 001 27 PARMA Ltd ...

Page 3: ...РА1 007 001РЭ 001 26 PARMA Ltd РА1 007 001РЭ 001 3 PARMA Ltd Appearance of Parma VAF A voltamperephasometer with its accessories ...

Page 4: ...ntal factors 8 5 Overall dimensions in packing are 330x235x105 mm 8 6 Gross weight is no less than 3 kg 8 7 The device is suitable for the use in the sphere of state metrological supervision in case when there are two unopened seals on its housing on the right and on the left belonging to the manufacturer enterprise and a Gosstandart body carrying out verification of the device 8 8 The device seal...

Page 5: ...or use without release verifica tion 7 8 Verification interval 7 8 1 Verification is made one time a year 8 MARKING PACKING AND SEALING 8 1 The following data is marked on a device name type manufac turer s trade mark national compatibility mark after type registration serial num ber year of release designation of input and output circuits nominal voltage type of current and power supply frequency...

Page 6: ...oltage current РА1 007 001РЭ 001 23 PARMA Ltd 7 6 4 4 VAF is considered to pass the test if the range of effective value of direct voltage is 0 0 460 V and the relative permissible error doesn t exceed the value determined in p 7 1 7 6 5 Verification of effective range and measurement error of effective value of alternating voltage measurement 7 6 5 1 Connect the device to UPPU 1M system 7 6 5 2 S...

Page 7: ...o value of voltage is shown on indicator 7 6 4 3 Make measurements in the following range points 10 100 200 300 460 V РА1 007 001РЭ 001 7 PARMA Ltd Um Im Measured value of voltage current 3 SAFETY REQUIREMENTS 3 1 The device belongs to class II according to GOST 26104 re garding electric injury protection 3 2 Protection level of the device case according to GOST 14254 80 IEC 529 is IP54 3 3 Only p...

Page 8: ...on linear distortion of supply line voltage no more than 15 Maximum consumed power no more than W From network power source 3 From battery source 0 5 Direct current power supply voltage is 6 V 4 galvanic cells R14P AR SIZE C 1 5 V SUM 2 Battery supply is switching automatically when network supply cable connector is removed from a jack РА1 007 001РЭ 001 21 PARMA Ltd Operation is made during Names ...

Page 9: ...thstand test voltage of effective value of 2 kV frequency of 50 Hz during 1 minute without damage 4 4 4 Electric insulation between network circuits and equivalent cir cuits and the case of the device can withstand test voltage of effective value of 4 kV frequency of 50 Hz during 1 minute without damage 4 5 Service requirements in overload conditions 4 5 1 The device can withstand overload of 2 Us...

Page 10: ...front panel 3 there is a power switch 5 display lighting button 6 double terminal 7 for connec tion of phase A voltage signal or Uref single terminal 8 for connection of phase B voltage signal double terminal 9 for connection of phase C voltage signal or Um jack 10 for connection of a clamp Iref jack 11 for connection of a clamp Im display 12 The jack for connection of the supply line of 220 V is ...

Page 11: ...ll be switched to required mode automatically when the signal comes to terminal B 6 2 9 2 Correct determination of phase shift order is possible if only all three phases are connected in accordance with marking on the de vice see fig 12 13 6 2 9 3 The result of phase shift determination is displayed as a text for example Figure 12 NO PHASE С NO PHASE А φuu 30 L 220 0V φui 30 C 999 mА 180 270 0 90 ...

Page 12: ...ent of signal appearance on Ureference terminals the device will be switched to the proper mode automatically 6 2 7 3 The upper line of display will indicate the value of phase shift between Ureference voltage and voltage applied to measurement chan nel Umeasuring Read indication from the upper line of the display 6 2 7 4 Prefix of current voltage phase shift uu Figure 10 6 2 8 Measurement of volt...

Page 13: ...is intended for formation of signals necessary for measurement of phase shifts of galvanically isolated circuits These may be current Im or voltage Um The reference channel consists of three comparators 1 2 3 two of them to form reference signals Im and Um 1 2 and one 3 to work with voltage signal Ub in the mode of determi nation of phase alternation order Comparators convert the input sinusoid si...

Page 14: ...А1 007 001РЭ 001 15 PARMA Ltd 6 2 2 Measurement of effective value of alternating voltage and its frequency 6 2 2 1 To measure the effective value of alternating voltage and its frequency apply alternating voltage to terminals named Um see fig 5 read display indication The measured voltage value is placed in the right upper part of the display and has a dimension of voltage V frequency value is in...