18
Parks Medical Electronics, Inc.
Aloha, Oregon U.S.A.
Maintenance and Service
Troubleshooting Guide
This table lists common problems and suggestions for troubleshooting them. If a problem persists after
these actions have been taken, contact Parks Medical Electronics, Inc.
Problem
Situation
Suggested Actions
(continued)
3.0
Hard to hear
blood sounds (poor
signal: noise ratio)
(continued)
3.1
Too much noise
3.1.4
Check for noisy probe:
Clean and dry the crystal end of the probe.
1.
Connect the dry probe to the Doppler unit, turn
2.
the Doppler unit on, and turn the volume up.
Wiggle the probe plugs.
3.
If you hear a crackling sound, the connection
is noisy.
Clean and tighten the probe plugs:
4.
Wipe the pin in the center of the plug and
•
the
fl
ange around it with contact cleaner or
isopropyl alcohol.
Squeeze the
fl
ange with your
fi
ngers (or
•
gently with pliers). Make the
fl
ange form
a smaller circle so that it grips the jacks on
the Doppler panel more tightly.
Check for noise again.
5.
If there is still noise, bend the cable here:
6.
Send the cable back to the factory for new
plugs if the noise originates here.
Flex the cable near the crystals. If the cable
7.
is noisy near the transducer, it cannot be
repaired. Replace probe.
Press gently on each
8.
crystal with your
fi
ngernail.
If a crackling or popping sound occurs, the
probe cannot be repaired. Replace probe.
3.1.5
Check for electromagnetic interference:
Noise that is probably caused by an
electromagnetic
fi
eld:
a tone or whistle that does not change pitch
•
a buzz or a hum
•
a noise like a radio station
•
a ticking sound that seems periodic
•
Con
fi
rmation criteria:
The noise worsens when the probe is held or
•
applied to the patient.
The noise disappears when the Doppler unit is
•
used at another site
Possibly if the noise lessens when the probe is
•
disconnected
Summary of Contents for 915-BL
Page 2: ...2 Parks Medical Electronics Inc Aloha Oregon U S A ...
Page 24: ...24 Parks Medical Electronics Inc Aloha Oregon U S A ...
Page 26: ...26 Parks Medical Electronics Inc Aloha Oregon U S A Appendix ...
Page 27: ...27 915 BL Dual Frequency Doppler Appendix ...
Page 28: ...28 Parks Medical Electronics Inc Aloha Oregon U S A Appendix ...