6
Figure 5. Digital MDM® Mixer Head Installation
11. With the (4) screws, secure the mixer head, the
bezel, and mounting shelf to the cabinet surface.
See Figures 4 and 5.
12. Tighten the (4) screws and snap the screw caps
over the screw heads.
1.5 Bag Tee and Scavenging Control Valve
Installation
If a scavenging control valve is to be installed, it is
installed next to the bag tee. See Figure 6. The bag
tee/scavenging control valve is supplied assembled to
the appropriate mounting hardware.
CAUTION
During daily operation, the bag tee/scavenging
control valve mount is occasionally stressed, for
example, when the breathing bag is removed or
installed. The attachment to the cabinet surface
must be sufficiently rigid to withstand the stress.
The bag tee/scavenging control valve assembly
may be installed on plywood attached to studs or
on a wood cabinet surface. It must not be installed
onto plaster board or other soft wall materials.
All hoses and gas connections must be
accessible
after the installation is complete.
Figure 6. Flush Mounting the Bag Tee and
Scavenging Control Valve.
1.5.1
Flush Mounting the Bag Tee/Scavenging
Control Valve For flush mounting, the bag
tee/scavenging control valve is supplied
assembled to a mounting plate; see Figure 6.
1. Position the bag tee/scavenging control valve
template at the desired location on the cabinet
surface.
2. Mark the (4) mounting holes and the (2) holes for
the gas hoses.
3.
Using a #20 drill, drill (4) 0.161” mounting holes.