Parker 023-064 Series Technical Manual Download Page 9

 

 

023-064X Technical Manual 

 

 

70-PHW-1018-R3.2 MTA 8.5 023-064X Refrigerant Leak Detector Manual.docx 

 

4

 

Operation 

 

The power supplied should be between 17 to 28 vac or between 21 to 30 vdc. If the power is too high 
or too low the sensor will not work properly. 
 
Each time the unit is powered on with proper power, the relay is disabled and the analog outputs stay at 
0 during the warm up period. This is to allow time for the sensor to warm up and to avoid false alarm 
due to high readings of the cold sensor. The warm up timer may be cancel by pressing any arrow. The 
reading may take a few more minutes to stabilize as the sensor may need more warm-up time if it was 
not powered for many days. The warm up time can be changed to up to 255 sec with the 52 W/U user 
setting. After power-up it will take 48 hours for maximum accuracy to be obtained, although accuracy is 
very good after an hour. 
When the gas concentration is higher than the threshold set by AL1 for the time A1Del the relay is 
energized. The display will display 1 at the bottom left of the display. The relay will de-energized A1Off 
seconds after the gas concentration drops below the threshold. 
AL1 is for relay settings, AL2 is for alarm 2 settings and AL3 is for alarm 3 and buzzer settings. 
See section 3.2 

User settings list

 to change any setting. 

The buzzer may sound when alarm 3 activates. To silence the buzzer, press any arrow. 
  
See section 3.3 

Test

 for override and test commands. 

 

5

 

Calibration 

Procedure specifically for refrigerant sensors: 
1.  Use certified bottled calibration gas mixed with air only, not nitrogen. Ensure that sensors are 

powered on for a minimum of the break-in period for the sensor. For these sensor types, this is two 
days. 

2.  Inject bottled span gas and wait until the gas reading stops going up. The span gas used must be 

within the range of sensor’s scale.

 

Use a flow rate of 0.1 LPM to 0.5 LPM. The gas fitting to sensor 

should have a small outlet hole. If not, the pressure will increase and distort the reading (high). 

3.  Adjust the blue potentiometer on the module to set the sensor reading displayed to that of the bottle 

mix. Do not change the factory calibration constants in calibration mode as above. 

4.  Remove span gas mix and supply air. Sensor will return to zero. 
5.  If sensor does not return to zero, it needs replacement 
 
Mode conveniently you can order a new replacement sensor module factory calibrated 
See section 1.3 

Ordering information

 

 

Summary of Contents for 023-064 Series

Page 1: ...or by any means electronic mechanical photocopying recording or otherwise without the prior written permission of Parker Hannifin Canada Micro Thermo Technologies is a trademark of Parker Hannifin Ca...

Page 2: ...ettings list 8 3 3 Test 8 4 Operation 9 5 Calibration 9 6 MT Alliance installation 10 7 To add a missing Sensor model in the list 12 Safety Information This manual describes information required to in...

Page 3: ...master slave operation or central control via a controller Pre calibrated plug and play sensor modules avoid the need to recalibrate when upgrading expired sensors Impact resistant water resistant enc...

Page 4: ...C 88 F decreasing linearity to 50 rh at 40 C 104 F 1 5 User Interface Back lit LCD display shows gas concentration user settings and alarm status Red LED alarm indicators gas level 1 and 2 High inten...

Page 5: ...to 5 feet 1 to 2 meters from the floor when the ceiling is 7 to 10 feet high If the ceiling height is higher than 10 feet for example for heavy equipment the carbon monoxide detectors are installed at...

Page 6: ...ual pull station panel 24Vac 5VA Transformer Sensor GND Sensor U Input To sensor node Binary Input from limit switch AL1 AL2 AL3 PWR BI S L H S R141b 123 10 Active Sensor Alarm PPM reading Jumper Up P...

Page 7: ...ermine the threshold of Sensor A at which the relay should operate default is 400ppm Setting 1 A1Del is the set time default is 30 sec 2 A1Off is the reset time default is 20 sec The relay reset by it...

Page 8: ...55 44 C temperature display 0 off 1 on Temp 0 1 0 45 audio alarm 3 off on Aud 0 1 1 49 Keyboard security lock 0 off 1 on KBL 0 1 0 50 Temperature modify correct TMod 9 9 C 51 High temperature alarm li...

Page 9: ...concentration drops below the threshold AL1 is for relay settings AL2 is for alarm 2 settings and AL3 is for alarm 3 and buzzer settings See section 3 2 User settings list to change any setting The bu...

Page 10: ...tration Type Local Source Micro Thermo Manufacturer and the Sensor Model and gas type you are going to monitor pre defined sensor model will be available in version MTA 8 5 For previous version create...

Page 11: ...alarm You can use predefined values by clicking on Pick Alarm Settings button Select the Hardware tab to review the sensor manufacturer and model and where you can select the node and input where this...

Page 12: ...nge accordingly to the table at section 1 3 Ordering information Electrical point values are constant and should always be 9 45V and 2V respectively Please pay special attention to put 1 of full range...

Reviews: