background image

-2- 

 

GENERAL SAFETY RULES

 

 

Please read all instructions. Failure to follow all instructions listed below may result in injury. If the 
equipment is used in a manner other than as specified in these operating instructions, the 
protection provided by the equipment may be impaired. Always wear eye protection. 

 

DO NOT OPERATE 

unit from a DC power source. 

 

Warning: 

Do not change the AC/DC selector switch with the probe on. 

 

Do not abuse the power cords. Never carry the instrument by the cord or attempt to unplug the 
instrument using the cord. Always operate the instrument with the standard installed cord. 
Changing or using a damaged cord can increase the risk of electrical shock. The cord should be 
checked periodically for any damage. 

 

Do not position the instrument such that it would be difficult to operate the disconnect device (plug) 
on the end of the power cord. 

 

The outside housing should remain intact and solid. Any damage, chipping, or separating exposing 
internal wires is a hazard. Instruments should not be used in this condition. The outside housing 
should be periodically checked for damage.

 

 

Duty Cycle: 

The 400 series are designed for a 50% duty cycle. Continuous operation may cause 

overheating and damage the Contour Probe. 

 

Operating Environment: 

Temperature: 32° to 104°F (0° to 40°C). Relative humidity: 10% to 95%, 

non-condensing. 

 

Shipping and Storage Environment: 

Temperature: 40° to 140°F (4.44° to 60°C). Relative 

humidity: 5% to 95%. Vibration and shock: As encountered in normal shipping and handling with 
no degradation. 

 

General Cleaning 

The outside surface of the instrument can be periodically 
wiped clean with a clean cloth and a mild general purpose 
cleaner. Avoid using cleaners such as lacquer thinner, or 
mineral spirits that could damage the outside housing. 

 

Never attempt field service. All 400 series Contour Probes 
should be returned to the factory for repairs. 

 

OPERATION: 

Connect the instrument plug into a 

Grounded power outlet of proper voltage. Place the 
Contour Probe legs upon the work surface with the 
suspected defect at right angles to the legs (Good contact 
will produce the best results). Push the test switch to 
energize the instrument. Lightly dust or float dry magnetic 
inspection powder over the area being inspected. Defect 
indications will be revealed in a direction shown on figure 2. 
Turn the Probe 90 degrees from the first test and repeat the 
process. This method may also be utilized when applying a 
wet medium. Check all procedures and standards for 
further details involving inspection specifics. 

 

FIG. 2 

Reviews: