parkell TurboSENSOR D560-110 Instructions For Use Manual Download Page 5

The sheath and ultrasonic insert must be 

washed in soap and water either manually or 

in an ultrasonic cleaner, and rinsed in running 

water prior to bagging them for processing in 

a steam autoclave.

•  Sheaths and inserts may be sterilized in any 

conventional steam autoclave following 

manufacturer’s instructions. A typical steam 

cycle is 250°F-268°F / 121°C-131°C for 30 

minutes at a minimum of 15 psi, followed by a 

15 minute minimum cool-down period. After 

sterilization, inspect the device in the 

autoclave bag for integrity. If suspicion about 

the item exists, discard it and order a 

replacement from Parkell or your dealer.

• 

DO NOT USE DRY HEAT OR CHEMCLAVE 

on the sheath or the insert.

HOW TO REPLACE THE FILTER DISK

The water filter disk coming off of the rear hose 

of the unit should be replaced when it becomes 

dark or clogged with debris so that water flow is 

not blocked (1-3 months). Replacement Filters 

are available from Parkell (Stock No. D419 

– ten per package).
1. Close any water valves, and bleed all water 

lines of residual water pressure.

2. Disconnect the scaler water supply hose 

from the water supply at the quick-connect.

3.  Remove the old filter disk by holding it in one 

hand and loosening it by twisting it off the 

hose connectors on either side. They rotate 

in opposite directions.

4.  Install the new filter disk in the same position 

as the old one, and tighten the connectors in 

reverse.

5. Reconnect the scaler water supply hose to 

the water supply via the quick-connect and 

check for leaks.

6.  Note: If the filter becomes clogged and you 

do not have a replacement disk, the hose 

connectors may be locked together 

TEMPORARILY without the filter disk to 

allow emergency operation only.

TROUBLESHOOTING

  Unit does not operate (“power on” 

indicator does not light):

• Transformer cable disconnected from wall 

outlet or scaler. Reconnect properly.

• Wall outlet not live. Reconnect properly.
• Foot Pedal or cable is damaged. Contact 

Parkell.

• Unit fuse has failed. Contact Parkell.

   Insert does not vibrate properly: 

•  Faulty, damaged or worn insert. Replace.
• Insert not correctly seated in handpiece. 

Reinsert properly.

• Power control not correctly adjusted. 

Readjust.

  Excess heat at handle or tip of insert: 

• Faulty, damaged or worn insert. Replace.
• Too little water flow. Insert may be blocked. 

Inspect and clear blockage, or replace 

insert.

• Power too high for the amount of water. 

Increase water flow.

• Excessive hand pressure applied. Correct 

technique.

  No water spray when foot pedal is 

depressed:

• Water line blocked or kinked. Correct.
• Water passage in the scaling insert clogged. 

Inspect and clear blockage, or replace 

insert.

• Water filter clogged. Replace.
• No water supply connected. Correct.

  Insert does not go into handpiece:

•  Insert stack is bent. Straighten plates 

carefully by hand and reinsert.

• O-ring is too tight. Lubricate O-ring with 

water and use twisting motion to seat insert.

• Autoclavable handpiece sheath is not 

attached to handpiece.

  Insert falls out of handpiece or water 

leaks from front of handpiece:

• O-ring damaged or worn. Replace O-ring on 

insert and/or handpiece.

  Water leaks at base of handpiece:

• Orange O-ring at end of handpiece is 

missing or damaged.

Summary of Contents for TurboSENSOR D560-110

Page 1: ...AC power disconnect device for the scaler The transformer should be located above and away from any sources of water that may enter the unit As with all electrical devices the unit should not be immer...

Page 2: ...Electric Shock Class 1 Type B applied part Protection Against Ingress of Liquids Foot Switch Scaler IPX1 drip proof Power Supply IPX0 Ordinary Mode of Operation of Equipment Intermittent 10 minutes o...

Page 3: ...the base with one hand and turn the sheath counter clockwise with the other taking care not to twist the hose The sheath will unscrew and slip off 2 To replace the sheath gently thread it onto the ha...

Page 4: ...ust power and water controls accordingly Do not test a scaler tip for vibration on your fingers while the unit is operating This is not a valid test of how scaling feels to the patient Hold the handpi...

Page 5: ...quick connect and check for leaks 6 Note If the filter becomes clogged and you do not have a replacement disk the hose connectors may be locked together TEMPORARILY without the filter disk to allow em...

Page 6: ...you have any questions or problems with the installation or use of your TurboSENSOR call Technical Support Service at 1 800 243 7446 M F from 8 15 AM to 5 15 PM EST Outside of the USA or Canada call...

Page 7: ......

Page 8: ...31 249 1242 E mail info parkell com www parkell com THIS PRECISION DENTAL DEVICE WAS DESIGNED MANUFACTURED AND IS SERVICED IN THE UNITED STATES OF AMERICA BY European Authorized Representative Not a d...

Reviews: