DEVICE DESCRIPTION

The Parkell TurboSENSOR™ is an autotune 

magnetostrictive ultrasonic tooth scaler that 

operates at either 25 KHz or 30 KHz.  

The TurboSENSOR automatically detects 

whether the handpiece contains a 25 KHz or 

30 KHz insert, and switches to the correct 

operating frequency. The TurboSENSOR has 

an expanded low-power range to improve 

patient comfort during debridement, and a 

power-boosting, user-controlled Turbo feature 

to increase scaling power when needed for 

heavy deposits. The unit is designed to work 

with Parkell inserts. For information about 

usage with non-Parkell inserts, please see  

our warranty policy.

INTENDED USE/INDICATIONS:

For removal of calculus, plaque and oral debris 

from teeth during dental prophylaxis/scaling. 

 

CONTRAINDICATIONS TO USE 

Because of the potential for electromagnetic 

interference, this device should not be used on 

patients or by clinicians with cardiac 

pacemakers, internal defibrillators, 

intracorporeal fluid pumps or any other 

implantable electronic devices, or in close 

proximity to sensitive patient monitoring 

devices such as pulse oximeters.

WARNINGS

• The water supply to the scaler should always 

be turned off whenever the device is being 

connected, disconnected, or when not in use. 

• The power supply cord functions as the AC 

power disconnect device for the scaler. 

• The transformer should be located above and 

away from any sources of water that may 

enter the unit. As with all electrical devices, 

the unit should not be immersed in water or 

other liquids. Do not reach for the device if it 

has fallen into liquid until power is 

disconnected, and do not use the device after 

it has fallen into liquid. Return the device to 

Parkell for servicing. 

• Modification of this device will void the 

warranty, and may violate safety codes, 

endangering the patient and/or the clinician.

• This equipment is not suitable for use in the 

presence of a flammable anesthetic gas 

mixture (when used along with air or oxygen).

A00414revL0218

INSTRUCTIONS FOR USE 

STOCK NO. D560-110 

(110VAC)  

STOCK NO. D560-230 

(230VAC) 

STOCK NO. D560-100 

(100VAC)

24-HOUR SAFETY HOTLINE 

1-800-535-5053

CAUTION: RX ONLY

 

This manual, in whole or in part, should not be considered a substitute for formal training in Ultrasonic Scaling.  

Appropriate professional education is STRONGLY RECOMMENDED prior to using this device clinically.

25KHZ OR 30KHZ INSERTS: 

Device automatically switches 

operating frequency to match 

(insert not included).

“ON” INDICATOR: 

Lights only when 

footswitch is 

depressed.

“TURBO” LIGHT: 

Indicates the device is operating 

in high-power “Turbo” mode  

(footswitch fully depressed).

“PERIO” LIGHT: 

Indicates low-power range where 

large changes in knob produce 

subtle changes in tip amplitude.

WATER CONTROL 

KNOB

AUTOCLAVABLE REMOVABLE  

HANDPIECE SHEATH.

Summary of Contents for TurboSENSOR D560-110

Page 1: ...AC power disconnect device for the scaler The transformer should be located above and away from any sources of water that may enter the unit As with all electrical devices the unit should not be immer...

Page 2: ...Electric Shock Class 1 Type B applied part Protection Against Ingress of Liquids Foot Switch Scaler IPX1 drip proof Power Supply IPX0 Ordinary Mode of Operation of Equipment Intermittent 10 minutes o...

Page 3: ...the base with one hand and turn the sheath counter clockwise with the other taking care not to twist the hose The sheath will unscrew and slip off 2 To replace the sheath gently thread it onto the ha...

Page 4: ...ust power and water controls accordingly Do not test a scaler tip for vibration on your fingers while the unit is operating This is not a valid test of how scaling feels to the patient Hold the handpi...

Page 5: ...quick connect and check for leaks 6 Note If the filter becomes clogged and you do not have a replacement disk the hose connectors may be locked together TEMPORARILY without the filter disk to allow em...

Page 6: ...you have any questions or problems with the installation or use of your TurboSENSOR call Technical Support Service at 1 800 243 7446 M F from 8 15 AM to 5 15 PM EST Outside of the USA or Canada call...

Page 7: ......

Page 8: ...31 249 1242 E mail info parkell com www parkell com THIS PRECISION DENTAL DEVICE WAS DESIGNED MANUFACTURED AND IS SERVICED IN THE UNITED STATES OF AMERICA BY European Authorized Representative Not a d...

Reviews: