2.2 Unpacking and first use
Caution!
If the product is installed by Park House Healthcare it is not the responsibility of
the user to perform the activities described in this item.
The bed is shipped by Park House Healthcare assembled and in a wooden, open-work chest. The bed
should be unpacked indoors to protect from possible damage
.
Prior to installation you must:
-
Read the this User Manual carefully
-
Remove all packaging and materials protecting the bed
Caution!
Package may be marked with following signs:
- multiuse package
- package for recycling
low-density polyethylene packaging
LDPE
Package waste is recyclable and should be segregated before removed to a disposal
site.
Returnable packaging should be returned to the producer of the medical product
..
-
Remove any additional equipment
-
Remove the wheel blocks and release the central locking system
-
With help of another person carefully remove the bed from the palette and lock the brakes
Caution!
Do not hold the plastic ABS covers whilst moving the bed
-
Ensure the equipment provided fits the documents enclosed
-
Ensure all accessories are fitted in accordance with the instructions stated In item
4 Optional
elements
-
Install the product in its required destination ensuring the environment meets the
requirements as described in
item 1.2 General requirements
-
If the place where the bed has been installed is equipped with earth grounding point you
should connect the bed with it by means of earth ground wire connected to earth grounding
socket in the bed connecting panel,
Caution!
All electric devices connected directly with the patient should be connected to the
bed earth grounding point
Connect the bed to the mains ensuring there is suitable electrical supply
Check all functions are in correct working prior to patient use
All bed functions should work silently and smoothly, without stopping or creaking etc.
Any noise produced from the bed and electric servo-motors does not exceed 65dB(A) measured at 1m
from the bed