Operating Functions
a) Transport
Mode
If required, disconnect the hoses from the pump and join together.
The system will equalise and retain seat pressure once the
pump has been shut off.
b) Infection
Control
Should the seat become infected then please refer to our
disinfection protocol for appropriate cleaning processes.
Support Setting Procedure
It is important to follow the correct support setting procedure to ensure the
patient receives adequate support (lift) while achieving maximum pressure
relief and comfort. Failure to do this could result in the patient neing put at
risk.
1. Lie or sit the patient on the cushion.
2. Reset pressure dial on pump according to patient weight guide (PWG).
Ensure the correct pressure setting has been selected for either lying
or sitting position.
3. MAXIMUM PATIENT WEIGHT 159 kgs / 25 stones
Weight in KG
Weight in lb
Weight in Stones
& lb
Pressure
Range
31 - 85kg
85 - 113kg
113 - 141kg
141 - 159kg
70 - 187lb
187 - 249lb
249 - 310lb
310 - 350lb
5 stone - 13 stone
33% turn away
from soft
66% turn away
from soft
Firm
Soft
13 stone - 17 stone
11lb
17 stone 11lb - 22
stone 2lb
22 stone 2lb - 25
stone
4. Cleaning
The following guidelines are suggested by Park House Healthcare Ltd
as being suitable infection control procedures. Further information is
available.
PUMPS
General cleaning may be affected by using a cloth dampened with a mild
detergent and water solution. This approach may be followed either by
wiping with a sodium hypochlorite solution to a dilute of 1000 ppm or by
using an alcoholic wipe.
Do not use hyper carbonate or phenol based cleaning solutions.
Do not use any abrasive compounds or cleaning pads.
SEAT CUSHIONS
General Cleaning
Using a single use wipe, clean the cover with a solution of neutral detergent
and hand hot water. Rinse thoroughly with clean water and a damp single
use wipe. NOTE: The top, bottom and all four sides of the cushion, including
under the zip
fl
aps MUST be cleaned.
Disinfecting the cover
If the cover is heavily soiled or has been exposed to bodily
fl
uids such as
blood, it will require a more thorough cleaning procedure:
Wipe the cover using a single use wipe and a 0.1% Chlorine Solution
(1,000ppm) and cold water. If required a 1% Chlorine Solution (10,000ppm)
and cold water can be used. Rinse thoroughly with clean water and a damp
single use wipe. Make sure the cover is completely dried before re
fi
tting to
the cushion.
Frequent or prolonged exposure to higher concentration disinfectant
solutions may prematurely age the fabric cover. Surfaces must be protected
during use and rinsed and thoroughly dried after application of a disinfectant.