TROUBLESHOOTING
Contact the manufacturer or distributor:
– in the event of faults that are not listed in this chapter
– if the suggested procedure does not correct the fault
Fault
Possible cause
Remedy
No aerosol is
coming out of
the nebuliser.
The nebuliser nozzle
is blocked.
Clean the nebuliser.
The connection
tubing is not connec-
ted properly.
Check that the tubing con-
nectors are connected firmly
to the compressor and the
nebuliser.
The connection
tubing is leaking.
Replace the connection
tubing.
7
TECHNICAL DATA
7.1 General nebuliser data
Size
5
10 cm × 9.5 cm × 4.5 cm
Weight
5
27.5 g to 29.5 g
Operating gases
Air, oxygen
Minimum compressor flow
3.0 l/min.
Minimum operating pressure
0.5 bar/50 kPa
Maximum compressor flow
6.0 l/min.
Maximum operating pressure
2.0 bar/200 kPa
Minimum fill volume
2 ml
Maximum fill volume
8 ml
5) Without mouthpiece; unfilled.
– 39 –