eFlow
®
rapid
41
en
5
CLEANING AND DISINFECTION
Nebuliser handset
The nebuliser handset (and the aerosol
head) must be cleaned immediately after
each use, and disinfected once a day.
Preparation
Disassemble the nebuliser system into its
constituent parts:
• Pull the connection cord out of the
nebuliser handset:
• Remove the mouthpiece from the
nebuliser handset.
• Carefully pull the blue expiratory
valve (
the slot.
• Open the medication reservoir and pour
out the medication residue.
• Disassemble the medication cap.
To do this, press the cap seal (
downwards and out of the cap (
• Open the aerosol chamber.
• Remove the aerosol head.
To do this, gently press the engaging
hooks on the aerosol head together
from the side:
• Remove the inspiratory valve from the
aerosol chamber.
Info:
The expiratory valve is held in place
by a thin fixing clip so that it cannot
get lost. This clip must not be
detached from the mouthpiece.
Summary of Contents for eFlow Rapid 178G1005
Page 1: ...Gebrauchsanweisung Instructions for Use nebuliser system...
Page 6: ...2 eFlow rapid 04 18...
Page 32: ...28 eFlow rapid 04 18...
Page 58: ...54 eFlow rapid 04 18...
Page 108: ...104 eFlow rapid 04 18...
Page 134: ...130 eFlow rapid 04 18...
Page 160: ...156 eFlow rapid 04 18...
Page 186: ...182 eFlow rapid 04 18...
Page 259: ......