6.8 CONNECTING THE LH(C/M) VIA USB:
Unfasten the protective cap on the USB connector of the LH(C/M):
Connect a USB cable with a micro-USB connector to the LH(C/M):
Refer to the “IHM_Capteurs” software manual (I158F) for the connection and configuration
procedure.
6.9 REPLACING THE DESICCANT SACHET (SILICA GEL TO PROTECT AGAINST MOISTURE):
When the ochre granules in the silica gel sachet attached to the LH(C/M) turn red, replace them
with a replacement sachet provided after removing it from its protective wrapper:
Desiccant sachets can be reactivated by oven-drying for 4 hours at 120°C. The granules should
turn back to an ochre colour.
Warning: Previously used sachets can be reused after they have been oven-dried.