WARNINGS AND PRECAUTIONS

 

 

Warnings

 

 

-

Read these instructions carefully to ensure proper use of the Parasym device. 

-

Do not inhale or swallow small parts. 

-

Do not wrap the lead wire around the neck. 

-

Do not apply stimulation in the bath or shower or while sleeping. 

-

Do not get the Parasym device wet.  

-

Do not apply stimulation while driving, operating machinery, or during any activity in which 
electrical stimulation can put the patient at risk of injury. 

-

Do not apply stimulation over open wounds or rashes, or over swollen, red, infected, or inflamed 
areas or skin eruptions (e.g., phlebitis, thrombophlebitis, varicose veins). 

-

Do not apply stimulation over, or in proximity to, cancerous lesions. 

-

Do not apply stimulation over the patient’s neck or mouth because this could cause severe 
muscle spasms resulting in closure of the airway, difficulty in breathing, or adverse effects on 
heart rhythm or blood pressure. 

-

Do not apply stimulation across the patient’s chest, because the introduction of electrical current 
into the chest may cause rhythm disturbances to the patient’s heart, which could be lethal. 

-

Do not apply electrodes near the thorax because the introduction of electrical current may 
increase the risk of cardiac fibrillation. 

-

Stimulation should not be applied across or through the head, directly on the eyes, covering the 
mouth, on the front of the neck, (especially the carotid sinus), or from electrodes placed on the 
chest and the upper back or crossing over the heart. 

-

Use this device only with the electrodes, and accessories recommended by the manufacturer as 
using others may be unsafe. Any electrodes that have current densities exceeding 2 mA/cm

2

 may 

require special attention of the operator. 

-

Potential hazard from simultaneous connection of a patient to a high frequency surgical ME 
equipment and the device that may result in burns and possible damage to the device. 

-

Operation in close proximity (e.g. 1 m) to a shortwave or microwave therapy ME equipment may 
produce instability in the stimulator output. 

-

The long-term effects of the chronic use of the device have not been evaluated. 

-

Contact your healthcare provider (HCP) if your symptoms worsen. 

-

Electronic monitoring equipment such as EKG alarms may not operate properly when the 
Parasym device is in close proximity while being used.  

-

Consult with your HCP before use because the device may cause changes in cardiac rhythm. 

 

 

Safety and efficacy of the Parasym device have not been evaluated in the following patients, and 
therefore it is not indicated for: 

-

Patients with an active implantable medical device, such as a cardiac pacemaker, hearing aid 
implant, or any implanted metallic or electronic device 

-

Patients with a history of baseline cardiac disease or atherosclerotic cardiovascular disease, 
including congestive heart failure (CHF), known severe coronary artery disease or recent 
myocardial infarction (within 5 years).  

-

Patients with diagnosed bradycardia or an irregular heart rhythm.  

-

Patients who have had surgery to cut the vagus nerve in the neck (cervical vagotomy). 

-

Patients diagnosed with narrowing of the arteries (carotid atherosclerosis). 

-

Patients whose pain syndromes are undiagnosed. 

-

Pediatric patients. 

-

Pregnant women. 

Summary of Contents for tVNS

Page 1: ...Parasym tVNS Device Quick Start Guide...

Page 2: ...1...

Page 3: ...for you that you can use every day Two AA alkaline batteries are included so you can start straight away Before using your device please make sure you have read all the precautions warnings and safet...

Page 4: ...the electrodes and accessories recommended by the manufacturer as using others may be unsafe Any electrodes that have current densities exceeding 2 mA cm 2 may require special attention of the operato...

Page 5: ...or cracked The device casing is cracked dented or appears to be damaged Discontinue use if you experience Lightheadedness dizziness or chest pain Excessive skin irritation Caring for Your Device Do no...

Page 6: ...6 of participants discontinued use due to side effects where the most common side effect reported was skin irritation at the stimulation site which typically ceased shortly after use The following si...

Page 7: ...Instructions For Use 6...

Page 8: ...7...

Page 9: ...Electrode Placement Connect the electrode to the tragus of the left ear running the cable behind the neck over the right shoulder 8...

Page 10: ...9...

Page 11: ...into channel 1 3 Electrode Placement 3 1 Clip the electrode to your left tragus See diagram on Page 8 Make sure this is a good connection with both gold plates having a firm contact with the skin 4 Pu...

Page 12: ...s of use per day This can be achieved by pressing the timer button to 60 min The device will automatically shut down after 60 minutes Please find a convenient time each day using the treatment at the...

Page 13: ...ception turn the device off then turn the device on and repeat this step to find a lower intensity Note If no stimulation is felt at Level 36 apply a small amount of moisture to the tragus and repeat...

Page 14: ...r treatment session should be one level below the sensory level where it can be perceived or felt For most people this is between Level 1 Level 15 however some people need to go to the maximum to find...

Page 15: ...ectedly This is likely to be the power save function If no buttons are pressed and the intensity level is zero the unit will power off automatically after a few minutes There is no stimulus simulation...

Page 16: ...Visit our Help Center Our Help Centre has answers to most common questions You can access this at www parasym co hc 15...

Page 17: ...ort team Reach out to us via our support messaging widget on our website www parasym co and we ll get right back to you You can also email us directly at hello parasym co and this will also go through...

Page 18: ...mmerse the device in water 4 Do not use any other cleaning solution than soap and water Testing the Device The device can be tested by making sure the batteries are in the device correctly pressing th...

Page 19: ...s shown in the image on page 8 Note the safety and efficacy of the Parasym device has not been established in individuals with Implantable medical devices a history of cardiac disease or arrhythmia di...

Page 20: ...ctangular Mode 5 Modes are available Mode A Constant Stimulation Mode B Burst I Two train per second 250ms on 250ms off 25 pulses within one train pulse width 200 S Mode C Burst II Two train per secon...

Page 21: ...ects on the environment Symbols Explanation Type BF Applied Part Refer to Instructions for Use Compliance with WEEE regulations Manufacturer European Representative IP22 Protected against access to ha...

Page 22: ...Version 4 2 www parasym co 21...

Reviews: